Skip to content

LORRAINE 3.5mm Medial Distal Humerus Plate, 6 Hole B-160100Recalls & reported problems

Bonebridge AG · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07640187462067), received between Jul 22, 2025 and Jul 22, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Bone Fracture(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 22, 2025Fracture; Fracture

    ON (B)(6) 2025, DR. (B)(6) REPORTED SCREW BREAKAGE AND LOOSENING IN A PATIENT TREATED WITH LORRAINE MEDIAL AND POSTEROLATERAL DISTAL HUMERUS PLATES. THE ISSUE OCCURRED ONLY ON THE MEDIAL PLATE. THE INITIAL SURGERY TOOK PLACE ON (B)(6) 2025. THE PATIENT IS A 58-YEAR-OLD MALE (180 CM, 80 KG) WITH NO KNOWN COMORBIDITIES OR SMOKING HISTORY. HE SUSTAINED A COMPLEX OPEN AO 13C FRACTURE AS PART OF A POLYTRAUMA. SURGICAL FIXATION WAS PERFORMED USING MEDIAL AND POSTEROLATERAL LORRAINE PLATES WITHOUT INTRAOPERATIVE COMPLICATIONS. AT 6 WEEKS POSTOPERATIVE, HEALING WAS CONSIDERED TO BE PROGRESSING NORMALLY; HOWEVER, RADIOGRAPHS SHOWED VISIBLE FRACTURE LINES WITH NO CONSOLIDATION. AT 12 WEEKS ((B)(6) 2025), ONE SCREW ON THE MEDIAL PLATE WAS FOUND BROKEN, WITH LOOSENING OF TWO ADDITIONAL SCREWS MEDIALLY AND ONE LATERALLY. NEW BONE TISSUE HAS FORMED ON THE OUTER SURFACE OF THE SHAFT (DIAPHYSIS), WITHIN THE PERIOSTEUM, AND MEDIALLY IN THE PERIARTICULAR REGION. HOWEVER, THE FRACTURE GAPS IN THE NEAR THE JOINT (INTRA-ARTICULAR) AND AT THE END OF THE BONE (METAPHYSEAL REGION) REMAINED CLEARLY VISIBLE, INDICATING DELAYED BONE UNION. THE SURGEON ATTRIBUTED THE COMPLICATION TO MICROMOVEMENTS AT THE FRAC

    Patient problems: Bone Fracture(s)

    MAUDE report 22565402 ↗

Other Plate, Fixation, Bone devices

All Plate, Fixation, Bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page