Skip to content

BRM TOOL SCREWS INSTRUMENT Slim cannulated screwdriver Hex 2.5Recalls & reported problems

Brm Extremities Srl · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08053782780143), received between Apr 24, 2026 and Apr 24, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2026Fracture

    ON (B)(6) 2026, A SURGICAL PROCEDURE WAS PERFORMED IN COLOMBIA. DURING THE PROCEDURE, A 4.0 DIAMETER SCREW WAS IMPLANTED. INITIALLY, THE NECESSARY DRILLING AND COUNTERSINKING WERE PERFORMED. SUBSEQUENTLY, THE 4.0 SCREW WAS INSERTED. UPON COMPLETION OF THE SCREW INSERTION, A CRACKING SOUND WAS HEARD. UPON REMOVAL OF THE HEXAGONAL CANNULATED SCREWDRIVER, IT WAS FOUND THAT THE FINAL PORTION WAS FRACTURED. INTRAOPERATIVE RADIOGRAPHIC ASSESSMENT WAS PERFORMED TO LOCATE EVENTUAL FRACTURED METALLIC FRAGMENTS. NO FRAGMENT WAS VISIBLE IN THE SOFT TISSUES AND THE SCREW WAS IMPLANTED IN THE CORRECT POSITION. SINCE THE SCREW WAS ALREADY FULLY INSERTED INTO THE BONE, THE SURGEON DECIDED TO COMPLETE THE PROCEDURE WITHOUT FURTHER ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24983428 ↗

Other Orthopedic Manual Surgical Instrument devices

All Orthopedic Manual Surgical Instrument devices →