LILLIPUT D902 PH.I.S.I.O.Recalls & reported problems
Sorin Group Italia Srl · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08033178100397), received between Nov 8, 2017 and May 2, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak3
- Air Leak1
- Blocked Connection1
- Disconnection1
- Medical Gas Supply Problem1
- Unexpected Therapeutic Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Coagulation Disorder1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 2, 2024Unexpected Therapeutic Results; Unexpected Therapeutic Results
SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING A PEDIATRIC CARDIAC SURGERY, AFTER FEW MINUTES OF CARDIAC BYPASS INITIATION, IT WAS FOUND THAT THE P02 LEVELS BLOOD GAS WERE 90MMHG WITH FI02 100%. MEDICAL TEAM REPEAT SAMPLE AND SHOWED 99MMHG. MEDICAL TEAM EVALUATED THE VALUE NOT ACCEPTABLE AND ELECTED TO CHANGE OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19228063 ↗MalfunctionMay 25, 2023Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT OF A LEAKAGE AT THE LILLIPUT OXYGENATOR ARTERIAL OUTLET DURING A PROCEDURE. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17002167 ↗MalfunctionJan 17, 2022Fluid/Blood Leak; Fluid/Blood Leak
SORIN GROUP (B)(4) RECEIVED A REPORT THAT, 4 LILLIPUT OXYGENATORS LEAKED FROM OXY SITE WHEN CONNECTED TO THE HEATER COOLER (1 UNIT LOT 2002260217 AND 3 UNITS LOT 2002260216). DURING FOLLOW UP WITH THE CUSTOMER, SORIN GROUP (B)(4) HAS RECEIVED THAT DID NOT CLARIFY IF WATER ENTERED THE RESERVOIR WHEN PRIMING THE WATER SIDE OF LILLIPUT OXYGENATORS. THE EVENTS OCCURRED PRIOR TO ANY PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13266472 ↗MalfunctionJan 17, 2022Fluid/Blood Leak; Fluid/Blood Leak
SORIN GROUP (B)(4) RECEIVED A REPORT THAT, 4 LILLIPUT OXYGENATORS LEAKED FROM OXY SITE WHEN CONNECTED TO THE HEATER COOLER (1 UNIT LOT 2002260217 AND 3 UNITS LOT 2002260216). DURING FOLLOW UP WITH THE CUSTOMER, SORIN GROUP (B)(4) HAS RECEIVED THAT DID NOT CLARIFY IF WATER ENTERED THE RESERVOIR WHEN PRIMING THE WATER SIDE OF LILLIPUT OXYGENATORS. THE EVENTS OCCURRED PRIOR TO ANY PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13266465 ↗MalfunctionNov 24, 2021Blocked Connection; Blocked Connection
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12867993 ↗MalfunctionSep 15, 2021Medical Gas Supply Problem; Medical Gas Supply Problem
SORIN GROUP (B)(4) HAS RECEIVED A REPORT THAT, DURING A PROCEDURE, THE D902 DIDECO LILLIPUT PHISIO OXYGENATOR WAS NOT OXYGENATING AS EXPECTED. MEDICAL TEAM ELECTED TO CHANGE-OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12475310 ↗MalfunctionNov 8, 2017Air Leak; Air Leak
SORIN GROUP (B)(6) RECEIVED A REPORT THAT AN ESCAPE OF AIR WAS OBSERVED FROM THE D902 LILLIPUT OXYGENATOR DURING A PROCEDURE AND A LARGE AMOUNT OF FOAM WAS NOTICED. THE AIR REPORTEDLY ESCAPED THROUGH THE UPPER RIGHT SIDE OF THE RESERVOIR FILTER. THE PATIENT REQUIRED TRANSFUSION THERAPY OF ALL FACTORS. THE PATIENT REPORTEDLY SUFFERED FROM CONSUMPTION COAGULOPATHY.
Patient problems: Coagulation Disorder
MAUDE report 7008906 ↗
FDA clearance and classification
The FDA regulates product code DTZ (oxygenator, cardiopulmonary bypass) under 21 CFR 870.4350 as a Class II device. 21 CFR 870.4350 ↗
Devices of this type normally need 510(k) clearance before sale.
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the LILLIPUT D902 PH.I.S.I.O.
Has the LILLIPUT D902 PH.I.S.I.O. been recalled?
No FDA recall record in our data names this exact device identifier (DI 08033178100397), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the LILLIPUT D902 PH.I.S.I.O.?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Nov 8, 2017 and May 2, 2024. In those reports, the most common specific device problems are “Fluid/Blood Leak” (3), “Air Leak” (1) and “Blocked Connection” (1); the most common patient problem is “Coagulation Disorder” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the LILLIPUT D902 PH.I.S.I.O.?
None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the LILLIPUT D902 PH.I.S.I.O., and is it still sold?
The FDA's device database lists Sorin Group Italia Srl as the company responsible for it (brand LILLIPUT, model D902 PH.I.S.I.O.). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other OXYGENATOR, CARDIOPULMONARY BYPASS devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths87 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-i HMO 50000-USA #SQUADROX-i Small Adult, no FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page