Flexima™ M001271860Recalls & reported problems
Boston Scientific Corporation · Nephrostomy Catheter System Kit · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729323372), received between Aug 21, 2019 and Oct 15, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Deformation Due to Compressive Stress1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2025Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem; Deformation Due to Compressive Stress; Adverse Event Without Identified Device or Use Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 17SEP2025. IT WAS REPORTED THAT THE SHEATH BUCKLED. A VTC/8/K FLEXIMA REGULAR WAS SELECTED FOR A NEPHROSTOMY PROCEDURE. DURING PREPARATION, THE 0.018 WIRE SNAGGED WITHIN THE THREE-PART NEEDLE SHEATH AND CAUSED DAMAGE TO THE DEVICE ITSELF. THE VTC/8/K FLEXIMA REGULAR WAS NOT USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE RADIOPAQUE MARKER WAS TORN AT THE MIDDLE TO TIP SECTION OF THE SHEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23302274 ↗MalfunctionAug 21, 2019Break; Break
IT WAS REPORTED THAT THE RADIOPAQUE MARKER DETACHED. A VTC/8/K FLEXIMA REGULAR WAS SELECTED FOR USE. DURING PROCEDURE, THE RADIOPAQUE MARKER BECAME DETACHED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S CONDITION WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8919075 ↗
Other CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY devices
- NA M0067001210Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY2 deaths48 reports
- Flexima™ M001271800Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY6 reports
- NA M0064201110Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY6 reports
- Skater™ All-Purpose and Nephrostomy Set 756608025Argon Medical Devices, Inc. · Catheter, Nephrostomy, General & Plastic Surgery5 reports
- Flexima™ M001271790Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY4 reports
- Re-Entry Malecot Nephrostomy Catheter Set M0064101060Boston Scientific Corporation · CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY4 reports
- SKATER™ All Purpose and Nephrostomy Drainage Set 756608035Argon Medical Devices, Inc. · Catheter, Nephrostomy, General & Plastic Surgery4 reports
- GP General Purpose Drainage Catheter GPL2-1030HUresil, LLC · Catheter, Nephrostomy, General & Plastic Surgery1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.