BIO-FLASH 3710-0344Recalls & reported problems
Biokit, S.A. · BIO-FLASH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08426950639725), received between Dec 27, 2024 and Dec 27, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “False Negative Result” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- False Negative Result1
- Insufficient Device Problem Information1
Patient problems
- Kidney injury1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 27, 2024False Negative Result; Insufficient Device Problem Information; False Negative Result; Insufficient Device Problem Information
THE QUANTA FLASH® GBM IS A CHEMILUMINESCENT IMMUNOASSAY (CIA) FOR THE SEMI-QUANTITATIVE DETECTION OF IGG ANTI-GLOMERULAR BASEMENT MEMBRANE (GBM) AUTOANTIBODIES IN HUMAN SERUM ON THE BIO-FLASH® INSTRUMENT. QUANTA FLASH® GBM IS AN AID IN THE DIAGNOSIS OF GOODPASTURE'S SYNDROME IN CONJUNCTION WITH CLINICAL FINDINGS AND OTHER LABORATORY TESTS. COMPLAINT (B)(4) REPORTED THAT A PATIENT WITH ACUTE KIDNEY INJURY PRESENTED WITH A NEGATIVE QUANTA FLASH GBM REAGENT (LOT NUMBER 244154) RESULT ON (B)(6) 2024. ON (B)(6) 2024 THE SAMPLE WAS RETESTED USING THE PHADIA MPO AND PR-3 ASSAYS, WHICH TESTED LOW POSITIVE FOR MPO AND NEGATIVE FOR PR-3. A BIOPSY WAS ORDERED (DATE UNKNOWN) BASED UPON THE PHADIA LOW POSITIVE MPO TEST RESULT. THE BIOPSY WAS POSITIVE FOR IGG ANTI-GLOMERULAR BASEMENT MEMBRANE (GBM) DISEASE. PLEX TREATMENT (PLASMA EXCHANGE) WAS ADMINISTERED FOLLOWING THE BIOPSY RESULTS. HOWEVER, THE CUSTOMER REPORTED THAT THE DELAY IN DIAGNOSIS LED TO IRREVERSIBLE DESTRUCTION OF ORGAN AND PERMANENT NEED FOR DIALYSIS.
Patient problems: Kidney injury
MAUDE report 21031210 ↗
Other System, Multipurpose For In Vitro Coagulation Studies devices
- ACL TOP 970 CL 00000280097Instrumentation Laboratory Company · System, Multipurpose For In Vitro Coagulation Studies2 recalls0 reports
- TEG5000 HEMOSTASIS ANALYZER 07-022Haemonetics Corporation · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall2 reports
- STA Satellite® 58104Diagnostica Stago · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall1 report
- ACL ELITE PRO 00001000311RInstrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE PRO 00001000311Instrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE 00000880311RInstrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- ACL ELITE 00000880311Instrumentation Laboratory Company · SYSTEM, MULTIPURPOSE FOR IN VITRO COAGULATION STUDIES1 recall0 reports
- Hemochron Whole Blood Microcoagulation System ELITEAccriva Diagnostics Holdings, Inc. · System, Multipurpose For In Vitro Coagulation Studies2 deaths13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page