EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAA (UDI 07898959543333)Recalls & reported problems
Micromar Indústria e Comércio Ltda · Class 2
All models of EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is the Micromar EasyDrill cranial perforator, a powered surgical drill bit used to make burr holes in the skull during craniotomy. Reports mostly describe the drill bit failing to stop automatically after passing through the skull, sometimes driving deeper than intended and damaging the membrane covering the brain, with bleeding and perforation also mentioned. There are 31 reports, 16 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 07898959543333), received between Dec 10, 2019 and Apr 5, 2022 — classified as 16 injuries, 15 malfunctions.
The most frequently reported specific device problem is “Device Remains Activated” (29 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Remains Activated29
- Failure to Auto Stop1
- Failure to Shut Off1
Patient problems
- Perforation15
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient4
- Hemorrhage/Blood Loss/Bleeding3
- Injury3
- Contusion1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2022Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT DURING CRANIOTOMY, WHILE CREATING THE SECOND BURR HOLE IN THE SKULL, THE PERFORATOR DID NOT AUTOMATICALLY STOP ONCE IT MADE IT THROUGH THE SKULL. ADDITIONAL INFORMATION REGARDING THE PATIENT IMPACT WAS BEING FOLLOWED UP FROM THE FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14014184 ↗MalfunctionFeb 17, 2022Failure to Shut Off
WHILE CREATING THE SECOND BURR HOLE IN THE SKULL, THE PERFORATOR DID NOT AUTOMATICALLY STOP ONCE IT MADE IT THROUGH THE SKULL.
Patient problems: Insufficient Information
MAUDE report 13548421 ↗MalfunctionOct 11, 2021Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT DURING MEDTRONIC LAB TESTING OF THE DEVICE, IT WAS NOTED THAT THE CLUTCH FAILED AND DIDN'T STOP AT A CERTAIN DEPTH ON THE BOVINE SCAPULA. IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12611588 ↗MalfunctionJun 23, 2021Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT THE BIT DID NOT STOP AFTER GOING THROUGH THE CRANIUM. IT WAS NOTED THAT THE PATIENT IS ALIVE BUT WITH INJURY. ON FOLLOW UP, IT WAS CONFIRMED THAT THERE WAS NO PATIENT INJURY/IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12047914 ↗MalfunctionJun 22, 2021Device Remains Activated; Device Remains Activated
IT WAS REPORTED THROUGH MAUDE REFERENCE NUMBER (B)(4) THAT THE CRANIAL PERFORATOR DID NOT STOP AT THE INNER TABLE AND IT WAS MANUALLY STOPPED BY THE SURGEON. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. NO FURTHER INFORMATION REGARDING THE FACILITY AND THE EVENT WAS PROVIDED IN THE MAUDE REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12046589 ↗MalfunctionJun 22, 2021Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT THE DEVICE WENT TOO FAR AND DIDN'T STOP AS IT SHOULD HAVE. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12040309 ↗MalfunctionApr 20, 2021Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT DURING BUR HOLE CREATION, THE PERFORATOR DID NOT STOP. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT IMPACT AND THERE WAS NO PROCEDURE DELAY. ON FOLLOW-UP, IT WAS REPORTED THAT THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11692715 ↗InjuryApr 13, 2021Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT DURING A CRANIOTOMY PROCEDURE, THE PERFORATOR DID NOT STOP AND DAMAGED THE DURA. IT WAS ALSO CONFIRMED UPON COMPLETION OF THE PROCEDURE, THAT THE BRAIN SURFACE WAS NOT INJURED. ON FOLLOW-UP, IT WAS REPORTED THAT THERE WAS NO PROCEDURE DELAY, THERE WAS NO BACK-UP DEVICE USED, AND THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY.
Patient problems: Perforation
MAUDE report 11660760 ↗InjuryMar 5, 2021Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT THE PERFORATOR PLUNGED ON A SUBDURAL. ON FOLLOW UP, IT WAS REPORTED THAT THE DEVICE DID NOT STOP DURING A SUBDURAL HEMATOMA EVACUATION. IT WAS ALSO REPORTED THAT THERE WERE NO ADDITIONAL INTERVENTIONS PERFORMED, THERE WAS NO PROCEDURE DELAY, AND THE PROCEDURE WAS COMPLETED USING THE REPORTED DEVICE. ON FURTHER FOLLOW-UP, IT WAS REPORTED THAT THERE WAS NO FURTHER INFORMATION REGARDING THE STATUS OF THE PATIENT POST-SURGERY AND THE DEVICE IS STILL IN THE HOSPITAL'S RISK MANAGEMENT.
