Quick summary

This Boston Scientific Coyote ES balloon catheter has 437 adverse event reports, with 3 injuries and no deaths. The reports mostly describe the balloon rupturing during inflation in angioplasty procedures, often on arteries in the legs. Most reports record no clinical signs or effects on the patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
437
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 437 adverse event reports that list this exact device (DI 08714729767176), received between Mar 13, 2017 and Aug 17, 2026 — classified as 3 injuries, 434 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (416 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 5 reportsJan 2024: 4 reportsFeb 2024: 4 reportsMar 2024: 4 reportsApr 2024: 4 reportsMay 2024: 5 reportsJun 2024: 2 reportsJul 2024: 9 reportsAug 2024: 2 reportsSep 2024: 5 reportsOct 2024: 6 reportsNov 2024: 5 reportsDec 2024: 3 reportsJan 2025: 6 reportsFeb 2025: 4 reportsMar 2025: 3 reportsApr 2025: 3 reportsMay 2025: 1 reportsJun 2025: 4 reportsJul 2025: 6 reportsAug 2025: 2 reportsSep 2025: 8 reportsOct 2025: 0 reportsNov 2025: 5 reportsDec 2025: 6 reportsJan 2026: 7 reportsFeb 2026: 8 reportsMar 2026: 2 reportsApr 2026: 6 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture416
  • Failure to Advance10
  • Adverse Event Without Identified Device or Use Problem9
  • Break4
  • Entrapment of Device3
  • Material Integrity Problem3
  • Difficult to Advance2
  • Inflation Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions292
  • No Consequences Or Impact To Patient142
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 437 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50273299 ↗
  • MalfunctionAug 17, 2026Material Rupture

    It was reported that balloon rupture occurred.The 50% stenosed target lesion was located in the mildly tortuous and non-calcified pulmonary artery. A 2mm x 20mm x 143cm Coyote ES balloon catheter was advanced for dilatation. The balloon was inflated multiple times at 10, 12, and 14 atmospheres for 30 seconds each; however, during the third inflation at 14 atmospheres, the balloon ruptured with a pinhole rupture. The rupture location could not be confirmed because the device was discarded. The balloon was removed using the normal method without difficulty, and another Coyote ES was used for further inflation. The procedure was completed successfully using different types of balloons. No patient complications were reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50270987 ↗
  • MalfunctionJul 13, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON SECOND INFLATION AT 12 ATMOSPHERES FOR 5 SECONDS. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25793166 ↗
  • MalfunctionJun 11, 2026Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE BALLOON RUPTURED, AND THE PROCEDURE WAS CANCELLED. A 2MM X 40MM X 145CM COYOTE? ES BALLOON CATHETER WAS SELECTED FOR USE IN A FEMORO-POPLITEAL LEFT LEG CROSSOVER ANGIOPLASTY PROCEDURE. THE PATIENT HAD VASCULAR CALCIFICATIONS. AFTER DILATION, THE BALLOON RUPTURED DURING ITS REMOVAL. THE DEVICE WAS REPLACED WITH A 2MM X 20MM X 143CM COYOTE? ES BALLOON CATHETER, AND THE SAME PROBLEM OCCURRED. BOTH BALLOONS WERE SUCCESSFULLY REMOVED FROM THE BODY. THE PROCEDURE WAS CANCELLED. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477435 ↗
  • MalfunctionApr 22, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 50% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON SECOND INFLATION AT 6 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24958591 ↗
  • MalfunctionApr 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS VESSEL OF BELOW THE KNEE. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE SECOND INFLATION AT 8 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24924703 ↗
  • MalfunctionApr 8, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 1 ATMOSPHERE FOR 1 SECOND, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM USING THE NORMAL METHOD, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24820349 ↗
  • MalfunctionApr 7, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 5 ATMOSPHERES FOR 2 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM USING THE NORMAL METHOD, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24798257 ↗
  • MalfunctionApr 5, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 20MM X 143CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 1 ATMOSPHERE FOR 1 SECOND. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24787040 ↗
  • MalfunctionApr 2, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED PULMONARY ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 6 ATMOSPHERES FOR A FEW SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24764963 ↗
  • MalfunctionMar 23, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING FIRST INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED VERTICALLY. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24661387 ↗
  • MalfunctionMar 2, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED PULMONARY ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 10 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24481820 ↗
  • MalfunctionFeb 25, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. A 2 MM X 20 MM X 143 CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24447182 ↗
  • MalfunctionFeb 24, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 20MM X 143CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 5 ATMOSPHERES FOR 2 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24436763 ↗
  • MalfunctionFeb 23, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 8 ATMOSPHERES FOR SEVERAL SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24422593 ↗
  • MalfunctionFeb 20, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. A 2MM X 20MM X 143CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24403362 ↗
  • MalfunctionFeb 16, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 20MM X 143CM COYOTE? ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE SECOND INFLATION AT 14 ATMOSPHERES FOR 5 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24349879 ↗
  • MalfunctionFeb 6, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2MM X 20MM X 143CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE SECOND INFLATION AT 8 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24283587 ↗
  • MalfunctionFeb 4, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 20MM X 143CM COYOTE? ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT 8 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH THIS DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24262111 ↗
  • MalfunctionFeb 3, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 20MM X 143CM COYOTE ES WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 14 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24248810 ↗
  • MalfunctionJan 28, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE THIRD INFLATION PERFORMED 10 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED FROM A PINHOLE. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH AA DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198167 ↗
  • MalfunctionJan 23, 2026Fluid/Blood Leak

    IT WAS REPORTED BALLOON LEAK OCCURRED. PATIENT WAS PRESENTED WITH PULMONARY HYPERTENSION. THE TARGET LESION WAS LOCATED IN THE PULMONARY ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING INFLATION, IT WAS NOTED THAT THE BALLOON LEAK AIR. THE BALLOON WAS REPLACED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24165027 ↗
  • MalfunctionJan 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 2MM X 20MM X 143CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON WAS RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24116518 ↗
  • MalfunctionJan 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ILIAC ARTERY. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 14 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24115455 ↗
  • MalfunctionJan 19, 2026Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL OF BELOW THE KNEE. A 2MM X 20MM X 143CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING FIRST INFLATION AT 6 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24109251 ↗

Other Catheter, angioplasty, peripheral, transluminal devices