INSPIRE 6F PH.I.S.I.O.Recalls & reported problems
Sorin Group Italia Srl · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 08033178112376), received between Aug 3, 2018 and Feb 20, 2026 — classified as 1 injury, 13 malfunctions.
The most frequently reported specific device problem is “Unexpected Therapeutic Results” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unexpected Therapeutic Results6
- Fluid/Blood Leak5
- Device Contamination with Chemical or Other Material1
- Increase in Pressure1
- Particulates1
- Restricted Flow rate1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2026Unexpected Therapeutic Results
SORIN GROUP ITALIA RECEIVED A REPORT OF A POOR GAS EXCHANGE PERFORMANCE RELATED TO TWO UNITS OF INSPIRE 6F OXYGENATOR. THERE WAS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24401203 ↗MalfunctionJan 23, 2026Restricted Flow rate
SORIN GROUP ITALIA RECEIVED A REPORT THAT, AT THE BEGINNING OF THE PROCEDURE, THE RESERVOIR OF AN INSPIRE 6F OXYGENATOR WAS BLOCKED AND IT WAS DECIDED TO CHANGE THE WHOLE CARDIOPLEGIA CIRCUIT. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24162695 ↗MalfunctionMay 9, 2025Fluid/Blood Leak; Fluid/Blood Leak
SORIN GROUP ITALY RECEIVED A REPORT THAT AN INSPIRE 6F OXYGENATOR HAD A SIGNIFICANT LEAKAGE FROM THE ARTERIAL LINE OUTLET DURING PROCEDURE. THE ARTERIAL LINE WAS ADJUSTED, BUT SINCE THE LEAKAGE CONTINUED, MEDICAL TEAM ELECTED TO REMOVE THE PATIENT FROM THE PUMP AND REPLACE THE OXYGENATOR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21994647 ↗MalfunctionMay 7, 2025Fluid/Blood Leak; Fluid/Blood Leak
SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, AT THE TIME OF STARTING THE VALVE REPLACEMENT SURGERY, A LEAK IN THE LUER CONNECTION OF THE ARTERIAL LINE OF THE INSPIRE OXYGENATOR WAS OBSERVED. MEDICAL TEAM ELECTED TO CHANGE OUT THE OXYGENATOR. THERE WAS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21974700 ↗MalfunctionJan 8, 2025Unexpected Therapeutic Results; Unexpected Therapeutic Results
SEE INITIAL REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21102481 ↗InjuryApr 9, 2024Unexpected Therapeutic Results; Unexpected Therapeutic Results
SORIN GROUP ITALIA RECEIVED A REPOR OF AN INSPIRE OXYGENATOR FAILURE DURING A PROCEDURE, THE FIO2 WAS 95% AND PO2 SHOWS ~20%. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19071303 ↗MalfunctionAug 2, 2023Increase in Pressure; Unexpected Therapeutic Results; Increase in Pressure; Unexpected Therapeutic Results
SORIN GROUP ITALIA RECEIVED A REPORT THAT HIGH TRANSMEMBRANE PRESSURE OCCURRED DURING A PROCEDURE. MEDICAL TEAM ELECTED TO ADMINISTER MAG 2GM AND ALBUMIN 12.5GM. SURGERY WAS COMPLETED WITHOUT ANY NEED FOR UNIT CHANGE-OUT. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17442871 ↗MalfunctionJun 7, 2023Fluid/Blood Leak; Fluid/Blood Leak
SORIN GROUP ITALIA RECEIVED A REPORT THAT, AFTER ENTERING INTO BYPASS, A BLOOD LEAK WAS OBSERVED FORM THE TEMPERATURE PROBE HOLDER OF THE INSPIRE 6F OXYGENATOR. MEDICAL TEAM ELECTED TO CHANGE OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17077048 ↗MalfunctionOct 26, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING SET UP OF THE HEART LUNG MACHINE, THE PERFUSIONIST NOTICED A PIECE OF FOREIGN MATTER IN THE INSPIRE 6F OXYGENATOR ARTERIAL OUTLET. THE OXYGENATOR WAS NOT USED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15672528 ↗MalfunctionSep 6, 2022Fluid/Blood Leak; Fluid/Blood Leak
SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING A MITRAL VALVE REPLACEMENT SURGERY, WHILE ON PUMP, BLOOD LEAK FROM THE PURGE LINE OF THE INSPIRE 6F OXYGENATOR WAS OBSERVED. MEDICAL TEAM ELECTED TO CHANGE-OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15359897 ↗MalfunctionAug 5, 2022Unexpected Therapeutic Results; Unexpected Therapeutic Results
SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT OF ARTERIAL BLOOD DESATURATION HALF AN HOUR AFTER THE BEGINNING OF THE CEC USING AN INSPIRE 6F OXYGENATOR. MEDICAL TEAM ELECTED TO CHANGE-OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15173416 ↗MalfunctionApr 15, 2022Unexpected Therapeutic Results; Unexpected Therapeutic Results
SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING A PROCEDURE USING AN INSPIRE 6F OXYGENATOR, THE OXYGEN VALUES IN THE PATIENT BLOOD WERE LOW. THERE IS NO REPORT OF ANY PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14117000 ↗MalfunctionMar 21, 2019Particulates; Particulates
SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING DEVICE SET-UP, A BLACK FOREIGN PARTICLE WAS IDENTIFIED BY THE OUTLET SIDE OF INSPIRE 6F INTEGRATED ARTERIAL FILTER. THE ISSUE WAS IDENTIFIED PRIOR TO ANY PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8440408 ↗MalfunctionAug 3, 2018Fluid/Blood Leak; Fluid/Blood Leak
SORIN (B)(4) RECEIVED A REPORT OF A BLOOD LEAK FROM THE LOWER PORTION OF AN INSPIRE 6F HOLLOW FIBER OXYGENATOR DURING A PROCEDURE. ACCORDING TO THE CUSTOMER, THE TOTAL BLOOD LOSS WAS OF 1000 ML AND THE PATIENT WAS HEMO-TRANSFUSED. THERE WAS NO REPORT OF PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7746125 ↗
Other OXYGENATOR, CARDIOPULMONARY BYPASS devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths90 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-i HMO 50000-USA #SQUADROX-i Small Adult, no FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page