Skip to content

PHYSICA system 9095.11.D50Recalls & reported problems

Limacorporate SPA · Headed Screw Dia. 3 mm x L. 50 mm · Class 1

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08033390123693), received between Mar 24, 2021 and Oct 21, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem3
  • Physical Resistance/Sticking3
  • Break1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2025Break; Break

    DURING A KNEE SURGERY (LATERAL UNI), TWO HEADED SCREW L.50 MM (PRODUCT CODE: 909511D50S, LOT: 2523343 - STER. (B)(4) PINS BROKE. BOTH BROKEN PARTS ARE RETAINED IN THE PATIENT'S FEMUR. THE EVENT OCCURRED AS THE SURGEON UNSCREWED THE FEMORAL CUT BLOCK. THE PROCEDURE WAS DELAYED OF 10 MINUTES. EVENT OCCURRED IN UNITED KINGDOM.

    Patient problems: Insufficient Information

    MAUDE report 23345066 ↗
  • MalfunctionMar 24, 2021Mechanical Problem; Physical Resistance/Sticking; Mechanical Problem

    DURING KNEE SURGERY PERFORMED ON (B)(6) 2021, IT WAS NOT POSSIBLE TO UNSCREW THE PIN EXAGONAL HEAD SCREW DIA.3X50 MM (PRODUCT CODE 9095.11.D50, LOT #19AT42Z). THE ISSUE OCCURRED DURING USE WITH THE POWERED PIN DRIVER EXAGONAL HEAD (PRODUCT CODE 9095.11.121, LOT 19AG08L). ACCORDING TO COMPLAINT SOURCE, PATIENT'S BONE WAS HARD. THE PIN WAS THEN REMOVED USING FORCEPS AND THE SURGERY WAS EXTENDED BY 20 MINUTES. THIS EVENT OCCURRED IN ITALY. NOTE: THIS COMPLAINT WAS REPORTED TO FDA AS THE PIN IS FDA CLEARED. HOWEVER, IN THIS CASE IT WAS USED IN COMBINATION WITH A POWERED PINDRIVER DURING A PHYSICA ZUK KNEE SURGERY, WHICH IS NOT COMMERCIALIZED IN THE USA.

    Patient problems: Insufficient Information

    MAUDE report 11560631 ↗
  • MalfunctionMar 24, 2021Mechanical Problem; Physical Resistance/Sticking; Mechanical Problem

    DURING KNEE SURGERY PERFORMED ON (B)(6) 2021, IT WAS NOT POSSIBLE TO UNSCREW THE PIN EXAGONAL HEAD SCREW DIA.3X50 MM (PRODUCT CODE 9095.11.D50, LOT #19AT07Y). THE ISSUE OCCURRED DURING USE WITH THE POWERED PIN DRIVER EXAGONAL HEAD (PRODUCT CODE 9095.11.121, LOT 19AG08L). ACCORDING TO COMPLAINT SOURCE, PATIENT'S BONE WAS HARD. THE PIN WAS THEN REMOVED USING FORCEPS AND THE SURGERY WAS EXTENDED BY 20 MINUTES. EVENT OCCURRED ON ITALY. NOTE: THIS COMPLAINT WAS REPORTED TO FDA AS THE PIN IS FDA CLEARED. HOWEVER, IN THIS CASE IT WAS USED IN COMBINATION WITH A POWERED PINDRIVER DURING A PHYSICA ZUK KNEE SURGERY, WHICH IS NOT COMMERCIALIZED IN THE USA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11560628 ↗
  • MalfunctionMar 24, 2021Mechanical Problem; Physical Resistance/Sticking; Mechanical Problem

    DURING KNEE SURGERY PERFORMED ON (B)(6) 2021, IT WAS NOT POSSIBLE TO UNSCREW THE PIN EXAGONAL HEAD SCREW DIA.3X50 MM (PRODUCT CODE 9095.11.D50, LOT 19AT07Y) . THE ISSUE OCCURRED DURING USE WITH THE POWERED PIN DRIVER EXAGONAL HEAD (PRODUCT CODE 9095.11.121, LOT 19AG08L). ACCORDING TO COMPLAINT SOURCE, PATIENT'S BONE WAS HARD. THE PIN WAS THEN REMOVED USING FORCEPS AND THE SURGERY WAS EXTENDED BY 20 MINUTES. EVENT OCCURRED IN ITALY. NOTE: THIS COMPLAINT WAS REPORTED TO FDA AS THE PIN IS FDA CLEARED. HOWEVER, IN THIS CASE IT WAS USED IN COMBINATION WITH A POWERED PINDRIVER DURING A PHYSICA ZUK KNEE SURGERY, WHICH IS NOT COMMERCIALIZED IN THE USA.

    Patient problems: Insufficient Information

    MAUDE report 11560626 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.