INSPIRE 8 PH.I.S.I.O.Recalls & reported problems

Sorin Group Italia Srl · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08033178112369), received between Jan 31, 2017 and Apr 9, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 13, 2018Class IITerminated

    Some integrated sterile INSPIRE family products contain an incorrect expiration date printed in the Unique Device Identification (UDI) bar code and readable code just beneath the barcode. Specifically, the expiration date is reported as DDMMYY instead of YYMMDD. The expiration date in the applicable field on the package labels are correct as printed.

    Z-1919-2018 · Root cause: Incorrect or no expiration date · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Increase in Pressure2
  • Coagulation in Device or Device Ingredient1
  • Device Operational Issue1
  • Low Readings1
  • Pressure Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    SORIN GROUP ITALY RECEIVED A REPORT RELATED TO A POSSIBLE BLOOD TO WATER LEAK OF AN INSPIRE 8 OXYGENATOR. THE ISSUE OCCURRED TOWARD THE END OF THE PROCEDURE, WHEN IT WAS NOTICED LIGHT PINK/RED TINGE IN THE WATER LINES. MEDICAL TEAM CHOSE TO FINISH THE SURGERY AS IT WAS VERY LIGHT TINGED IN COLOR AND THE CASE WAS ALMOST COMPLETE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19070311 ↗
  • MalfunctionMar 22, 2024Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient

    SORIN GROUP ITALY RECEIVED A REPORT OF A SUDDEN BLOOD ACTIVATION INSIDE INSPIRE 8 OXYGENATOR DURING BYPASS ON A LONG CASE. MEDICAL TEAM ELECTED TO CHANGE OUT THE AFFECTED UNIT. NO FURTHER DETAILS ABOUT SURGICAL PHASE (BEATING HEART SURGERY OR AORTIC CROSS-CLAMP) AND DURATION OF THE CHANGE OUT OPERATION WERE COMMUNICATED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18957694 ↗
  • MalfunctionAug 2, 2021Increase in Pressure; Increase in Pressure

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12257992 ↗
  • MalfunctionAug 11, 2017Low Readings; Device Operational Issue; Low Readings; Device Operational Issue

    SORIN GROUP (B)(4) RECEIVED A REPORT THAT LOW OXYGEN LEVELS WERE FOUND BY THE PERFUSIONIST FOR THE INSPIRE 8 HOLLOW FIBER OXYGENATOR THEN THE BLOOD GAS PARAMETER WAS ANALYZED UPON THE START OF BYPASS. THE PERFUSIONIST DECIDED TO SWITCH TO A NEW MEMBRANE OXYGENATOR TO CONTINUE. THE CHANGE OUT WAS DONE AFTER COOLING DOWN OF THE PATIENT. OXYGEN LEVELS WERE RESTORED AFTER THE CHANGE OUT. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6787410 ↗
  • MalfunctionFeb 17, 2017Increase in Pressure; Pressure Problem; Increase in Pressure; Pressure Problem

    SORIN GROUP (B)(4) RECEIVED A REPORT THAT HIGH TRANSMEMBRANE PRESSURE WAS OBSERVED FOR THE INSPIRE 8 HOLLOW FIBER OXYGENATOR AT THE BEGINNING OF A PROCEDURE. THE MEDICAL TEAM ELECTED TO CONTINUE THE CASE. DURING THE COOLING PHASE, THE OXYGENATOR TRANSMEMBRANE PRESSURE INCREASED AND THE MEDICAL TEAM ELECTED REPLACE THE DEVICE AT THIS TIME. THE WHOLE PROCEDURE WAS PROLONGED OF 40 MINUTES. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6338721 ↗
  • MalfunctionJan 31, 2017Fluid/Blood Leak; Fluid/Blood Leak

    SORIN GROUP (B)(4) RECEIVED A REPORT THAT TRACES OF BLOOD WERE OBSERVED IN THE WATER OUTLET OF THE INSPIRE 8 DUAL HOLLOW FIBER OXYGENATOR ABOUT 50 MINUTES AFTER STARTING BYPASS. THE MEDICAL TEAM OPTED TO CONTINUE THE SURGERY WITHOUT CHANGING THE OXYGENATOR. THE PROCEDURE WAS COMPLETED WITHOUT ISSUE. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6289851 ↗

Other OXYGENATOR, CARDIOPULMONARY BYPASS devices

All OXYGENATOR, CARDIOPULMONARY BYPASS devices →