The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08052705496253), received between Jun 13, 2023 and Aug 2, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 2, 2024Fluid/Blood Leak; Mechanical Problem; Fluid/Blood Leak; Mechanical Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS NOT HARMED. THE REPLACEMENT OXYGENATOR WORKED WELL. THE PUMP SPEED WAS 2400 RPM, THE FLOW WAS 0.8 L/MIN, AND THE GAS FLOW WAS 0.5 L/MIN. THERE WAS A DIALYSIS FILTER BETWEEN THE PUMP AND THE OXYGENATOR. THE ACTIVATED CLOTTING TIME (ACT) WAS APPROXIMATELY 200. AFTER A COUPLE OF HOURS OF SUPPORT, THE PATIENT STARTED TO WAKE UP AND COUGH, AND THE INTENSIVE CARE NURSE NOTICED BLOOD DRIPPING FROM THE BOTTOM OF THE OXYGENATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19893678 ↗MalfunctionJun 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THE CENTER HAD AN OUT OF BOX FAILURE ON AN OXYGENATOR. THE OXYGENATOR WAS LEAKING. THERE WAS NO VISIBLE DAMAGE REPORTED TO THE BOX OR THE OXYGENATOR. IT WAS REMOVED FROM USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17119843 ↗
FDA clearance and classification
The FDA regulates product code DTZ (oxygenator, cardiopulmonary bypass) under 21 CFR 870.4350 as a Class II device. 21 CFR 870.4350 ↗
Devices of this type normally need 510(k) clearance before sale.
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the A.M.G. PP Sterile A.M.G. MODULE PP NO T.TP.STER.
Has the A.M.G. PP Sterile A.M.G. MODULE PP NO T.TP.STER. been recalled?
No FDA recall record in our data names this exact device identifier (DI 08052705496253), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the A.M.G. PP Sterile A.M.G. MODULE PP NO T.TP.STER.?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jun 13, 2023 and Aug 2, 2024. In those reports, the most common specific device problems are “Fluid/Blood Leak” (2) and “Mechanical Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the A.M.G. PP Sterile A.M.G. MODULE PP NO T.TP.STER.?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the A.M.G. PP Sterile A.M.G. MODULE PP NO T.TP.STER., and is it still sold?
The FDA's device database lists Eurosets Srl as the company responsible for it (brand A.M.G. PP Sterile, model A.M.G. MODULE PP NO T.TP.STER.). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Oxygenator, Cardiopulmonary Bypass devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths87 reports
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-i HMO 50000-USA #SQUADROX-i Small Adult, no FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page