Skip to content

EasyDrill Cranial Perforator Ø14 x Ø11 x 1,5mm DM0210FAA (UDI 07898959543609)Recalls & reported problems

Micromar Indústria e Comércio Ltda · Class 2

All models of EasyDrill Cranial Perforator Ø14 x Ø11 x 1,5mm (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07898959543609), received between Nov 4, 2024 and Dec 22, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Device Remains Activated” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Remains Activated6
  • Fracture3
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Perforation2
  • Insufficient Information1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 22, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT IN OPEN CRANIOTOMY PROCEDURE TO REMOVE BRAIN TUMOR THE BIT DID NOT STOP AUTOMATICALLY AND TORE THE PATIENT'S DURA MATER.THE PROCEDURE WAS EXTENDED FOR 5 MINUTES AND IT WAS COMPLETED BY THE BACKUP PRODUCTS.

    Patient problems: Perforation

    MAUDE report 23861727 ↗
  • MalfunctionDec 19, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A BURR HOLE PROCEDURE, THREE PERFORATOR BROKE WHILE DRILLING. IT WAS REPORTED THAT PROCEDURE WAS DELAYED BY 20 MINUTES AND PROCEDURE WAS COMPLETED USING A MANUAL DRILL INSTEAD OF THE INTENDED POWERED DEVICE. AT THE TIME OF REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT. ADDITIONAL DETAILS REGARDING PATIENT IMPACT IS BEING FOLLOWED FROM THE FACILITY. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23847016 ↗
  • MalfunctionDec 19, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A BURR HOLE PROCEDURE, THREE PERFORATOR BROKE WHILE DRILLING. IT WAS REPORTED THAT PROCEDURE WAS DELAYED BY 20 MINUTES AND PROCEDURE WAS COMPLETED USING A MANUAL DRILL INSTEAD OF THE INTENDED POWERED DEVICE. AT THE TIME OF REPORT, THERE WAS NO KNOWN IMPACT TO A PATIENT. ADDITIONAL DETAILS REGARDING PATIENT IMPACT IS BEING FOLLOWED FROM THE FACILITY. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23847014 ↗
  • MalfunctionOct 22, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING A CRANIOTOMY PROCEDURE, THE PERFORATOR WAS USED WITH THE MIDAS REX MOTOR TO DRILL INTO THE CRANIAL BONE. THE SURGEON OBSERVED THAT THE BURR DID NOT STOP ROTATING AFTER COMPLETING THE BONE PERFORATION.  IT WAS ALSO REPORTED THAT NO INTERVENTION WAS PERFORMED AND THE PROCEDURE WAS COMPLETED USING THE SAME PRODUCT. IT WAS ALSO REPORTED THAT THERE WAS NO PROCEDURE DELAY AND NO PATIENT INJURY. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WAS NO PATIENT OR STAFF IMPACT AND THE RIGHT PARIETAL BONE OF THE SKULL WAS THE PERFORATED. IT WAS ALSO STATED THAT THE PATIENT HAVE RELEVANT NO BONE DISEASE OR COMORBIDITIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356587 ↗
  • MalfunctionMar 12, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING A STANDARD TUMOR RESECTION PROCEDURE, FAILURE OF EASY DRILL PERFORATOR. IT WAS REPORTED THAT THE PATIENT WAS ALIVE WITH NO INJURY. IT WAS ALSO REPORTED THAT THERE WAS NO INTERVENTION PLANNED. ON FOLLOW-UP IT WAS REPORTED THAT THERE WAS NO PATIENT OR STAFF IMPACT AND THERE WAS NO DELAY IN PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21583858 ↗
  • MalfunctionMar 4, 2025Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A PROCEDURE TO PERFORATE THE SKULL TO RESECT THE TUMOR THE PRODUCT GOT BROKEN AND THE UPPER PART FELL OFF FROM THE DEVICE. IT WAS ALSO REPORTED THAT THERE WAS NO INTERVENTION PERFORMED AND THERE WAS NO DELAY IN PROCEDURE. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. ADDITIONAL INFORMATION REGARDING THE PATIENT IMPACT WAS BEING FOLLOWED UP FROM THE FACILITY. ON ADDITIONAL FOLLOW THE PATIENT SHOWED NO SIGNS OF BONE DISEASES. THE SURGERY WAS PERFORMED DUE TO A HEMATOMA IN THE PARIETAL BONE AREA. TWO HOLES WERE CREATED USING A PERFORATOR, BUT THE DEVICE ENCOUNTERED AN ISSUE WITH THE FIRST HOLE. THE EM200 MOTOR WAS USED DURING THE PROCEDURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21515466 ↗
  • MalfunctionDec 3, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING SHUNT INSERTION UNDER LAPAROSCOPY NAVIGATION, THE PERFORATOR DID NOT STOP WHILE DRILLING THROUGH THE SKULL BONE. IT WAS REPORTED THAT THE PATIENT WAS ALIVE - NO INJURY AND THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT. UPON FOLLOWUP IT WAS CONFIRMED THAT THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20835852 ↗
  • MalfunctionDec 1, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT IN THE PROCEDURE OF CRANIOTOMY THE DRILL DOESN'T MAKE CORRECT TURNS . IT STOPS FOR A WHILE THEN TURNS BACK ON. IT WAS REPORTED THAT INTERVENTION HAS BEEN PERFORMED. THE PROCEDURE WAS COMPLETED WITH THE BACKUP PRODUCT. UPON FOLLOWUP IT WAS CONFIRMED THAT THERE WAS NO PATIENT IMPACT. THERE WAS A DELAY FOR 30 MINUTES DURING THE PROCEDURE. NO INTERVENTION PERFORMED OR PLANNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20814364 ↗
  • InjuryNov 4, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT WHEN DRILLING BURR HOLE, THE PERFORATOR DID NOT STOP AUTOMATICALLY WHEN FINISHED DRILLING BONE CONTINUING TO DRILL AND TEAR DURA MATER. IT WAS ALSO REPORTED THAT THERE WAS A PROCEDURE DELAY OF 30 MINUTES AND DURAL REPAIR WAS REQUIRED. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT.

    Patient problems: Perforation

    MAUDE report 20597442 ↗

Other Drills, Burrs, Trephines & Accessories (Compound, Powered) devices

All Drills, Burrs, Trephines & Accessories (Compound, Powered) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page