Linear™ SC-2158-70Recalls & reported problems

Boston Scientific Neuromodulation Corporation · 70cm 8 Contact Lead Kit · Class 3 · Implant

Quick summary

This is a Boston Scientific Linear spinal cord stimulator lead used for pain relief. There are 23 adverse event reports naming it, with 20 injuries and no deaths. The reports mostly describe leads migrating out of position, high impedance readings, fractures, and patients not getting enough pain relief, often leading to lead replacement or removal procedures. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 08714729767718), received between May 22, 2017 and Jan 19, 2026 — classified as 20 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Unexpected Therapeutic Results” (10 reports).

17 other recalls cover devices cleared under the same FDA submission (P030017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Therapeutic Results10
  • Migration8
  • High impedance7
  • Fracture4
  • Adverse Event Without Identified Device or Use Problem3
  • Bent1
  • Kinked1
  • Migration or Expulsion of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Inadequate Pain Relief6
  • No Code Available4
  • Pain3
  • Erosion1
  • Implant Pain1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 19, 2026Unexpected Therapeutic Results; Migration

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE PAIN RELIEF AND THE LEADS MIGRATED. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICES WERE NOT RETURNED.

    Patient problems: Pain

    MAUDE report 24110367 ↗
  • InjurySep 9, 2025Migration; Migration

    IT WAS REPORTED THAT THE SPINAL CORD STIMULATOR LEADS MIGRATED. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICES WERE NOT RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23005356 ↗
  • InjuryJan 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) LEADS CONTINUALLY KEPT PUSHING THROUGH THE PATIENTS SKIN. THE PHYSICIAN GRABBED SOME SCISSORS AND CLIPPED THE LEAD AT THE SKULL.

    Patient problems: Erosion

    MAUDE report 21114395 ↗
  • InjuryDec 31, 2024High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENTS SPINAL CORD STIMULATION (SCS) LEAD HAD HIGH IMPEDANCES. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE WAS KEPT BY THE FACILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21049071 ↗
  • InjuryJul 24, 2024High impedance; High impedance

    IT WAS REPORTED THAT THE PATIENTS LEADS HAD HIGH IMPEDANCES. THE PATIENT UNDERWENT A LEAD REVISION PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE LEADS REMAIN IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823282 ↗
  • InjuryJul 16, 2024Unexpected Therapeutic Results; Unexpected Therapeutic Results

    IT WAS REPORTED THAT THE LEADS WERE CAUSING PAIN IN THE PATIENTS BACK. IT WAS ALSO NOTED THAT THE SPINAL CORD STIMULATION (SCS) WAS NOT HELPING WITH THE PATIENTS PAIN. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE. ALL EXPLANTED DEVICE COMPONENTS WILL NOT BE RETURNED.

    Patient problems: Implant Pain; Pain

    MAUDE report 19751201 ↗
  • InjuryJun 21, 2024Migration; Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE DUE TO MIGRATION. THE PATIENT WAS DOING WELL POSTOPERATIVELY AND THE EXPLANTED LEADS WERE DISCARDED BY THE MEDICAL FACILITY AND WILL NOT BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19588308 ↗
  • InjuryJun 21, 2024High impedance; Unexpected Therapeutic Results; Migration; High impedance; Unexpected Therapeutic Results; Migration

    IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) PATIENT EXPERIENCED INADEQUATE STIMULATION AND HIGH IMPEDANCE READINGS WERE DISCOVERED ON THE LEAD. ADDITIONALLY, IMAGING THAT WAS TAKEN CONFIRMED THAT THE LEAD HAD MIGRATED. REPROGRAMMING DID NOT RESOLVE THE INADEQUATE STIMULATION. THEREFORE, THE PATIENT UNDERWENT A LEAD REVISION PROCEDURE DURING WHICH THE LEAD WAS EXPLANTED AND TWO NEW LEADS WERE IMPLANTED. THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEAD WILL NOT BE RETURNED AS IT WAS DISCARDED BY THE MEDICAL FACILITY.

