Skip to content

SALGINA 2.5mm Volar Distal Radius Plate, Standard, 3 Hole, Left B-111200Recalls & reported problems

Bonebridge AG · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07640187462630), received between Jan 13, 2026 and Jan 13, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Bone Fracture(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 13, 2026Detachment of Device or Device Component

    ON 16 DECEMBER, BONEBRIDGE RECEIVED MARKET FEEDBACK FROM (B)(6) SWITZERLAND REPORTING AN OSTEOSYNTHESIS MATERIAL REMOVAL (OSME) AND REVISION OSTEOSYNTHESIS PERFORMED AFTER THE 6-WEEK FOLLOW-UP DUE TO SECONDARY DISLOCATION OF THE DISTAL LOCKING SCREWS OF A SALGINA DISTAL RADIUS PLATE. THE TREATING SURGEON DESCRIBED THE PATIENT AS BEING VERY ACTIVE AND STATED THAT POSTOPERATIVE NON-COMPLIANCE WAS A HIGHLY LIKELY CONTRIBUTING FACTOR. ACCORDING TO THE SURGEON EARLY EXCESSIVE LOADING OF THE OPERATED HAND, INCLUDING SUPPORTING BODY WEIGHT SHORTLY AFTER SURGERY, WAS CONSIDERED A LIKELY CONTRIBUTING FACTOR. ACCORDING TO THE OPERATIVE REPORT, ALL DISTAL LOCKING SCREWS WERE INSERTED USING A TORQUE-LIMITING DEVICE WITH POWER ASSISTANCE AND WERE MANUALLY CHECKED AND RETIGHTENED AT THE END OF THE PROCEDURE. IMAGING & CLINICAL FINDINGS: POSTOPERATIVE IMAGING CONFIRMED THAT ALL DISTAL SCREWS WERE POSITIONED EXTRA-ARTICULARLY. AT THE 6-WEEK FOLLOW-UP, CT IMAGING SHOWED SECONDARY DISPLACEMENT OF THE DISTAL FRAGMENT BLOCK, WITH THE SCREWS REMAINING FIXED IN BONE WHILE THE FRAGMENT BLOCK HAD SUBSIDED RELATIVE TO THE PLATE. THIS FINDING INDICATES A LOSS OF LOCKING INTEGRITY AT THE SCREW-PLATE INTERFAC

    Patient problems: Bone Fracture(s)

    MAUDE report 24052100 ↗

Other Plate, Fixation, Bone devices

All Plate, Fixation, Bone devices →

Sources and updates