Patient problems: Perforation
MAUDE report 11425628 ↗InjuryFeb 12, 2021Device Remains Activated; Device Remains Activated
ON FOLLOW-UP, IT WAS REPORTED THAT THERE WAS NO FURTHER INFORMATION THAT CAN BE PROVIDED REGARDING THE EVENT AND THE STATUS OF THE PATIENT POST-SURGERY.
Patient problems: Perforation
MAUDE report 11325139 ↗InjuryJan 20, 2021Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT THE PERFORATOR DID NOT STOP AND CUT THE DURA. IT WAS ALSO REPORTED THAT A PIECE OF BONE LODGED INTO THE BRAIN. ON FOLLOW-UP, IT WAS REPORTED THAT THE EVENT HAPPENED DURING A DEEP BRAIN STIMULATION PROCEDURE AND THERE WAS A 20-MINUTE DELAY IN THE PROCEDURE TO REMOVE THE BONE FROM THE BRAIN. IT WAS ALSO REPORTED THAT THE PATIENT NEED TO REMAIN FLAT FOR 24 HOURS. ON FURTHER FOLLOW-UP, IT WAS REPORTED THAT THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY AND THE DEVICE WAS DISCARDED.
Patient problems: Perforation
MAUDE report 11202105 ↗InjuryDec 2, 2020Device Remains Activated; Device Remains Activated
ON FOLLOW UP, IT WAS REPORTED THAT THE PERFORATOR PLUNGED INTO THE TUMOR THAT WAS SUPERFICIAL. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS CONTINUED NORMALLY WITHOUT DELAY AND THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY. THE DEVICE WILL NOT BE RETURNED AS IT WAS DISCARDED BY THE CUSTOMER.
Patient problems: Perforation
MAUDE report 10931584 ↗InjuryNov 18, 2020Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT DURING A PROCEDURE, THE PERFORATOR DID NOT STOP WHICH CAUSED BLEEDING AND DAMAGE TO THE PATIENT'S DURA. ON FOLLOW UP, IT WAS REPORTED THAT THE EVENT HAPPENED DURING A VP SHUNT PLACEMENT AND THERE WAS A DELAY ON THE PROCEDURE TO PREVENT BLOOD LOSS. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED USING THE REPORTED DEVICES AND THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Perforation
MAUDE report 10864128 ↗MalfunctionNov 18, 2020Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE PERFORATOR. NO PATIENT IMPACT REPORTED. ON FOLLOW UP, IT WAS REPORTED THAT THE PERFORATOR DID NOT STOP. IT WAS ALSO REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT AND THE DEVICE HAS BEEN DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10863787 ↗MalfunctionNov 17, 2020Device Remains Activated; Device Remains Activated
ON FOLLOW UP, IT WAS CONFIRMED THAT THE PROCEDURE WAS COMPLETED USING THE REPORTED DEVICE AND THERE WAS NO ISSUE WITH THE PATIENT PO ST-SURGERY. IT WAS ALSO REPORTED THAT THE DEVICE WAS DISCARDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10856925 ↗MalfunctionOct 30, 2020Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT DURING A REVISION OF LEFT SHUNT VALVE AND TUBING, THE CRANIAL PERFORATOR CONTINUES TO RUN, AND WAS MANUALLY STOPPED BY THE SURGEON. ADDITIONAL INFORMATION ABOUT THE EVENT IS BEING FOLLOWED UP FROM THE FACILITY CONTACT PERSON.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10762320 ↗InjuryOct 5, 2020Device Remains Activated; Device Remains Activated
ON FOLLOW-UP, IT WAS CONFIRMED THAT THERE WERE NO ADDITIONAL ACTIONS REQUIRED AND THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY.
Patient problems: Perforation
MAUDE report 10631924 ↗MalfunctionOct 5, 2020Device Remains Activated; Device Remains Activated
ON FOLLOW UP, IT WAS REPORTED THAT THE EVENT HAPPENED DURING A CRANIOTOMY PROCEDURE AND THERE WAS NO DELAY OR ADDITIONAL INTERVENTIONS PERFORMED. IT WAS ALSO REPORTED THAT THE DEVICE DID NOT HIT THE PATIENT'S DURA AND NO FURTHER INFORMATION CAN BE PROVIDED REGARDING THE STATUS OF THE PATIENT POST-SURGERY.