    Patient problems: Pain

    MAUDE report 19586786 ↗
  • MalfunctionOct 24, 2022Fracture; High impedance; Fracture; High impedance

    IT WAS REPORTED THAT THE PATIENTS LEAD HAD HIGH IMPEDANCES. THE PATIENTS LEADS WERE EXPLANTED WHEREIN THE LEFT CERVICAL LEADS DISTAL TIP WERE LEFT BEHIND. THE PATIENT WAS DOING WELL POSTOPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15663950 ↗
  • InjurySep 30, 2022Unexpected Therapeutic Results; Unexpected Therapeutic Results

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INADEQUATE STIMULATION DESPITE REPROGRAMMING ATTEMPT. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE OLD LEADS WERE REPLACED WITH A MAGNETIC RESONANCE IMAGING (MRI) CAPABLE ONE AND WERE MOVED HIGHER. THE PATIENT WAS DOING WELL POSTOPERATIVELY AND THE EXPLANTED LEADS WERE DISCARDED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 15514758 ↗
  • InjurySep 6, 2022Unexpected Therapeutic Results; Adverse Event Without Identified Device or Use Problem; Unexpected Therapeutic Results; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN TWO LEADS WERE ADDED FOR AN UNKNOWN REASON.

    Patient problems: Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15366723 ↗
  • InjuryAug 1, 2022Unexpected Therapeutic Results; Unexpected Therapeutic Results

    IT WAS REPORTED THAT THE PATIENTS LEADS IN THE CERVICAL AREA WERE NOT GETTING COVERAGE ON THE LEFT DUE TO LOCATION ON THE FAR RIGHT. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND REPOSITIONED AT THE MIDLINE. THE PATIENT WAS DOING WELL POST-OPERATIVELY. THE EXPLANTED LEADS WILL NOT BE RETURNED PER PHYSICIAN PREFERENCE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 15141964 ↗
  • InjuryJun 15, 2022Unexpected Therapeutic Results; Unexpected Therapeutic Results

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION. NO DEVICE MALFUNCTION SUSPECTED. THE PATIENT UNDERWENT LEAD EXPLANT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE COMPONENTS WILL NOT BE RETURNED AS IT WAS DISCARDED BY THE FACILITY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 14698271 ↗
  • InjuryAug 6, 2021High impedance; Unexpected Therapeutic Results; High impedance; Unexpected Therapeutic Results

    IT WAS REPORTED THAT THE PATIENT WAS NOT RECEIVING ANY STIMULATION DUE TO LEADS WERE NO LONGER WORKING AND HAD HIGH IMPEDANCE. THE PATIENT UNDERWENT LEADS REPLACEMENT PROCEDURE AND WAS DOING GOOD POSTOPERATIVELY. THE EXPLANTED LEADS WERE DISCARDED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 12289016 ↗
  • MalfunctionJul 21, 2021Fracture; Fracture

    IT WAS REPORTED THAT AFTER A NON-DEVICE RELATED SURGERY, THE PATIENTS LEADS HAVE BEEN CUT OR DAMAGED. ADDITIONAL INFORMATION WAS RECEIVED THAT AN ELECTROCAUTERY WAS USED DURING THE NON-DEVICE RELATED SURGERY. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEADS WERE DISCARDED. PER THE PHYSICIANS IMPLANT MANUAL, ELECTROCAUTERY IS A KNOWN SOURCE OF HIGH VOLTAGE TRANSIENT SIGNAL AND WARNS AGAINST THE USE OF ELECTROCAUTERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12201903 ↗
  • InjuryJun 18, 2021Unexpected Therapeutic Results; Migration; Unexpected Therapeutic Results; Migration

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED LOSS OF COVERAGE DUE TO LEAD MIGRATION. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEADS WERE NOT RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12028472 ↗
  • InjuryJun 3, 2021Migration; Migration

    IT WAS REPORTED THAT THE PATIENTS LEADS MIGRATED. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE WILL NOT BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11935927 ↗
  • InjuryFeb 7, 2020Unexpected Therapeutic Results; Migration; Unexpected Therapeutic Results; Migration

    A REPORT WAS RECEIVED THAT THE PATIENT WAS HAVING INADEQUATE STIMULATION DUE TO LEAD MIGRATION. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE.

    Patient problems: No Code Available

    MAUDE report 9684087 ↗
  • InjurySep 6, 2019High impedance; High impedance

    A REPORT WAS RECEIVED THAT THE PATIENTS LEADS HAD HIGH IMPEDANCES. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.

