NA M0064201110Recalls & reported problems

Boston Scientific Corporation · Percutaneous Access Set · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729009733), received between May 1, 2019 and Feb 11, 2025 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Device Dislodged or Dislocated1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding4
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1
  • Urinary Tract Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A NEPHROSTOMY CATHETER SET WAS USED TO ESTABLISH PERCUTANEOUS NEPHROSTOMY ACCESS DURING A PERCUTANEOUS NEPHROSTOMY PROCEDURE PERFORMED ON (B)(6) 2025. FOLLOWING HOSPITAL DISCHARGE, THE PATIENT DEVELOPED A URINARY INFECTION DURING THE CATHETER INDWELLING PERIOD. NO ADDITIONAL ADVERSE EVENTS WERE REPORTED. PER PHYSICIAN'S ASSESSMENT, THE EVENT WAS NOT RELATED TO THE DEVICE.

    Patient problems: Urinary Tract Infection

    MAUDE report 21352924 ↗
  • InjuryFeb 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A NEPHROSTOMY CATHETER SET WAS USED TO ESTABLISH PERCUTANEOUS NEPHROSTOMY ACCESS AND URINE DRAINAGE DURING A PERCUTANEOUS NEPHROSTOMY WITH PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE PERFORMED ON (B)(6) 2024. FOLLOWING HOSPITAL DISCHARGE, THE PATIENT EXPERIENCED BLEEDING DURING THE CATHETER INDWELLING PERIOD, REQUIRING INTERVENTION. ANGIOEMBOLIZATION WAS PERFORMED FOR RENAL BLEEDING, THOUGH DETAILS OF ANY ADDITIONAL INTERVENTION WERE NOT REPORTED. THE CATHETER WAS REMOVED ON (B)(6) 2024, WITH NO SERIOUS ADVERSE EVENTS REPORTED IN CONNECTION TO THIS INCIDENT. PER PHYSICIAN'S ASSESSMENT, THE EVENT WAS NOT RELATED TO THE DEVICE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21343140 ↗
  • MalfunctionJan 28, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A NEPHROSTOMY CATHETER SET WAS USED TO ESTABLISH PERCUTANEOUS NEPHROSTOMY ACCESS AND URINE DRAINAGE DURING A PERCUTANEOUS NEPHROSTOMY PROCEDURE PERFORMED ON (B)(6) 2024. FOLLOWING THE PROCEDURE, THE PATIENT PULLED ON TUBE POST PROCEDURE AND DISLODGED IT FROM COLLECTING SYSTEM. THE CATHETER WAS REMOVED ON (B)(6) 2024. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21260279 ↗
  • InjuryJan 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A NEPHROSTOMY PERCUTANEOUS ACCESS SET WAS USED DURING A PERCUTANEOUS NEPHROSTOMY WITH PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE. ON (B)(6) 2024, DURING HOSPITALIZATION AND PRIOR TO DISCHARGE THE PATIENT EXPERIENCED BLEEDING. REPORTEDLY, THERE WAS NO RELATIONSHIP BETWEEN THE ADVERSE EVENT AND THE NEPHROSTOMY PERCUTANEOUS ACCESS SET. ADDITIONALLY, THERE WAS NO REPORTED MALFUNCTION OR DEFICIENCY OF THE NEPHROSTOMY PERCUTANEOUS ACCESS SET. THE DEVICE WAS REMOVED ON (B)(6) 2024. NO INTERVENTION WAS REQUIRED FROM THE PATIENT OTHER THAN BEDREST. THERE WERE NO ADDITIONAL PATIENT COMPLICATIONS REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21123958 ↗
  • InjuryJan 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A NEPHROSTOMY PERCUTANEOUS ACCESS SET WAS USED DURING A PERCUTANEOUS NEPHROSTOMY WITH PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE. ON (B)(6) 2024, DURING THE PROCEDURE THE PATIENT EXPERIENCED BLEEDING. REPORTEDLY, THERE WAS NO RELATIONSHIP BETWEEN THE ADVERSE EVENT AND THE NEPHROSTOMY PERCUTANEOUS ACCESS SET. ADDITIONALLY, THERE WAS NO REPORTED MALFUNCTION OR DEFICIENCY OF THE NEPHROSTOMY PERCUTANEOUS ACCESS SET. THE DEVICE WAS REMOVED ON (B)(6) 2024. THERE WERE NO ADDITIONAL PATIENT COMPLICATIONS REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21123956 ↗
  • InjuryMay 1, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PERCUTANEOUS ACCESS SET WAS USED IN THE KIDNEY DURING A PERCUTANEOUS NEPHROSTOMY PROCEDURE PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE DEVICE WAS USED AS A MICROPUNCTURE SYSTEM; HOWEVER, THE INDICATION FOR USE OF THE DEVICE IS FOR PERCUTANEOUS NEPHROSTOMY. REPORTEDLY, THIS EVENT CAUSED DELAY TO THE PROCEDURE WHICH RESULTED TO PAIN AND BLEEDING OF THE PATIENT. THE REPORTED INJURY DOES NOT REQUIRE FURTHER MEDICAL INTERVENTION AND THE PROCEDURE WAS COMPLETED WITH ANOTHER PERCUTANEOUS ACCESS SET. ADDITIONALLY, IT WAS CONFIRMED THAT THERE WAS NO ISSUE NOTED ON THE DEVICE. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pain

    MAUDE report 8570927 ↗

FDA clearance and classification

The FDA regulates product code GBO (catheter, nephrostomy, general & plastic surgery) under 21 CFR 878.4200 as a Class I device. 21 CFR 878.4200 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA M0064201110

Has the NA M0064201110 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729009733), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA M0064201110?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between May 1, 2019 and Feb 11, 2025. In those reports, the most common specific device problem is “Device Dislodged or Dislocated” (1); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (4), “Pain” (1) and “Urinary Tract Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA M0064201110?

None of the reports are classified as deaths, 5 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA M0064201110 FDA approved?

Devices of this type (product code GBO) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Boston Scientific Corporation.

Who makes the NA M0064201110, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand NA, model M0064201110). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY devices