Patient problems: No Consequences Or Impact To Patient; Perforation
MAUDE report 10631844 ↗InjuryOct 5, 2020Device Remains Activated; Device Remains Activated
ON FOLLOW-UP, IT WAS REPORTED THAT THE EVENT HAPPENED DURING A CRANIOTOMY PROCEDURE AND THERE WAS NO DELAY OR ADDITIONAL INTERVENTIONS REQUIRED. IT WAS ALSO CONFIRMED THAT THERE WAS NO PATIENT IMPACT AND THERE WAS NO FURTHER INFORMATION REGARDING THE STATUS OF THE PATIENT POST-SURGERY.
Patient problems: No Consequences Or Impact To Patient; Perforation
MAUDE report 10628824 ↗MalfunctionSep 11, 2020Device Remains Activated; Device Remains Activated
ON FOLLOW-UP, IT WAS REPORTED THAT THE 10-MINUTE DELAY WAS TO PREPARE ANOTHER PERFORATOR TO COMPLETE THE PROCEDURE. IT WAS ALSO MENTIONED THAT THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY. ALSO, THERE WAS NO FURTHER INFORMATION THAT CAN BE PROVIDED REGARDING THE DETAILS OF THE EM200.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10521426 ↗InjuryAug 27, 2020Device Remains Activated; Device Remains Activated
ON FOLLOW UP, IT WAS CONFIRMED THAT THE DURA WAS NOT HIT BECAUSE THE BRAIN WAS SUNKEN DUE TO THE SUBDURAL HEMATOMA. NO FURTHER INFORMATION CAN BE PROVIDED FOR THE AD03 DRIVER DETAILS.
Patient problems: Injury; No Consequences Or Impact To Patient
MAUDE report 10462877 ↗InjuryAug 6, 2020Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT DURING A CRANIOTOMY PROCEDURE FOR A TUMOR RESECTION, THE PERFORATOR DID NOT STOP SPINNING WHILE CREATING THE INITIAL BUR HOLE. IT HAPPENED WHILE PASSING THROUGH THE SKULL THUS PLUNGING INTO THE TISSUE. IT WAS ALSO ADDED THAT THERE WAS A PROCEDURE DELAY TO CONTROL THE BLEEDING. THE REMAINING BUR HOLES WERE COMPLETED USING AN ACRA-CUT PERFORATOR. ON FOLLOW UP, NO FURTHER INFORMATION CAN BE PROVIDED REGARDING THE STATUS OF THE PATIENT POST-SURGERY.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Perforation
MAUDE report 10376638 ↗InjuryAug 4, 2020Device Remains Activated; Device Remains Activated
IT WAS REPORTED THROUGH MAUDE REFERENCE NUMBER (B)(4) THAT THE PERFORATOR FAILED TO STOP AT DURA LEVEL WHICH RESULTED TO DURA CONTUSION. IT WAS ALSO ADDED THAT THE PATIENT HAD NO FURTHER ISSUES POST-SURGERY. NO FURTHER INFORMATION REGARDING THE FACILITY AND THE EVENT WAS PROVIDED IN THE MAUDE REPORT.
Patient problems: Contusion; Perforation
MAUDE report 10364869 ↗MalfunctionJun 24, 2020Failure to Auto Stop
THE CRANIAL PERFORATOR DIDN'T STOP AT THE INNER TABLE; IT WAS MANUALLY STOPPED BY THE SURGEONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10188821 ↗InjuryJun 17, 2020Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT THE PERFORATOR MALFUNCTIONED. NO PATIENT IMPACT REPORTED. ON FOLLOW UP, IT WAS CONFIRMED THAT THE CLUTCH OF THE PERFORATOR DID NOT WORK CAUSING DURA DAMAGE.
Patient problems: Perforation
MAUDE report 10165420 ↗
Other Drills, Burrs, Trephines & Accessories (Compound, Powered) devices
- CODMAN® 261221Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall223 reports
- Zyphr 5100060001Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall199 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)67 reports
- CODMAN® 261222Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall7 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)54 reports
- CODMAN® 261230Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall3 reports
- CODMAN® 261223Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall2 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)37 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page