Skip to content

SEDECAL MobileDiagnost wDRRecalls & reported problems

Sedecal · This mobile digital radiographic system is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08436046002173), received between Oct 19, 2018 and Oct 19, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Oct 14, 2021Open, Classified

    While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.

    Z-1083-2024 · Root cause: Use error · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Device Fell1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 19, 2018Device Dislodged or Dislocated; Device Fell

    THIS IS A MOBILE DIAGNOSTIC X-RAY UNIT. THE COLLIMATOR ON THE UNIT WAS SERVICED. AS A RESULT OF IMPROPER RE-ATTACHMENT, THE COLLIMATOR FELL OFF THE TUBE-HEAD DURING SUBSEQUENT USE. IT WAS THEN LEFT DANGLING BY ITS CABLE. NO ONE WAS INJURED. A SERVICE PERSON SUBSEQUENTLY RE-ATTACHED THE COLLIMATOR IN A PROPER FASHION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7984277 ↗

Other Solid State X-Ray Imager (Flat Panel/Digital Imager) devices

All Solid State X-Ray Imager (Flat Panel/Digital Imager) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page