Linear™ ST SC-2218-50Recalls & reported problems
Boston Scientific Neuromodulation Corporation · 50cm 8 Contact Lead Kit · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Linear ST spinal cord stimulator lead is an implanted lead used for pain relief. Its 2,716 reports mostly describe leads moving out of place, stimulation not giving the expected results, and high electrical resistance. The reports include no deaths and 2,523 injuries, and patient problems listed include pain and inadequate pain relief. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 2,729
- Deaths reported
- 0
- Injuries reported
- 2,536
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2,729 adverse event reports that list this exact device (DI 08714729767725), received between Dec 8, 2016 and Sep 29, 2026 — classified as 2,536 injuries, 193 malfunctions.
The most frequently reported specific device problem is “Migration” (1,358 reports).
17 other recalls cover devices cleared under the same FDA submission (P030017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Migration1,358
- Unexpected Therapeutic Results1,341
- High impedance606
- Adverse Event Without Identified Device or Use Problem494
- Fracture175
- Inappropriate/Inadequate Shock/Stimulation48
- Charging Problem34
- Migration or Expulsion of Device32
Patient problems
- Pain748
- No Clinical Signs, Symptoms or Conditions677
- No Code Available615
- Inadequate Pain Relief377
- Undesired Nerve Stimulation219
- Bacterial Infection80
- Discomfort76
- No Known Impact Or Consequence To Patient65
Recent adverse event reports
The 25 most recent of 2,729 reports, in the reporter's own words as published by the FDA.
InjurySep 29, 2026Unexpected Therapeutic Results
It was reported that the patient was not getting good leg coverage along with back coverage. The patient underwent a lead replacement procedure and was doing well postoperatively. The explanted products were discarded per hospital policy.
Patient problems: Pain
MAUDE report 50928371 ↗InjurySep 28, 2026High impedance
It was reported that the Spinal Cord Stimulator (SCS) leads of the patient had impedances. The patient underwent a lead replacement procedure and was doing well postoperatively. The explanted devices were not returned as the facility discarded the devices.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915505 ↗InjurySep 25, 2026Inappropriate/Inadequate Shock/Stimulation; Migration
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50903115 ↗InjurySep 24, 2026High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50892268 ↗InjurySep 24, 2026Unexpected Therapeutic Results; Battery Problem; Migration
Patient problems: Burning Sensation; Numbness; Pain; Paresthesia; Swelling/ Edema
MAUDE report 50886568 ↗InjurySep 22, 2026Unexpected Therapeutic Results; Migration
Patient problems: Pain
MAUDE report 50862904 ↗InjurySep 22, 2026Unexpected Therapeutic Results; Migration
It was reported that the patients stimulation therapy was not adequately targeting the areas requiring pain relief. It was noted that the leads had migrated which was confirmed through imaging. Program optimization was performed, and the Spinal Cord Stimulator (SCS) device was functioning as intended. The patient underwent a procedure wherein an additional SCS lead was implanted. The patient was reported to be doing well postoperatively. The devices remain implanted and in use.
Patient problems: Pain
MAUDE report 50858981 ↗InjurySep 22, 2026Migration
It was reported that the patient's Spinal Cord Stimulation (SCS) leads had moved down. The patient underwent Spinal Cord Stimulator (SCS) replacement procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50858218 ↗InjurySep 22, 2026Packaging Problem
It was reported that the patient became ill and was vomiting the patient was under anesthesia.
Patient problems: Vomiting
MAUDE report 50851480 ↗InjurySep 21, 2026High impedance
It was reported that three lead contacts were out of range. The patient underwent a Spinal Cord Stimulator (SCS) system replacement procedure and was doing well postoperatively. The explanted products were discarded per hospital policy.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50835265 ↗InjurySep 18, 2026Unexpected Therapeutic Results
Patient problems: Pain
MAUDE report 50831457 ↗InjurySep 18, 2026High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50831261 ↗InjurySep 18, 2026Positioning Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50829230 ↗InjurySep 18, 2026Unexpected Therapeutic Results; Adverse Event Without Identified Device or Use Problem
It was reported that the spinal cord stimulation (SCS) patient underwent a revision because patient wasnt getting relief in her new pain areas. The patient was reported to be receiving good coverage of the pain areas postoperatively and was scheduled for follow-up with the local representative. The explanted Montage IPG was discarded by the facility and was not returned for analysis.
Patient problems: Pain
MAUDE report 50819543 ↗InjurySep 17, 2026Charging Problem; Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced pain along midline spine, as well as burning and numbness in the thoracic spine. The patient also reported the need to frequently recharge the Implantable Pulse Generator (IPG) and required Magnetic Resonance Imaging (MRI). Due to the reported symptoms, the patient had pain medication.
