Precision™ SC-1110-02Recalls & reported problems
Boston Scientific Neuromodulation Corporation · Implantable Pulse Generator Kit · Class 3 · Implant
All models of Precision (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Precision spinal cord stimulator is an implanted pulse generator used for pain relief. Of its 6,510 reports, 23 describe deaths and 5,937 describe injuries. Many describe charging problems, batteries that drain early, and pain relief that is not as expected. Recent reports often describe the device not holding a charge and being replaced in a procedure. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 6,519
- Deaths reported
- 26
- Injuries reported
- 5,943
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6,519 adverse event reports that list this exact device (DI 08714729767688), received between Nov 26, 2012 and Sep 29, 2026 — classified as 26 deaths, 5,943 injuries, 550 malfunctions.
The most frequently reported specific device problem is “Charging Problem” (2,846 reports).
17 other recalls cover devices cleared under the same FDA submission (P030017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Charging Problem2,846
- Adverse Event Without Identified Device or Use Problem1,779
- Unexpected Therapeutic Results1,111
- Defective Device552
- Premature Discharge of Battery423
- Failure to Charge224
- Device Inoperable205
- Delayed Charge Time171
Patient problems
- No Code Available2,900
- No Clinical Signs, Symptoms or Conditions1,657
- No Known Impact Or Consequence To Patient838
- Pain529
- Inadequate Pain Relief329
- No Information279
- Discomfort214
- Undesired Nerve Stimulation93
Recent adverse event reports
The 25 most recent of 6,519 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933792 ↗InjurySep 25, 2026Charging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899098 ↗InjurySep 24, 2026Charging Problem
It was reported that the patient was experiencing frequent Implantable Pulse Generator (IPG) charging. The patient underwent an IPG replacement procedure and was doing well postoperatively. The explanted device was not returned.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50878499 ↗InjurySep 22, 2026Charging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50862914 ↗InjurySep 22, 2026Charging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50862906 ↗InjurySep 22, 2026Unexpected Therapeutic Results
It was reported that the patient felt stimulation and burning sensation all around the body. The patient noted that they are supposed to feel stimulation at the back. The patient underwent an Implantable Pulse Generator (IPG) explant procedure.
Patient problems: Burning Sensation; Inappropriate Device Stimulation of Tissue
MAUDE report 50858098 ↗InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the Implantable Pulse Generator (IPG) was no longer functioning as intended. The patient underwent an Implantable Pulse Generator (IPG) replacement procedure and was doing well postoperatively. The explanted device was not returned.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50851442 ↗InjurySep 21, 2026Fracture; Nonstandard Device; Use of Device Problem
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 50848926 ↗InjurySep 21, 2026Nonstandard Device
It was reported that the patient underwent a Spinal Cord Stimulator (SCS) explant procedure due to device never worked as intended. No further information has been obtained despite good faith efforts.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50846483 ↗InjurySep 21, 2026Charging Problem
It was reported that the patient's implanted pulse generator (IPG) was unable to charge. The patient subsequently underwent a surgical procedure. During the procedure, following incision at the battery site, a white discharge was observed, and wound irrigation specimens were obtained. The midline incision was also opened, no discharge was noted at that location. However, the physician observed granulation tissue beneath the currently implanted paddle lead. Given these findings, the physician elected not to implant new equipment and instead explanted the existing system. The physician reportedly did not believe the probable infection was device related and instead suspected an underlying inflammatory condition. Culture was obtained and the result was not yet available.
Patient problems: Fluid Discharge; Inflammation; Unspecified Infection
MAUDE report 50841698 ↗DeathSep 18, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient implanted with a spinal cord stimulation (SCS) system passed away. The cause of death is unknown. No further information has been obtained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50831205 ↗InjurySep 18, 2026Overheating of Device; Unexpected Therapeutic Results
Patient problems: Burning Sensation; Fall; Pain
MAUDE report 50831158 ↗InjurySep 18, 2026Battery Problem
It was reported that the patient's IPG had reached End of Life. The patient underwent a procedure wherein the implantable pulse generator (IPG) and spinal cord stimulator (SCS) leads were explanted. The explanted devices were discarded by the medical facility and will not be returned.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817245 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Stroke/CVA
MAUDE report 50812339 ↗DeathSep 17, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a spinal cord stimulation (SCS) patient passed away. According to the available information, the death was attributed to natural causes; however, the exact cause of death has not been confirmed and it is not known by the physicians office. No additional information has been obtained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50805167 ↗DeathSep 17, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the spinal cord stimulation (SCS) patient passed away. The cause of death is not believed to be related to stimulation; however, it remains unconfirmed. Physicians office does not have a record of patient passing away. No further information has been obtained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803186 ↗InjurySep 17, 2026Nonstandard Device
It was reported that the Implantable Pulse Generator (IPG) was no longer functioning as intended. The patient underwent an Implantable Pulse Generator (IPG) explant procedure. No further information has been obtained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50797840 ↗InjurySep 16, 2026Charging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50795990 ↗InjurySep 16, 2026Nonstandard Device; Adverse Event Without Identified Device or Use Problem; Wireless Communication Problem
It was reported that the patient has difficulty connecting remote control to SCS. It was noted that the patients implantable pulse generator (IPG) was no longer functioning as intended. The patient underwent a spinal cord stimulation (SCS) system replacement procedure. During the revision the physician noticed fraying of the paddle lead. The patient was doing well postoperatively. The explanted device components were kept by the facility.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50789020 ↗InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient was experiencing pain and discomfort. The patient underwent Implantable Pulse Generator (IPG) explant procedure. Patient was doing well postoperatively. Explanted device not available for return, the facility disposed of it.
