Skip to content

Flexima™ M001271790Recalls & reported problems

Boston Scientific Corporation · Nephrostomy Catheter System · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08714729323303), received between Oct 4, 2022 and Apr 2, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Difficult to Advance2
  • Defective Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Internal Organ Perforation1
  • Perforation1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2024Difficult to Advance; Difficult to Advance

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 06MAR2024. IT WAS REPORTED THAT THE DEVICE WAS DIFFICULT TO ADVANCE. AN 8F FLEXIMA NEPHROSTOMY CATHETER SYSTEM WAS SELECTED FOR USE. DURING THE PROCEDURE, A HARD RESISTANCE WAS ENCOUNTERED UPON PUSHING THE CATHETER IN. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS REPORTED, AND THE PATIENT'S CONDITION WAS NORMAL. HOWEVER, INVESTIGATION OF THE RETURNED DEVICE REVEALED THAT THE SUTURE WAS DETACHED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19021708 ↗
  • InjuryJul 31, 2023Break; Break

    IT WAS REPORTED THAT THE STIFFENER GOT DETACHED AND PERFORATED THE KIDNEY HILUM. THE TARGET LESION WAS LOCATED IN THE KIDNEY AREA. A FLEXIMA CATHETER DRAINAGE WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE HUB OF THE STIFFENER GOT DETACHED, CAUSING THE STIFFENER TO ADVANCE THE CATHETER AND PERFORATE THE KIDNEY HILUM. THE DEVICE WAS REMOVED WITH EASE. A NEW NEPHROSTOMY TUBE WAS PLACED, SINCE THE PERFORATED RENAL PELVIS WILL HEAL ON ITS OWN. THE PHYSICIAN REVIEWED THE CT AND THERE WAS NO BLOOD VESSEL MEDIAL TO THE SITE OF PERFORATION. NO PATIENT COMPLICATIONS WERE REPORTED. DEVICE EVALUATED BY MANUFACTURER: THE DEVICE WAS RETURNED FOR ANALYSIS. VISUAL INSPECTION WAS PERFORMED. ONLY HUB CANNULA AND AN ACCESSORY WERE RETURNED FOR ANALYSIS. IT IS EVIDENCING THE SEPARATION OF THE METAL CANNULA FROM THE HUB. THE POUCH WAS RETURNED, AND IT WAS OBSERVED THAT THE POUCH INFORMATION MATCHES WITH THE COMPLAINT INFORMATION. NO ISSUES WERE OBSERVED IN THE DEVICES.

    Patient problems: Perforation

    MAUDE report 17426226 ↗
  • InjuryJul 28, 2023Break; Difficult to Advance; Break; Difficult to Advance

    IT WAS REPORTED THAT PATIENT INJURY OCCURRED. AN 8.3 FLEXIMA DRAINAGE CATHETER WAS SELECTED FOR RIGHT NEPHROSTOMY PLACEMENT IN THE KIDNEY. DURING THE PROCEDURE, IT WAS NOTED THAT THE RENAL COLLECTING SYSTEM WAS RUPTURED DUE TO INABILITY TO FEED OFF CATHETER FROM METAL STIFFENER. THE PATIENT HAS FULLY RECOVERED.

    Patient problems: Internal Organ Perforation

    MAUDE report 17418984 ↗
  • MalfunctionOct 4, 2022Break; Defective Device; Break; Defective Device

    IT WAS REPORTED THAT THE CATHETER BROKE. A FLEXIMA NEPHROSTOMY DRAINAGE CATHETER HAD BEEN PLACED FOR DRAINAGE OF THE KIDNEY. IT WAS NOTED THERE WAS DIFFICULTY PULLING THE SUTURE AND THE CATHETER BROKE. A NEW FLEXIMA CATHETER WAS USED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15538783 ↗

Other CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY devices

All CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page