Patient problems: Burning Sensation; Numbness; Pain
MAUDE report 50807019 ↗InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this Spinal Cord Stimulation (SCS) device was explanted due to an unknown reason. The location of the replaced device is unknown. Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50790106 ↗MalfunctionSep 16, 2026Migration
It was reported that the patient was experiencing a stimulation in the unwanted areas due to a lead that migrated that was confirmed via X-ray.
Patient problems: Inappropriate Device Stimulation of Tissue
MAUDE report 50784156 ↗InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Erosion
MAUDE report 50779903 ↗InjurySep 15, 2026High impedance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50779699 ↗InjurySep 15, 2026Migration
It was reported that the patients leads had migrated. The patient underwent a revision procedure during which the leads were repositioned to the desired location.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50779645 ↗InjurySep 15, 2026Device Dislodged or Dislocated
It was reported that the physician was having difficulty inserting the lead due to patient's anatomy during a spinal cord stimulation (SCS) implant procedure. There were three most proximal contacts detached from the lead when the physician attempted to insert the lead multiple times. The lead was not implanted to the patient's body and all contacts were removed. The patient was doing well postoperatively and lead was discarded by the medical facility and will not be returned.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50769602 ↗InjurySep 15, 2026Inappropriate/Inadequate Shock/Stimulation; Migration
Patient problems: Discomfort; Inappropriate Device Stimulation of Tissue
MAUDE report 50766031 ↗InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50765711 ↗InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient's legs had given out and experienced several episodes of losing balance over the past few days. The physician indicated that these symptoms might be related to the patient's other health conditions. The patient went to the Emergency Room (ER) for further evaluation, including thoracic and lumbar MRI imaging. The patient was reported to be doing well.
Patient problems: Ambulation or Postural Difficulties; Muscle Weakness/Atrophy
MAUDE report 50753869 ↗InjuryAug 18, 2026Migration
It was reported that one of the patients spinal cord stimulation (SCS) lead had migrated. The patient underwent a lead revision procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50308222 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030017. One submission often covers many models, so check the product and lot details in each record.
- Z-2879-2026 · Jun 17, 2026 · Open, ClassifiedINFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spin
- Z-2404-2026 · Apr 29, 2026 · Open, ClassifiedBoston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
- Z-2617-2024 · Jul 17, 2024 · Open, ClassifiedWAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
- Z-2122-2020 · Apr 24, 2019 · TerminatedThe LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for prescription use only as an aid in the management of chronic i
- Z-2120-2020 · Apr 24, 2019 · TerminatedThirty-five (35) cm 8 Contact Extension Kit, part number M365SC3138350. Package Contents: Lead Extension Kit (1) Lead Extension (1) Hex Wrench (1) Tunneling Tool Assembly (1) Devic
- Z-2121-2020 · Apr 24, 2019 · TerminatedLinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's Spinal Cord Stimulation (SCS) systems by delivering electric
- Z-2124-2020 · Apr 24, 2019 · TerminatedFifty-five (55) cm 8 Contact Extension Kit, part number M365SC3138550 - Product Usage: Lead Extensions are designed to connect the percutaneous leads to the Implantable Pulse Gene
- Z-2119-2020 · Apr 24, 2019 · TerminatedArtisanTM 70 cm 2 x 8 Surgical Lead Kit, part number M365SC8216700. 2 x 8 Surgical Paddle Lead Kit: (1) One Paddle Lead (4); Four Suture Sleeves; (2) Lead Position Labels, left
- Z-2123-2020 · Apr 24, 2019 · TerminatedPrecision TM M8 Adapter, 15 cm, part number M365SC9218150 - Product Usage: The adapter is part of a spinal cord stimulation system for prescription use only indicated as an aid in
- Z-1409-2010 · Mar 19, 2010 · TerminatedBionic NAVIGATOR Clinician Programmer Model Number SC-7150-4
- Z-0271-2009 · Sep 22, 2008 · TerminatedPrecision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System labeled as BIONICS SCS-CHARGING KIT CONTENTS Charger (1) Base Station (1) Belt Po
- Z-1441-2012 · Apr 22, 2008 · TerminatedOMG (Observational Mechanical Gateway) Connector - A Model Number SC-9305 The Precision Spinal Cord Stimulator System is indicated as an aid in the management of chronic intractabl
- Z-0960-2008 · Oct 5, 2007 · TerminatedAdvanced Bionics Precision Implantable Pulse Generator Kit, Model Number: SC-1110 with firmware version prior to Revision 3.02, Advanced Bionics Corportation, Sylmar, CA 91342
- Z-0055-2008 · Aug 3, 2007 · TerminatedPrecision Implantable Pulse Generator (IPG), Model Number SC-110
- Z-0700-2007 · Jan 30, 2007 · TerminatedThe Precision Implantable Pulse Generator, Model Number SC-1110
…and 2 more.
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page