Patient problems: Discomfort; Pain
MAUDE report 50786404 ↗InjurySep 10, 2026Nonstandard Device
It was reported that the Implantable Pulse Generator (IPG) was no longer functioning as intended after receiving the end of life message. The patient underwent a spinal cord stimulation (SCS) system replacement procedure and was doing well postoperatively. The explanted device was components will not be return per facility policy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50716537 ↗DeathSep 4, 2026Adverse Event Without Identified Device or Use Problem
It was reported that a patient implanted with a spinal cord stimulation (SCS) system passed away. The cause of death is unknown.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50602373 ↗InjurySep 3, 2026Unexpected Therapeutic Results; Migration
Patient problems: Pain
MAUDE report 50599416 ↗InjuryAug 28, 2026Wireless Communication Problem
It was reported that the remote control kept on disconnecting from the device after a few minutes and patient does not like to feel the stimulation. The patient underwent an Implantable Pulse Generator (IPG) replacement procedure and was doing well post operatively. No other information could be obtained.
Patient problems: Inappropriate Device Stimulation of Tissue
MAUDE report 50506265 ↗InjuryAug 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this Spinal Cord Stimulation (SCS) device was explanted due to an unknown reason. The location of the device is unknown. Information is unavailable per customer account.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50421700 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030017. One submission often covers many models, so check the product and lot details in each record.
- Z-2879-2026 · Jun 17, 2026 · Open, ClassifiedINFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spin
- Z-2404-2026 · Apr 29, 2026 · Open, ClassifiedBoston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
- Z-2617-2024 · Jul 17, 2024 · Open, ClassifiedWAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
- Z-2119-2020 · Apr 24, 2019 · TerminatedArtisanTM 70 cm 2 x 8 Surgical Lead Kit, part number M365SC8216700. 2 x 8 Surgical Paddle Lead Kit: (1) One Paddle Lead (4); Four Suture Sleeves; (2) Lead Position Labels, left
- Z-2124-2020 · Apr 24, 2019 · TerminatedFifty-five (55) cm 8 Contact Extension Kit, part number M365SC3138550 - Product Usage: Lead Extensions are designed to connect the percutaneous leads to the Implantable Pulse Gene
- Z-2120-2020 · Apr 24, 2019 · TerminatedThirty-five (35) cm 8 Contact Extension Kit, part number M365SC3138350. Package Contents: Lead Extension Kit (1) Lead Extension (1) Hex Wrench (1) Tunneling Tool Assembly (1) Devic
- Z-2123-2020 · Apr 24, 2019 · TerminatedPrecision TM M8 Adapter, 15 cm, part number M365SC9218150 - Product Usage: The adapter is part of a spinal cord stimulation system for prescription use only indicated as an aid in
- Z-2121-2020 · Apr 24, 2019 · TerminatedLinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's Spinal Cord Stimulation (SCS) systems by delivering electric
- Z-2122-2020 · Apr 24, 2019 · TerminatedThe LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for prescription use only as an aid in the management of chronic i
- Z-1409-2010 · Mar 19, 2010 · TerminatedBionic NAVIGATOR Clinician Programmer Model Number SC-7150-4
- Z-0271-2009 · Sep 22, 2008 · TerminatedPrecision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System labeled as BIONICS SCS-CHARGING KIT CONTENTS Charger (1) Base Station (1) Belt Po
- Z-1441-2012 · Apr 22, 2008 · TerminatedOMG (Observational Mechanical Gateway) Connector - A Model Number SC-9305 The Precision Spinal Cord Stimulator System is indicated as an aid in the management of chronic intractabl
- Z-0960-2008 · Oct 5, 2007 · TerminatedAdvanced Bionics Precision Implantable Pulse Generator Kit, Model Number: SC-1110 with firmware version prior to Revision 3.02, Advanced Bionics Corportation, Sylmar, CA 91342
- Z-0055-2008 · Aug 3, 2007 · TerminatedPrecision Implantable Pulse Generator (IPG), Model Number SC-110
- Z-0700-2007 · Jan 30, 2007 · TerminatedThe Precision Implantable Pulse Generator, Model Number SC-1110
…and 2 more.
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page