    Patient problems: No Code Available

    MAUDE report 8970299 ↗
  • InjuryMar 29, 2019Migration; Migration

    A REPORT WAS RECEIVED THAT THE PATIENTS LEADS HAD MIGRATED. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.

    Patient problems: No Code Available

    MAUDE report 8465678 ↗
  • InjuryAug 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A REPORT WAS RECEIVED THAT THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE FOR AN UNKNOWN REASON.

    Patient problems: No Code Available

    MAUDE report 7767783 ↗
  • InjuryMay 10, 2018Fracture; Fracture

    A REPORT WAS RECEIVED THAT THE PATIENT HAD A FRACTURED LEAD. THE PATIENT WILL UNDERGO A REVISION PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7505204 ↗
  • MalfunctionMay 22, 2017Bent; Fracture; High impedance; Kinked; Migration or Expulsion of Device; Bent; Fracture; High impedance; Kinked; Migration or Expulsion of Device

    A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING LOST OF COVERAGE. IT WAS ALSO REPORTED THAT THE LEADS HAD MIGRATED AND A LEAD WAS HAVING FOUR CONTACTS OUT. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN TWO LEADS AND CLICK ANCHORS WERE REPLACED. DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT WAS REPORTEDLY DOING WELL POSTOPERATIVELY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 6583469 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030017. One submission often covers many models, so check the product and lot details in each record.

  • Z-2879-2026 · Jun 17, 2026 · Open, Classified
    INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spin
  • Z-2404-2026 · Apr 29, 2026 · Open, Classified
    Boston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
  • Z-2617-2024 · Jul 17, 2024 · Open, Classified
    WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
  • Z-2119-2020 · Apr 24, 2019 · Terminated
    ArtisanTM 70 cm 2 x 8 Surgical Lead Kit, part number M365SC8216700. 2 x 8 Surgical Paddle Lead Kit: (1) One Paddle Lead (4); Four Suture Sleeves; (2) Lead Position Labels, left
  • Z-2121-2020 · Apr 24, 2019 · Terminated
    LinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's Spinal Cord Stimulation (SCS) systems by delivering electric
  • Z-2120-2020 · Apr 24, 2019 · Terminated
    Thirty-five (35) cm 8 Contact Extension Kit, part number M365SC3138350. Package Contents: Lead Extension Kit (1) Lead Extension (1) Hex Wrench (1) Tunneling Tool Assembly (1) Devic
  • Z-2123-2020 · Apr 24, 2019 · Terminated
    Precision TM M8 Adapter, 15 cm, part number M365SC9218150 - Product Usage: The adapter is part of a spinal cord stimulation system for prescription use only indicated as an aid in
  • Z-2122-2020 · Apr 24, 2019 · Terminated
    The LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for prescription use only as an aid in the management of chronic i
  • Z-2124-2020 · Apr 24, 2019 · Terminated
    Fifty-five (55) cm 8 Contact Extension Kit, part number M365SC3138550 - Product Usage: Lead Extensions are designed to connect the percutaneous leads to the Implantable Pulse Gene
  • Z-1409-2010 · Mar 19, 2010 · Terminated
    Bionic NAVIGATOR Clinician Programmer Model Number SC-7150-4
  • Z-0271-2009 · Sep 22, 2008 · Terminated
    Precision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System labeled as BIONICS SCS-CHARGING KIT CONTENTS Charger (1) Base Station (1) Belt Po
  • Z-1441-2012 · Apr 22, 2008 · Terminated
    OMG (Observational Mechanical Gateway) Connector - A Model Number SC-9305 The Precision Spinal Cord Stimulator System is indicated as an aid in the management of chronic intractabl
  • Z-0960-2008 · Oct 5, 2007 · Terminated
    Advanced Bionics Precision Implantable Pulse Generator Kit, Model Number: SC-1110 with firmware version prior to Revision 3.02, Advanced Bionics Corportation, Sylmar, CA 91342
  • Z-0055-2008 · Aug 3, 2007 · Terminated
    Precision Implantable Pulse Generator (IPG), Model Number SC-110
  • Z-0700-2007 · Jan 30, 2007 · Terminated
    The Precision Implantable Pulse Generator, Model Number SC-1110

…and 2 more.

Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices

All STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.