INSPIRE 6F PH.I.S.I.O.Recalls & reported problems

Sorin Group Italia Srl · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 08033178112376), received between Aug 3, 2018 and Feb 20, 2026 — classified as 1 injury, 13 malfunctions.

The most frequently reported specific device problem is “Unexpected Therapeutic Results” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Therapeutic Results6
  • Fluid/Blood Leak5
  • Device Contamination with Chemical or Other Material1
  • Increase in Pressure1
  • Particulates1
  • Restricted Flow rate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2026Unexpected Therapeutic Results

    SORIN GROUP ITALIA RECEIVED A REPORT OF A POOR GAS EXCHANGE PERFORMANCE RELATED TO TWO UNITS OF INSPIRE 6F OXYGENATOR. THERE WAS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24401203 ↗
  • MalfunctionJan 23, 2026Restricted Flow rate

    SORIN GROUP ITALIA RECEIVED A REPORT THAT, AT THE BEGINNING OF THE PROCEDURE, THE RESERVOIR OF AN INSPIRE 6F OXYGENATOR WAS BLOCKED AND IT WAS DECIDED TO CHANGE THE WHOLE CARDIOPLEGIA CIRCUIT. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24162695 ↗
  • MalfunctionMay 9, 2025Fluid/Blood Leak; Fluid/Blood Leak

    SORIN GROUP ITALY RECEIVED A REPORT THAT AN INSPIRE 6F OXYGENATOR HAD A SIGNIFICANT LEAKAGE FROM THE ARTERIAL LINE OUTLET DURING PROCEDURE. THE ARTERIAL LINE WAS ADJUSTED, BUT SINCE THE LEAKAGE CONTINUED, MEDICAL TEAM ELECTED TO REMOVE THE PATIENT FROM THE PUMP AND REPLACE THE OXYGENATOR. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21994647 ↗
  • MalfunctionMay 7, 2025Fluid/Blood Leak; Fluid/Blood Leak

    SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, AT THE TIME OF STARTING THE VALVE REPLACEMENT SURGERY, A LEAK IN THE LUER CONNECTION OF THE ARTERIAL LINE OF THE INSPIRE OXYGENATOR WAS OBSERVED. MEDICAL TEAM ELECTED TO CHANGE OUT THE OXYGENATOR. THERE WAS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21974700 ↗
  • MalfunctionJan 8, 2025Unexpected Therapeutic Results; Unexpected Therapeutic Results

    SEE INITIAL REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21102481 ↗
  • InjuryApr 9, 2024Unexpected Therapeutic Results; Unexpected Therapeutic Results

    SORIN GROUP ITALIA RECEIVED A REPOR OF AN INSPIRE OXYGENATOR FAILURE DURING A PROCEDURE, THE FIO2 WAS 95% AND PO2 SHOWS ~20%. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19071303 ↗
  • MalfunctionAug 2, 2023Increase in Pressure; Unexpected Therapeutic Results; Increase in Pressure; Unexpected Therapeutic Results

    SORIN GROUP ITALIA RECEIVED A REPORT THAT HIGH TRANSMEMBRANE PRESSURE OCCURRED DURING A PROCEDURE. MEDICAL TEAM ELECTED TO ADMINISTER MAG 2GM AND ALBUMIN 12.5GM. SURGERY WAS COMPLETED WITHOUT ANY NEED FOR UNIT CHANGE-OUT. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17442871 ↗
  • MalfunctionJun 7, 2023Fluid/Blood Leak; Fluid/Blood Leak

    SORIN GROUP ITALIA RECEIVED A REPORT THAT, AFTER ENTERING INTO BYPASS, A BLOOD LEAK WAS OBSERVED FORM THE TEMPERATURE PROBE HOLDER OF THE INSPIRE 6F OXYGENATOR. MEDICAL TEAM ELECTED TO CHANGE OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17077048 ↗
  • MalfunctionOct 26, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING SET UP OF THE HEART LUNG MACHINE, THE PERFUSIONIST NOTICED A PIECE OF FOREIGN MATTER IN THE INSPIRE 6F OXYGENATOR ARTERIAL OUTLET. THE OXYGENATOR WAS NOT USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15672528 ↗
  • MalfunctionSep 6, 2022Fluid/Blood Leak; Fluid/Blood Leak

    SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING A MITRAL VALVE REPLACEMENT SURGERY, WHILE ON PUMP, BLOOD LEAK FROM THE PURGE LINE OF THE INSPIRE 6F OXYGENATOR WAS OBSERVED. MEDICAL TEAM ELECTED TO CHANGE-OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15359897 ↗
  • MalfunctionAug 5, 2022Unexpected Therapeutic Results; Unexpected Therapeutic Results

    SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT OF ARTERIAL BLOOD DESATURATION HALF AN HOUR AFTER THE BEGINNING OF THE CEC USING AN INSPIRE 6F OXYGENATOR. MEDICAL TEAM ELECTED TO CHANGE-OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15173416 ↗
  • MalfunctionApr 15, 2022Unexpected Therapeutic Results; Unexpected Therapeutic Results

    SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING A PROCEDURE USING AN INSPIRE 6F OXYGENATOR, THE OXYGEN VALUES IN THE PATIENT BLOOD WERE LOW. THERE IS NO REPORT OF ANY PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14117000 ↗
  • MalfunctionMar 21, 2019Particulates; Particulates

    SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, DURING DEVICE SET-UP, A BLACK FOREIGN PARTICLE WAS IDENTIFIED BY THE OUTLET SIDE OF INSPIRE 6F INTEGRATED ARTERIAL FILTER. THE ISSUE WAS IDENTIFIED PRIOR TO ANY PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8440408 ↗
  • MalfunctionAug 3, 2018Fluid/Blood Leak; Fluid/Blood Leak

    SORIN (B)(4) RECEIVED A REPORT OF A BLOOD LEAK FROM THE LOWER PORTION OF AN INSPIRE 6F HOLLOW FIBER OXYGENATOR DURING A PROCEDURE. ACCORDING TO THE CUSTOMER, THE TOTAL BLOOD LOSS WAS OF 1000 ML AND THE PATIENT WAS HEMO-TRANSFUSED. THERE WAS NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7746125 ↗

Other OXYGENATOR, CARDIOPULMONARY BYPASS devices

All OXYGENATOR, CARDIOPULMONARY BYPASS devices →