MUSTANG™ H74939171060470Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of MUSTANG (131)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Mustang balloon catheter is a balloon used to widen narrowed arteries. Its 383 reports include 10 injuries and no deaths, and mostly describe the balloon rupturing during inflation. Most reports record no patient harm. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 384
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 384 adverse event reports that list this exact device (DI 08714729793335), received between Aug 24, 2018 and Sep 21, 2026 — classified as 10 injuries, 374 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (345 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture345
- Inflation Problem13
- Adverse Event Without Identified Device or Use Problem9
- Leak/Splash9
- Difficult to Remove7
- Entrapment of Device7
- Fluid/Blood Leak4
- Material Integrity Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions264
- No Consequences Or Impact To Patient113
- No Patient Involvement2
- Thrombosis/Thrombus2
- Device Embedded In Tissue or Plaque1
- Hematoma1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 384 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Material Rupture
It was reported that balloon burst occurred.A 6.0 x 40, 75cm Mustang balloon catheter was advanced for dilation. During the procedure, the pressure was within the normal range, but the balloon burst. The procedure was completed with another of the same device. There were no patient complications as a result of this event, and the patient was stable post-procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50836669 ↗MalfunctionSep 18, 2026Material Rupture
It was reported that balloon rupture occurred.The 99% stenosed target lesion was located in the moderately tortuous and severely calcified shunt. A 6.0 x 40, 75cm Mustang balloon catheter was advanced for dilation. During the procedure, the balloon ruptured upon first inflation at 10 atmospheres for 2 seconds. The device was removed without any problem using the normal method, and the procedure was completed with a different device. There were no patient complications as a result of this event, and the patient was in good condition post-procedure.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50821780 ↗MalfunctionAug 17, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50278476 ↗MalfunctionAug 17, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50274623 ↗MalfunctionAug 17, 2026Material Rupture
It was reported that a balloon rupture occurred.The 99% stenosed target lesion was located in the moderately tortuous and moderately calcified forearm vessel. A 6.0 x 40, 75cm Mustang balloon catheter was advanced for dilation. During the procedure, the balloon ruptured upon first inflation at 10 atmospheres for 5 seconds. The device was removed without any problem, and the procedure was completed with a different device. There were no patient complications as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50274605 ↗MalfunctionAug 17, 2026Material Rupture
It was reported that balloon rupture occurred.The 100% stenosed target lesion was located in the severely calcified superficial femoral artery. A 6.0 x 40, 75cm Mustang balloon catheter was advanced for dilatation. During the fourth inflation at approximately nominal pressure, after three prior inflations at rated pressure for 30 seconds each, the balloon ruptured with a pinhole defect at the junction between the proximal shoulder and the balloon. The balloon was removed without difficulty, and no fragments remained. The procedure was completed with another device of the same model. No patient complications were reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50270991 ↗MalfunctionAug 14, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50255880 ↗MalfunctionJul 28, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50053789 ↗MalfunctionJul 20, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25872453 ↗MalfunctionJul 19, 2026Difficult to Advance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25868096 ↗MalfunctionJul 12, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED AT 6 ATMOSPHERES. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25791357 ↗MalfunctionJun 29, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ILIAC ARTERY. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE THIRD INFLATION AT 24 ATM FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25661276 ↗MalfunctionJun 22, 2026Inflation Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 04JUN2026. IT WAS REPORTED THAT BALLOON FAILED TO INFLATE. THE SEVERELY STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED ARTERIOVENOUS FISTULA. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE THIRD INFLATION AT 23 ATMOSPHERES, THE BALLOON DID NOT FULLY EXPAND WELL IN THE LESION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED A BALLOON PINHOLE LOCATED APPROXIMATELY 1MM PROXIMAL FROM THE PROXIMAL MARKERBAND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25580129 ↗MalfunctionJun 4, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED ILIAC ARTERY. A 6.0 X 40, 75CM MUSTANG BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 20 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM USING THE NORMAL METHOD, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25392671 ↗MalfunctionJun 3, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED LEFT HAND SHUNT VESSEL. DURING SHUNT PTA, AFTER THE WIRE PASSED THROUGH THE LESION, A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE INITIAL INFLATION AT 8 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25373949 ↗MalfunctionMay 28, 2026Inflation Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 06MAY2026. IT WAS REPORTED THAT FAILURE TO INFLATE OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED CEPHALIC VEIN AND BRACHIAL VEIN. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON WAS UNABLE TO BE INFLATED ON THE SECOND INFLATION AT THE LESION AT 10 ATMOSPHERES FOR 5 SECONDS. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. THE PATIENT CONDITION FOLLOWING PROCEDURE WAS STABLE. HOWEVER. RETURNED DEVICE ANALYSIS REVEALED LONGITUDINAL TEAR IN THE BALLOON MATERIAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25315435 ↗MalfunctionMay 18, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING FOURTH INFLATION AT 20 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25203277 ↗MalfunctionMay 13, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY (SFA). A 6.0 X 40, 75CM BALLOON MUSTANG CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 24 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25160554 ↗MalfunctionMay 11, 2026Fluid/Blood Leak; Material Rupture
IT WAS REPORTED THAT BALLOON TEAR OCCURRED. THE 68% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 6.0 X 40, 75CM MUSTANG BALLOON WAS SELECTED FOR USE. HOWEVER, DURING THE FIRST DILATION, A TEAR IN THE BALLOON WAS NOTED, AND THE BALLOON WAS LEAKING LIQUID. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS STABLE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25134686 ↗MalfunctionMay 11, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON PINHOLE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND MODERATELY CALCIFIED ARTERIOVENOUS FISTULA. HOWEVER, THE BALLOON FAILED TO INFLATE DURING THE PROCEDURE AND A BALLOON PINHOLE WAS NOTED. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25134454 ↗MalfunctionApr 27, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED INTERNAL SHUNT. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 10 ATMOSPHERES FOR FEW SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24995554 ↗MalfunctionApr 8, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE DYNAMIC AND STATIC FISTULA. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON WAS BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24821500 ↗MalfunctionApr 2, 2026Inflation Problem
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED VESSEL IN THE INTERNAL SHUNT FOREARM. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, FOLLOWING FLUSHING AND AIR ASPIRATION, INFLATION WAS ATTEMPTED; HOWEVER, THE INFLATION PRESSURE DID NOT INCREASE. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24769089 ↗MalfunctionApr 1, 2026Fluid/Blood Leak; Inflation Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 12MAR2026. IT WAS REPORTED THAT BALLOON LEAKAGE AND INFLATION ISSUE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED CEPHALIC VEIN AND RADIAL ARTERY. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON WAS PRESSURIZED TO 12 ATMOSPHERES, AND IT WAS FOUND THAT THE BALLOON WAS LEAKING FLUID AND COULD NOT BE EXPANDED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED A LONGITUDINAL TEAR IN THE BALLOON MATERIAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24758127 ↗MalfunctionApr 1, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT SUPERFICIAL FEMORAL ARTERY. A 6.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 10 ATMOSPHERES FOR 20 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24758118 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
- MUSTANG™ H74939171060440Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC202 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page