A.M.G.PMP PEDIATRIC A.M.G.PMP PEDIATRIC STERILERecalls & reported problems
Eurosets Srl · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08052705493351), received between Feb 18, 2025 and Apr 8, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak5
- Gas/Air Leak1
- Improper or Incorrect Procedure or Method1
- Insufficient Device Problem Information1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Embolism/Embolus1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Fluid/Blood Leak; Gas/Air Leak
IT WAS REPORTED THAT WHILE PRIMING, OXYGENATOR WAS NOTICED THAT THE TOP WAS FILLING WITH WATER AND THE OXYGEN PORT WAS BREACHING. THIS WAS A PEDIATRIC OXYGENATOR, AND THIS HAPPENED DURING THE BLOOD PRIME. NO BLOOD LEAK WAS NOTED. THE OXYGENATOR WAS EXCHANGED. AFTER FURTHER INFORMATION, POSSIBLE BLOOD LEAK WITH WATER WAS ALSO CONCERN. THEY HAD TO CHANGE THE COMPLETE CIRCUIT CONCERNED FOR BLOOD CONTAMINATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24821511 ↗InjuryDec 23, 2025Fluid/Blood Leak; Mechanical Problem; Improper or Incorrect Procedure or Method; Fluid/Blood Leak; Mechanical Problem; Improper or Incorrect Procedure or Method
THE BLOOD FLOW AND GAS FLOW RATE WAS UNKNOWN. THE OXYGENATOR INLET AND OUTLET PRESSURES AT THE TIME THE ISSUE OCCURRED WAS UNKNOWN SINCE IT WAS NOT MONITORED. THE ISSUE OCCURRED AT AN UNKNOWN MOMENT AFTER INITIATION OF SUPPORT. IT WAS UNKNOWN IF THE DEVICE WOULD BE RETURNING FOR EVALUATION. THE PUMP MOTOR SPEED (RPM) WAS UNKNOWN. THE ACT/PTT WAS UNKNOWN STATUS POST THE ROSS PROCEDURE. THERE WERE NO ISSUES DURING PRIMING OF THE OXYGENATOR PRIOR TO SUPPORT. THE SOURCE OF THE LEAK WAS UNKNOWN. NO BLOOD WAS DRIPPING FROM THE OXYGENATOR. NO PHOTOS OR VIDEOS OF THE LEAK/DRIP WERE AVAILABLE. NO PHOTOS OF THE THROMBUS/CLOT WERE AVAILABLE. IT WAS CONFIRMED WITH THE CHIEF PERFUSION WHO ADDED THE ATTACHED EMAIL WITH FURTHER INFORMATION. AFTER A PHONE CALL WITH THE PERFUSIONIST, FOR WHAT IT IS WORTH, THE CHIEF PERFUSIONIST DID NOT DETERMINE THE OXYGENATOR TO BE OF FAULT. THIS SEEMED LIKE A USER ERROR OR LACK OF EDUCATION AND INEXPERIENCE AND WAS AN UNNECESSARY CHANGE OUT.
Patient problems: Embolism/Embolus
MAUDE report 23881925 ↗MalfunctionJun 30, 2025Fluid/Blood Leak; Insufficient Device Problem Information; Fluid/Blood Leak; Insufficient Device Problem Information
IT WAS REPORTED THAT AN OXYGENATOR WAS RETURNED FOR AN UNKNOWN REASON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22361915 ↗MalfunctionMar 26, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE OXYGENATOR BEGAN LEAKING BLOOD AFTER TWO DAYS. WHOLE RED BLOOD IN THE GAS OUTLET PHASE OF THE LOWER PORTION OF THE OXYGENATOR WAS NOTED. THERE WAS NO INCREASE IN PRESSURE NOTED AND NO DECREASE IN GAS TRANSFER. THE PUMP SPEED AND BLOOD FLOW RATE WAS 2850 RPMS AND 1.28 LPM. THE BLOOD INLET PRESSURE WAS 200MMHG AND THE BLOOD OUTLET PRESSURE WAS 176MMHG. THE ACTIVATED CLOTTING TIME/PARTIAL THROMBOPLASTIN TIME (ACT/PTT) WAS 70.8. THE PATIENT WAS ON SUPPORT AND STABLE WHEN THEY ELECTED TO CHANGE THE OXYGENATOR FOR A NEW ONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21700652 ↗MalfunctionFeb 18, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE AFTER PRIMING THE AMG PMP OXYGENATOR AND INSTITUTING EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) THE ICU STAFF NOTICED A LARGE AMOUNT OF DILUTED BLOODY FLUID IN THE LOWER CHAMBER OF THE AMG BELOW THE FIBER BUNDLE. THE STAFF IMMEDIATELY CHANGED THE OXYGENATOR UNDER CONTROLLED CIRCUMSTANCES AND DID NOT RELATE ANY ISSUES OR SEQUALAE TO THE PATIENT. THE SITE NOTED THAT THE PUMP SPEED WAS 3900 ROTATIONS PER MINUTE (RPM), BLOOD FLOW WAS 1.6 LITERS PER MINUTE, INLET PRESSURE WAS 280 MILLIMETERS OF MERCURY (MMHG), AND OUTLET PRESSURE WAS 246 MMHG. THE PARTIAL THROMBOPLASTIN TIME (PTT) WAS 69.2. NO ISSUES WERE REPORTED DURING THE PRIMING OF THE OXYGENATOR. THE ISSUE OCCURRED APPROXIMATELY 22 HOURS AFTER INITIATION OF SUPPORT. THE DEVICE WAS DISPOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21392365 ↗
Other Oxygenator, Cardiopulmonary Bypass devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths90 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-i HMO 50000-USA #SQUADROX-i Small Adult, no FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page