Sensor M0066703090Recalls & reported problems
Boston Scientific Corporation · Nitinol Wire with Hydrophilic Tip · Class 1
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08714729376378), received between Jul 27, 2018 and Feb 27, 2026 — classified as 4 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Detachment of Device or Device Component2
- Nonstandard Device2
- Adverse Event Without Identified Device or Use Problem1
- Device Contamination with Chemical or Other Material1
- Difficult to Advance1
- Peeled/Delaminated1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient1
- Unspecified Infection1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryFeb 27, 2026Break
IT WAS REPORTED THAT DURING A CYSTOSCOPY WITH RIGHT LASER LITHOTRIPSY AND SUCTION ASPIRATION OF STONE, RIGHT RETROGRADE PYELOGRAM, AND RIGHT URETERAL STENT PLACEMENT, A SENSOR WIRE WAS USED. THE SURGEON OBSERVED THAT THE SENSOR WIRE HAD BROKEN OFF IN THE KIDNEY. THE BROKEN WIRE WAS SUCCESSFULLY RETRIEVED WITHOUT DAMAGE TO SURROUNDING ANATOMICAL STRUCTURES. THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. NO HARM OCCURRED TO THE PATIENT, AND THE PATIENT RECOVERED AS EXPECTED AND WAS DISCHARGED HOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24472342 ↗MalfunctionJan 30, 2026
DURING A CYSTOSCOPY WITH RT LASER LITHOTRIPSY AND SUCTION ASPIRATION OF STONE, RT RETROGRADE PYELOGRAM AND RT URETAL STENT, THE SENSOR WIRE WAS NOTED BY THE SURGEON TO HAVE BROKEN OFF IN THE KIDNEY. THE WIRE WAS RETRIEVED WITHOUT DAMAGE TO THE STRUCTURE. CASE CONTINUED WITHOUT FURTHER INCIDENT, NO HARM TO THE PATIENT, PATIENT RECOVERED AND DISCHARGED HOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24220489 ↗MalfunctionJan 7, 2026Break; Nonstandard Device; Unraveled Material
IT WAS REPORTED THAT, DURING A TRANSURETHRAL LITHOTRIPSY PROCEDURE, A SENSOR WIRE WAS USED, FRAYING WAS OBSERVED ALONG THE LENGTH OF THE GUIDEWIRE, AND A TEAR WAS NOTED IN THE MIDDLE SECTION. IT WAS OBSERVED THAT IT GOT CAUGHT IMMEDIATELY AFTER THE GUIDEWIRE WAS STARTED TO USE. THE GUIDEWIRE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23995891 ↗MalfunctionNov 10, 2025Nonstandard Device; Difficult to Advance; Nonstandard Device; Difficult to Advance
IT WAS REPORTED THAT DURING THE PREPARATION, THE PHYSICIAN COMMENTED THAT THE SENSOR GUIDEWIRE WAS NOT SLIDING SMOOTHLY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE AND THERE WERE NO KNOWN PATIENT COMPLICATIONS REPORTED. THE ANALYSIS OF THE RETURNED DEVICE IDENTIFIED SLEEVE DETACHMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23513484 ↗MalfunctionFeb 13, 2025Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING PREPARATION, THE COATING OF THE SENSOR WIRE PEELED. THE PROCEDURE WAS COMPLETED OF ANOTHER OF THE SAME DEVICE AND THERE WERE NO PATIENT INJURY REPORTED. THE DEVICE WAS RETURNED, AND IT WAS IDENTIFIED THAT THE SLEEVE WAS DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21376979 ↗InjuryNov 2, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSOR WIRE WAS USED DURING A URETEROSCOPY PROCEDURE IN THE URETER OR BLADDER PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, WHEN THE GUIDE WIRE WAS PULLED OUT FROM THE PATIENT, THE BLUE COVERING WAS BROKEN. THE BROKEN PIECE OF THE GUIDEWIRE WAS SUCCESSFULLY RETRIEVED, AND AN X-RAY WAS COMPLETED AT THE END TO ENSURE NOTHING WAS IN THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE ORIGINAL SENSOR WIRE DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED THAT THE BROKEN PIECE OF WIRE WAS SUCCESSFULLY REMOVED USING GRASPERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18059700 ↗InjuryJun 6, 2023Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSOR WIRE WAS USED IN THE KIDNEY DURING AN ENDOSCOPY PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE TIP OF THE SENSOR WIRE BROKE OFF. THE PATIENT WILL REQUIRE ANOTHER PROCEDURE TO REMOVE THE STONE AND THE FRAGMENT THAT BROKE OFF WILL BE RETRIEVED AT THAT TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17070229 ↗MalfunctionMay 12, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSOR WIRE GUIDEWIRE WAS USED IN THE KIDNEY DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2023. DURING PROCEDURE, THE SENSOR WIRE SHRINK SLEEVE DETACHED. THE PROCEDURE WAS COMPLETED WITH A NEW SENSOR GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16920741 ↗InjuryNov 23, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT SENSOR WIRE GUIDEWIRE WAS USED IN THE KIDNEY AND URETER DURING A URETEROSCOPY WITH LASER LITHOTRIPSY PROCEDURE DATE IS UNKNOWN. ACCORDING TO THE COMPLAINANT, THE PATIENT WAS SENT HOME FROM THE PREVIOUS URETEROSCOPY WITH LASER LITHOTRIPSY PROCEDURE WITH A PORTION OF THE DETACHED COREWIRE FROM THE SENSOR WIRE. ON (B)(6) 2020 THE PATIENT RETURNED TO THE HOSPITAL SEPTIC. A URETEROSCOPY PROCEDURE WAS PERFORMED ON THE SAME DAY TO REMOVE THE RETAINED FRAGMENT. IT WAS REPORTED THE PATIENT IS EXPECTED TO FULLY RECOVER.
Patient problems: Unspecified Infection
MAUDE report 10885833 ↗MalfunctionJul 23, 2020Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT SENSOR WIRE GUIDEWIRE WAS USED DURING A KIDNEY STONE REMOVAL PROCEDURE ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SENSOR COREWIRE OF THE GUIDEWIRE BROKE WHEN IT WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NEW SENSOR WIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10315939 ↗MalfunctionJul 27, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SENSOR¿ GUIDEWIRE WAS USED IN A PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION OUTSIDE THE PATIENT, IT WAS NOTICED THAT THERE WAS A FOREIGN OBJECT AT THE TIP OF THE SENSOR GUIDEWIRE. THE PROCEDURE WAS COMPLETED WITH A NEW SENSOR¿ GUIDEWIRE. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient
MAUDE report 7726349 ↗
Other STYLET FOR CATHETER, GASTRO-UROLOGY devices
- NiCore Nitinol Guidewire 150NSS35C. R. Bard, Inc. · STYLET FOR CATHETER, GASTRO-UROLOGY229 reports
- Sensor M0066703080Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY118 reports
- Sensor M0066703050Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY41 reports
- Bard® Hydro-Glide™ Guidewire 145FS35C. R. Bard, Inc. · STYLET FOR CATHETER, GASTRO-UROLOGY31 reports
- NiCore Nitinol Guidewire 150NFS35C. R. Bard, Inc. · STYLET FOR CATHETER, GASTRO-UROLOGY30 reports
- Amplatz Super Stiff M0066401090Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY21 reports
- Amplatz Super Stiff M0066401080Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY18 reports
- Sensor M0066703120Boston Scientific Corporation · STYLET FOR CATHETER, GASTRO-UROLOGY17 reports