A.M.G. PP Sterile A.M.G. MODULE PP NO T.TP.STER.Recalls & reported problems

Eurosets Srl · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08052705496253), received between Jun 13, 2023 and Aug 2, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 2, 2024Fluid/Blood Leak; Mechanical Problem; Fluid/Blood Leak; Mechanical Problem

    ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS NOT HARMED. THE REPLACEMENT OXYGENATOR WORKED WELL. THE PUMP SPEED WAS 2400 RPM, THE FLOW WAS 0.8 L/MIN, AND THE GAS FLOW WAS 0.5 L/MIN. THERE WAS A DIALYSIS FILTER BETWEEN THE PUMP AND THE OXYGENATOR. THE ACTIVATED CLOTTING TIME (ACT) WAS APPROXIMATELY 200. AFTER A COUPLE OF HOURS OF SUPPORT, THE PATIENT STARTED TO WAKE UP AND COUGH, AND THE INTENSIVE CARE NURSE NOTICED BLOOD DRIPPING FROM THE BOTTOM OF THE OXYGENATOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19893678 ↗
  • MalfunctionJun 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE CENTER HAD AN OUT OF BOX FAILURE ON AN OXYGENATOR. THE OXYGENATOR WAS LEAKING. THERE WAS NO VISIBLE DAMAGE REPORTED TO THE BOX OR THE OXYGENATOR. IT WAS REMOVED FROM USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17119843 ↗

Other Oxygenator, Cardiopulmonary Bypass devices

All Oxygenator, Cardiopulmonary Bypass devices →