Jagwire M00556610Recalls & reported problems
Boston Scientific Corporation · High Performance Guidewire · Class 2
All models of Jagwire (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729747260), received between Sep 21, 2017 and Jun 21, 2023 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Break4
- Detachment of Device or Device Component2
- Peeled/Delaminated2
- Defective Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
- Perforation1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 21, 2023Peeled/Delaminated; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED IN THE SIGMOID COLON DURING A COLON STENT IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE GUIDEWIRE WAS PEELED AND DETACHED. THE PROCEDURE WAS COMPLETED WITH A NEW JAGWIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF THE CORE WIRE BROKEN. PLEASE SEE BLOCK H10 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17172173 ↗MalfunctionNov 29, 2022Break; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Break; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE GUIDEWIRE TIP SOFT PART OF ABOUT FIVE MM CAME OFF AND REMAINED IN THE BILE DUCT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS: THE COREWIRE WIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15882517 ↗MalfunctionOct 19, 2021Break; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem; Break; Peeled/Delaminated; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY(ERCP) PROCEDURE ON (B)(6) 2021. DURING THE PROCEDURE, PTFE COATING WAS PEELED. THE PROCEDURE WAS COMPLETED WITH A NEW JAGWIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF COREWIRE BREAK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12661202 ↗MalfunctionMay 27, 2021Break; Defective Device; Adverse Event Without Identified Device or Use Problem; Break; Defective Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE WAS USED IN A PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. DURING PREPARATION, THE GUIDEWIRE COREWIRE TIP WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH A NEW JAGWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JUNE 7, 2021: DURING PREPARATION WHEN THE COATING ON THE TIP WAS ACTIVATED USING SALINE BEFORE INSERTING INTO PATIENT, THE PHYSICIAN NOTICED THE GUIDEWIRE TIP WAS KINKED/BENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11895197 ↗MalfunctionDec 10, 2018Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED IN THE COMMON BILE DUCT DURING STONE REMOVAL PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, AFTER THE CANNULA WAS INSERTED INTO THE PATIENT, THE JAGWIRE GUIDEWIRE WAS ADVANCED INTO THE BILE DUCT THROUGH THE CANNULA. THE CANNULA WAS PULLED OUT AND AN ENDOSCOPIC SPHINCTEROTOMY DEVICE WAS THEN ADVANCED OVER THE GUIDEWIRE. HOWEVER, IT WAS NOTICED THAT THE CORE WIRE OF THE GUIDEWIRE WAS BROKEN WHICH FAILED THE OTHER DEVICES TO BE INSERTED. THE STUMP WAS CUT USING SCISSORS, AND THE PROCEDURE WAS CONTINUED. THE PROCEDURE WAS COMPLETED WITH A SECOND JAGWIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8147384 ↗InjurySep 21, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT # 3005099803-2017-02798 AND 3005099803-2017-02884 FOR THE ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2017 THAT A JAGWIRE GUIDEWIRE WAS USED TOGETHER WITH A WALLFLEX DUODENAL SOFT UNCOVERED STENT WHICH WAS IMPLANTED TO TREAT A SEVERE STENOSIS DUE TO A DUODENAL TUMOR FROM THE DUODENAL BULB TO THE DESCENDING LIMB OF THE DUODENUM DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2017. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE PHYSICIAN HAD DIFFICULTY ADVANCING THE GUIDEWIRE (THE SUBJECT OF MFR. REPORT#3005099803-2017-02884) THROUGH THE STENOSIS, WHICH REPORTEDLY TOOK ABOUT 90 MINUTES. AFTER THE STENT (THE SUBJECT OF THIS REPORT) WAS SUCCESSFULLY IMPLANTED, THE PHYSICIAN PERFORMED FLUOROSCOPY TO CHECK THE STENT'S PLACEMENT. IT WAS NOTED THAT THE CONTRAST MEDIA NEAR THE HORIZONTAL LIMB OF THE DUODENUM WAS NOT CLEAR AROUND THE DUODENUM AND THE PHYSICIAN SUSPECTED THAT THERE WAS A PERFORATION. A COMPUTED TOMOGRAPHY (CT) SCAN
Patient problems: Perforation
MAUDE report 6883872 ↗
FDA clearance and classification
The FDA regulates product code OCY (endoscopic guidewire, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Jagwire M00556610
Has the Jagwire M00556610 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729747260), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Jagwire M00556610?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Sep 21, 2017 and Jun 21, 2023. In those reports, the most common specific device problems are “Break” (4), “Detachment of Device or Device Component” (2) and “Peeled/Delaminated” (2); the most common patient problem is “Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Jagwire M00556610?
None of the reports are classified as deaths, 1 as an injury and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Jagwire M00556610 FDA approved?
Devices of this type (product code OCY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Boston Scientific Corporation.
Who makes the Jagwire M00556610, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Jagwire, model M00556610). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic guidewire, gastroenterology-urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology81 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death48 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.