Skip to content

Coyote™ ES H74939135204010Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Coyote ES (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Coyote ES balloon catheter is used to widen narrowed blood vessels. Most of the 586 reports describe the balloon rupturing, often during the first inflation. The reports include 6 injuries and no deaths, and reports fell from 62 in the previous 12 months to 39 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
585
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 585 adverse event reports that list this exact device (DI 08714729767190), received between Aug 31, 2018 and Aug 17, 2026 — classified as 6 injuries, 579 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (561 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 4 reportsJan 2024: 6 reportsFeb 2024: 7 reportsMar 2024: 6 reportsApr 2024: 6 reportsMay 2024: 11 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 4 reportsSep 2024: 8 reportsOct 2024: 12 reportsNov 2024: 4 reportsDec 2024: 4 reportsJan 2025: 5 reportsFeb 2025: 8 reportsMar 2025: 6 reportsApr 2025: 6 reportsMay 2025: 2 reportsJun 2025: 7 reportsJul 2025: 6 reportsAug 2025: 8 reportsSep 2025: 2 reportsOct 2025: 4 reportsNov 2025: 4 reportsDec 2025: 7 reportsJan 2026: 11 reportsFeb 2026: 5 reportsMar 2026: 4 reportsApr 2026: 4 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture561
  • Adverse Event Without Identified Device or Use Problem9
  • Entrapment of Device9
  • Break8
  • Failure to Advance7
  • Material Deformation4
  • Difficult to Advance3
  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions441
  • No Consequences Or Impact To Patient146

Recent adverse event reports

The 25 most recent of 585 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 17, 2026Material Rupture

    It was reported that a balloon rupture occurred.The 99% stenosed target lesion was located in the moderately tortuous and moderately calcified left popliteal artery. A 2mm x 40mm x 145cm Coyote ES MR balloon catheter was advanced for dilation. During the procedure, the balloon ruptured upon first inflation at 12 atmospheres for 10 seconds. The device was removed without any problem, and the procedure was completed with a different device. There were no patient complications as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50274613 ↗
  • MalfunctionAug 11, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50205079 ↗
  • MalfunctionJul 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. DURING ENDOVASCULAR TREATMENT (EVT) OF A CHRONIC TOTAL OCCLUSION (CTO) LESION, A 2.0 MM X 40 MM COYOTE BALLOON CATHETER WAS ADVANCED FOLLOWING GUIDEWIRE PASSAGE. HOWEVER, DUE TO THE HIGHLY CALCIFIED LESION, THE BALLOON RUPTURED DURING THE FIRST INFLATION AT 8 ATMOSPHERES. DILATATION WAS CONTINUED USING A HIGH-PRESSURE (HP) BALLOON OF THE SAME SIZE, FOLLOWED BY A 2.5 MM X 150 MM COYOTE BALLOON CATHETER. DURING THE SECOND INFLATION AT 6 ATMOSPHERES, THE SECOND COYOTE BALLOON ALSO RUPTURED. THE PROCEDURE WAS COMPLETED USING A HIGH-PRESSURE BALLOON OF THE SAME SIZE FOLLOWED BY POST-DILATATION WITH A 3 MM HIGH-PRESSURE BALLOON. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25874044 ↗
  • MalfunctionApr 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED POPLITEAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 14 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24924698 ↗
  • MalfunctionApr 13, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 8 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED FROM A LONGITUDINAL TEAR. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24858791 ↗
  • MalfunctionApr 7, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT 14 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24801736 ↗
  • MalfunctionApr 6, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 30 SECONDS. THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24788739 ↗
  • MalfunctionMar 30, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED BELOW-THE-KNEE VESSEL. A 2MM X 40MM X 145CM COYOTE? ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 14 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24715325 ↗
  • MalfunctionMar 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES BALLOON WAS ADVANCED FOR DILATION. HOWEVER, IT WAS NOTED THAT THE BALLOON RUPTURED DURING FIRST INFLATION AT 8 ATMOSPHERES FOR 3 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM USING THE NORMAL METHOD, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24651986 ↗
  • InjuryMar 17, 2026Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. THIS 2MM X 40MM X 145CM COYOTE ES BALLOON WAS SELECTED FOR USE DURING A TIBIAL DILATION PROCEDURE. DURING THE PROCEDURE, THE BALLOON CATHETER WAS FOUND TO BE SEVERED INSIDE THE INTRODUCER. THE PORTION INSIDE THE INTRODUCER WAS SUCCESSFULLY RETRIEVED USING FORCEPS. THE PROCEDURE WAS THEN COMPLETED USING DIFFERENT DEVICES. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24619676 ↗
  • MalfunctionMar 10, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED CEPHALIC VEIN. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AFTER THE FIRST INFLATION AT 8 ATMOSPHERES FOR 20 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24565319 ↗
  • MalfunctionFeb 25, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ANTERIOR TIBIAL ARTERY. A 2 MM X 40 MM X 145 CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 5 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED VERTICALLY AT ITS CENTER. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24448958 ↗
  • MalfunctionFeb 25, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 40MM X 145CM COYOTE BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT THE NOMINAL PRESSURE FOR 3 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM USING THE NORMAL METHOD, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24445497 ↗
  • MalfunctionFeb 4, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED POPLITEAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 14 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24259734 ↗
  • MalfunctionFeb 3, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 6 ATMOSPHERES FOR 1 SECOND, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251918 ↗
  • MalfunctionFeb 3, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY A 2MM X 40MM X 145CM COYOTE ES WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 6 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED VERTICALLY. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24248802 ↗
  • MalfunctionJan 29, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR TIBIAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES MR BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 10 ATMOSPHERES FOR 10 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24208610 ↗
  • MalfunctionJan 26, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED IN A PINHOLE FORM. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24176529 ↗
  • MalfunctionJan 26, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED IN A PINHOLE FORM. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24176524 ↗
  • MalfunctionJan 26, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 40MM X 145CM COYOTE ES BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE SECOND INFLATION AT 14 ATMOSPHERES FOR 60 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24176344 ↗
  • MalfunctionJan 26, 2026Material Rupture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 40MM X 145CM COYOTE ES BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 15 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24176187 ↗
  • MalfunctionJan 26, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR SEVERAL SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24176184 ↗
  • MalfunctionJan 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 2 SECONDS, THE BALLOON RUPTURED FROM A PINHOLE AT NEAR THE CENTER OF THE BALLOON. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO PATIENT INJURY AS A RESULT OF THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24121697 ↗
  • MalfunctionJan 20, 2026Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED VESSEL BELOW THE KNEE. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 8 ATMOSPHERES, THE BALLOON RUPTURED IN A PINHOLE FORM. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24118417 ↗
  • MalfunctionJan 20, 2026Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT SUPERFICIAL FEMORAL ARTERY. A 2MM X 40MM X 145CM COYOTE ES BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE FIRST INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED VERTICALLY ALONG ITS ENTIRE LENGTH. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24116911 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page