Coyote™ ES H74939135254010Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Coyote ES (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Coyote ES balloon catheter is a thin tube with a balloon used to widen narrowed blood vessels. Of its 128 reports, 127 are malfunctions and 1 is an injury, with no deaths; most describe the balloon rupturing during inflation, with another device used to complete the procedure. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 127
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 127 adverse event reports that list this exact device (DI 08714729767220), received between Dec 3, 2018 and Aug 17, 2026 — classified as 1 injury, 126 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (123 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture123
- Detachment of Device or Device Component2
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Remove1
- Failure to Advance1
- Fluid/Blood Leak1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions93
- No Consequences Or Impact To Patient34
- No Patient Involvement1
- Thrombosis/Thrombus1
- Vascular Dissection1
Recent adverse event reports
The 25 most recent of 127 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 17, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50278447 ↗MalfunctionMar 30, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL OF BELOW THE KNEE. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24729570 ↗MalfunctionFeb 16, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24349875 ↗MalfunctionFeb 10, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2.5MM X 40MM X 146CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 7 ATMOSPHERES FOR 3 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24309273 ↗MalfunctionFeb 6, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT NOMINAL PRESSURE. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24283568 ↗MalfunctionFeb 5, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24270836 ↗MalfunctionDec 9, 2025Fluid/Blood Leak; Material Rupture; Fluid/Blood Leak; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. THE LESION WAS ALSO FIRM, CREATING A SITUATION THAT REQUIRES THE PRODUCT TO DEMONSTRATE BOTH CROSSING AND INFLATION CAPABILITIES. A 2.5 MM X 40 MM X 146 CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 2 SECONDS, ANGIOGRAPHIC IMAGING REVEALED CONTRAST AGENT LEAKAGE, AND THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23760151 ↗MalfunctionNov 27, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23666174 ↗MalfunctionNov 25, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE THIRD INFLATION AT 14 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23649931 ↗MalfunctionNov 21, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE SEVENTH INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23616476 ↗MalfunctionNov 21, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT EXTERNAL ILIAC ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23615822 ↗MalfunctionNov 10, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED VESSEL. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES UNTIL REACHING 6 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23518722 ↗MalfunctionNov 5, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY (SFA). A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE FIRST INFLATION AT 14 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23478032 ↗MalfunctionOct 15, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES AND A 4MM X 40MM X 146CM COYOTE ES WERE SELECTED FOR USE IN AN ENDOVASCULAR TREATMENT. DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED HORIZONTALLY AT THE MIDDLE. THE DEVICE WAS RETRACTED INTO THE SHEATH AND WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23277953 ↗MalfunctionSep 8, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. A 2.5MM X 40MM X 146CM COYOTE BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR 20 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22995122 ↗MalfunctionSep 3, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2.5 MM X 40 MM X 146 CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22956670 ↗MalfunctionAug 12, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 40MM X 145CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND A PINHOLE WAS NOTED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22783276 ↗MalfunctionJul 1, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. A 2.5MM X 40MM X 146CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 12 ATMOSPHERES FOR SEVERAL SECONDS, THE BALLOON VERTICALLY RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22379180 ↗MalfunctionJun 26, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2.5MM X 40MM X 146CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT NOMINAL ATMOSPHERES FOR 3 SECONDS, THE BALLOON RUPTURED AT THE MIDDLE PART OF THE BALLOON. THE DEVICE WAS REMOVED, AND A PINHOLE WAS NOTED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22342605 ↗MalfunctionJun 23, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22305682 ↗MalfunctionApr 30, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL OF BELOW KNEE. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING INFLATION, IT WAS NOTED THAT THE BALLOON HAD ALREADY RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21937707 ↗MalfunctionMar 20, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION, WHICH WAS 100% STENOSED, WAS SITUATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED DURING THE FOURTH INFLATION AT 14 ATMOSPHERES, WHICH LASTED FOR SEVERAL SECONDS. THE DEVICE WAS SUCCESSFULLY REMOVED WITHOUT ANY ISSUES, AND THE PROCEDURE WAS COMPLETED USING AN ALTERNATIVE DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT REMAINED IN GOOD CONDITION FOLLOWING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21656875 ↗MalfunctionJan 20, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED LEFT ANTERIOR TIBIAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING FIRST INFLATION, THE BALLOON RUPTURED WHEN THE PRESSURE WAS INCREASED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21193126 ↗MalfunctionNov 8, 2024Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A SNARE WAS REQUIRED TO REMOVE A DETACHED BALLOON. A COYOTE BALLOON WAS SELECTED FOR USE. UNDER LOCAL ANESTHESIA, A RETROGRADE PUNCTURE WAS PERFORMED OF THE RIGHT COMMON FEMORAL ARTERY (CFA). A 4 FR SHEATH WAS INSERTED AND A PIGTAIL CATHETER WAS PLACED IN THE INFRARENAL AORTA. INJECTION OF A MECHANICAL CONTRAST MEDIUM FOLLOWED BY DIGITAL SUBTRACTION ANGIOGRAPHY (DSA) WAS PERFORMED TO VISUALIZE THE ILIAC VESSELS ON THE LEFT. ADDITIONAL CONTRAST MEDIUM INJECTED AND STAGED DSA OF THE LEFT ARTERIAL LEG CIRCULATION REVEALED HIGH-GRADE STENOSIS OF THE ANTERIOR TIBIAL ARTERY, AS WELL AS TWO FURTHER DOWNSTREAM SMALLER STENOSES. A NON-BSC GUIDEWIRE WAS INSERTED OVER THE STENOSIS. SUBSEQUENTLY, A COYOTE ES BALLOON WAS INSERTED HOWEVER, IT COULD NOT BE ADVANCED OVER THE ILIAC BIFURCATION. THE COYOTE ES BALLOON WAS WITHDRAWN BUT THE DISTAL PART OF THE CATHETER WAS SHEARED. THE PROXIMAL PART OF THE CATHETER WAS REMOVED BUT THE DISTAL PART REMAINED IN THE VESSEL HANGING OVER THE ILIAC BIFURCATION. AN 8 FR SHEATH WAS INSERTED AND MULTIPLE ATTEMPTS TO RETRIEVE THE DEVICE WERE UNSUCCESSFUL. DURING AN INTERIM CONTROL ANGIOGRAPHY, A DISSECTION OF THE RIGHT EXTERNAL ILIAC ARTERY W
Patient problems: Thrombosis/Thrombus; Vascular Dissection
MAUDE report 20645755 ↗MalfunctionSep 23, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED POPLITEAL ARTERY. A 2.5MM X 40MM X 146CM COYOTE ES WS ADVANCED FOR DILATATION. DURING FIRST INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20279619 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135302010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal1 death117 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page