CRE Wireguided M00558420Recalls & reported problems
Boston Scientific Corporation · Esophageal/Pyloric Wireguided Balloon Dilatation Catheter · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific CRE wireguided balloon dilatation catheter is used to widen narrowed areas in the esophagus and bile duct during endoscopy. The 33 reports mostly describe the balloon bursting or leaking fluid during use, sometimes springing back smaller or failing to inflate, which left the procedure to be finished with another catheter; no pieces were reported left inside patients. There are no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 08714729339359), received between Dec 6, 2018 and Sep 15, 2026 — classified as 36 malfunctions.
The most frequently reported specific device problem is “Material Puncture/Hole” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole17
- Burst Container or Vessel16
- Leak/Splash6
- Inflation Problem4
- Use of Device Problem4
- Material Split, Cut or Torn3
- Adverse Event Without Identified Device or Use Problem2
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Consequences Or Impact To Patient15
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Tear, Rip or Hole in Device Packaging
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50765971 ↗MalfunctionSep 11, 2026Burst Container or Vessel
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50744519 ↗MalfunctionSep 11, 2026Burst Container or Vessel
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50744485 ↗MalfunctionJun 16, 2026Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE FOR A STONE PERFORMED ON 17 MAY 2026. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER REACHING A DIAMETER OF 13.5MM. NO PIECE OF BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENTS CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25520931 ↗MalfunctionJun 10, 2026Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE FOR A STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER REACHING A DIAMETER OF 13.5MM. NO PIECE OF BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENTS CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25452721 ↗MalfunctionApr 13, 2026Burst Container or Vessel; Material Puncture/Hole; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. PRIOR TO THE PROCEDURE, THE PHYSICIAN ATTEMPTED TO INFLATE THE DEVICE; HOWEVER, THE BALLOON RUPTURED AND A HOLE WAS OBSERVED, RENDERING THE DEVICE UNUSABLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THE INSTRUCTIONS FOR USE (IFU) INDICATE THAT THIS BALLOON SHOULD NOT BE PRE-INFLATED PRIOR TO USE IN THE PROCEDURE. HOWEVER, THE CUSTOMER REPORTED THAT THE BALLOON WAS PRE-INFLATED OUTSIDE OF THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24858753 ↗MalfunctionMar 10, 2026Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ENDOSCOPIC ESOPHAGEAL DILATION PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, FOR THE FIRST DILATION, THE PRESSURE COULD NOT BE MAINTAINED. AFTER ABOUT TWO MINUTES, THE BALLOON SUDDENLY BECAME SMALLER, SO IT WAS REMOVED FOR CHECKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24564976 ↗MalfunctionDec 29, 2025Inflation Problem; Material Split, Cut or Torn; Inflation Problem; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILE DUCT DILATATION WITH CALCULI PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE NURSE PLACED THE CRE BALLOON AT THE PAPILLA AND DILATED IT WITH THE PRESSURE HANDLE, BUT THE LIQUID COULD NOT ENTER THE BALLOON. THE BALLOON COULD NOT BE OPENED AFTER REPEATED OPERATIONS. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF A BALLOON LONGITUDINALLY TORN. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23914423 ↗MalfunctionJul 11, 2025Material Puncture/Hole; Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn; Material Puncture/Hole; Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE DUODENAL PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WATER LEAKED DURING DILATION. IT WAS REPORTED THAT THE BALLOON HAD A HOLE. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: THE INSTRUCTIONS FOR USE (IFU) INDICATE THAT THIS BALLOON SHOULD NOT BE PRE-INFLATED BEFORE USE IN THE PROCEDURE. HOWEVER, THE CUSTOMER REPORTED THAT THE BALLOON WAS PRE-INFLATED BEFORE USE IN THE PATIENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF A BALLOON CIRCUMFERENTIAL TEAR. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22484512 ↗MalfunctionMar 20, 2025Inflation Problem; Material Puncture/Hole; Inflation Problem; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING A DILATION OF ESOPHAGEAL STRICTURE PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE BALLOON COULD NOT BE INFLATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF A BALLOON PINHOLE WHEN USED IN THE ESOPHAGUS. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21658510 ↗MalfunctionJan 4, 2024Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem; Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE ORIFICE OF COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, WHEN THE PRESSURE REACHED 6 ATM THE BALLOON BURST. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18448729 ↗MalfunctionJan 2, 2024Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE ORIFICE OF COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE PRESSURE CRITICAL VALUE WAS NOT REACHED WHEN THE DEVICE SUDDENLY BURST AT 7 MM. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18429048 ↗MalfunctionAug 30, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING A TRANSGASTROSCOPIC BALLOON DILATATION PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE PHYSICIAN INJECTED LIQUID TO INFLATE AND FOUND THAT THE PRESSURE GAUGE COULD NOT GO UP. THEN, THE PHYSICIAN TOOK OUT THE BALLOON AND INJECTED LIQUID OUTSIDE OF THE PATIENT'S BODY AND FOUND THAT THE BALLOON WAS LEAKING LIQUID. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17656896 ↗MalfunctionAug 29, 2023Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING AN ESOPHAGEAL STENT IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE BALLOON WAS LEAKING LIQUID AND THE CUSTOMER REPORTED THE BALLOON HAD A HOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17643140 ↗MalfunctionMay 15, 2023Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE ORIFICE OF THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE BALLOON BURST AT 5.5KPA/9MM AND WAS LEAKING. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16931616 ↗MalfunctionJan 28, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE BALLOON BURST DURING INFLATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13389459 ↗MalfunctionOct 4, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE ESOPHAGUS DURING AN ESOPHAGEAL STENT IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE BALLOON BURST DURING INFLATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12568792 ↗MalfunctionJun 22, 2021Inflation Problem; Material Puncture/Hole; Inflation Problem; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING A BALLOON DILATION ASSISTED LITHOTOMY PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE DEVICE WAS ATTEMPTED TO BE INFLATED; HOWEVER, THE BALLOON COULD NOT INFLATE COMPLETELY. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THIS BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12042468 ↗MalfunctionJun 14, 2021Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
THIS REPORT PERTAINS TO ONE OF TWO CRE WIREGUIDED DILATATION BALLOONS USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO CRE WIREGUIDED DILATATION BALLOONS WERE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON WAS LEAKING. THE SAME ISSUE OCCURRED WITH THE SECOND CRE WIREGUIDED DILATATION BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIRE GUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THIS BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11995174 ↗MalfunctionJun 14, 2021Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
THIS REPORT PERTAINS TO ONE OF TWO CRE WIREGUIDED DILATATION BALLOONS USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO CRE WIREGUIDED DILATATION BALLOONS WERE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON WAS LEAKING. THE SAME ISSUE OCCURRED WITH THE SECOND CRE WIREGUIDED DILATATION BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THIS BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11994986 ↗MalfunctionFeb 17, 2021Burst Container or Vessel; Inflation Problem; Burst Container or Vessel; Inflation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE PAPILLA DURING AN ENDOSCOPIC BALLOON SPHINCTEROPLASTY PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON CANNOT BE INFLATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE BALLOON BURST; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11341199 ↗MalfunctionJan 26, 2021Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE SIGMOID DURING A SIGMOID DILATION PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON BURST. REPORTEDLY, THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11228815 ↗MalfunctionNov 13, 2020Leak/Splash; Material Puncture/Hole; Material Split, Cut or Torn; Leak/Splash; Material Puncture/Hole; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE CYSTIC DUCT CALCULI DURING A CHOLEDOCHOLITHOTOMY PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON WAS LEAKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE BALLOON FOUND A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10840046 ↗MalfunctionOct 13, 2020Leak/Splash; Material Puncture/Hole; Leak/Splash; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED IN THE DUODENUM DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON WAS LEAKING. THE PROCEDURE WAS COMPLETED WTH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10673132 ↗MalfunctionAug 10, 2020Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE WIREGUIDED DILATATION BALLOON WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON WAS LEAKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE BALLOON HAD A PINHOLE; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10388206 ↗
Other Colonoscope and accessories, flexible/rigid devices
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-HQ190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths6,804 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-HQ190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls5 deaths6,244 reports
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-H190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid2 recalls7 deaths5,572 reports
- EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190DLOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2,830 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS PCF-H190LOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid3 recalls2 deaths2,513 reports
- EVIS EXERA lll COLONOVIDEOSCOPE OLYMPUS CF-H190IOlympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid1 recall1,967 reports
- PENTAX Medical EC38-i10cLHoya Corporation · Colonoscope and accessories, flexible/rigid1,618 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.