Skip to content

BIOPSYBELL DISKOMRecalls & reported problems

Biopsybell Srl · DISC DECOMPRESSION NEEDLE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08033860019747), received between Apr 3, 2024 and Apr 3, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Spinal Column Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2024Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    ON 04MAR2024 OUR US DISTRIBUTOR SOLUTION MEDICAL IN THE PERSON OF (B)(6) NOTIFIED OUR CUSTOMER CARE ABOUT THE SIGNALATION. ON 06MAR2024 AFTER FIRST REQUEST FOR DETAIL BY QUALITY WE RECEIVED THE FOLLOWING INFORMATION: A BRIEF DESCRIPTION RELEASED BY SURGICAL CENTRE: CERVICAL PERCUTANEOUS DISCECTOMY. ¾ INCH OF THE WIRE BROKE OFF IN PATIENT. PATIENT THEN TRANSPORTED TO HOSPITAL TO HAVE PIECE REMOVED. PHYSICIAN RAN THE DISKOM DEVICE FOR 5 MINUTES INSTEAD OF 2 MINUTES IN A BACK-AND-FORTH MOTION PATIENT NAME - INITIALS: (B)(6) - AGE OF PATIENT / BIRTH DATE: (B)(6) 1988 - MALE - 190LBS AND FURTHER INFORMATION BELOW: 1. CERVICAL PERCUTANEOUS DISCECTOMY PERFORMED IN C5-C6 2. ABOUT ¾ INCH OF THE PROBE WIRE BRAKE. THE MOMENT IT HAPPENED IS UNKNOWN, THE PHYSICIAN SAID SHE REALIZED IT WHEN SHE NOTICED THE PIECE OF WIRE IN THE PATIENT ON THE X RAY IMAGE 3. PICTURE AVAILABLE. THE PATIENT WAS TRANSPORTED TO HOSPITAL. HAD SURGERY TO REMOVE THE WIRE. THE PHYSICIAN ON MONDAY (B)(6) 2024 SAID THE PATIENT WAS AT HOME DOING FINE. 4. THE SURGERY CENTER ACCIDENTLY THREW OUT THE STICKER WITH THE SERIAL #. ALL THAT IS KNOWN IS THE LOT # ( 2401114). 5. THERE HAVE BEEN NO OTHER ISSUES ASSOCIATED WITH THIS LOT

    Patient problems: Spinal Column Injury

    MAUDE report 19032655 ↗

Other Arthroscope devices

All Arthroscope devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page