MUSTANG™ H74939171050470Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of MUSTANG (131)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific MUSTANG balloon catheter, used to widen narrowed blood vessels, has 212 adverse event reports, with no deaths and 5 injuries. Most describe the balloon bursting during inflation, and some say another catheter was used to finish the procedure. Many entries report no patient complications. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 212
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 212 adverse event reports that list this exact device (DI 08714729793366), received between Mar 27, 2017 and Jul 13, 2026 — classified as 5 injuries, 207 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (195 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture195
- Leak/Splash7
- Inflation Problem6
- Difficult to Remove4
- Deflation Problem3
- Fluid/Blood Leak3
- Detachment of Device or Device Component2
- Entrapment of Device2
Patient problems
- No Clinical Signs, Symptoms or Conditions133
- No Consequences Or Impact To Patient77
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 212 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 13, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED LEFT FOREARM. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 15 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM USING THE NORMAL METHOD AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25807885 ↗MalfunctionJun 8, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE ILIAC ARTERY. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIRST INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25436532 ↗MalfunctionApr 13, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS TRANSLUMINAL BALLOON DILATATION OF AN ARTIFICIAL ARTERIOVENOUS FISTULA IN THE LEFT FOREARM UNDER LOCAL ANESTHESIA WITH B-MODE ULTRASOUND GUIDANCE. VASCULAR ACCESS WAS SUCCESSFULLY OBTAINED UNDER ULTRASOUND GUIDANCE. SEVERE VASCULAR STENOSIS WAS NOTED. A 5.0 X 40 MM, 75 CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING INFLATION, THE BALLOON RUPTURED PRIOR TO REACHING THE RATED BURST PRESSURE. THE DEVICE WAS REMOVED AND REPLACED WITH ANOTHER BALLOON OF THE SAME SPECIFICATION, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT, AND THE PATIENT REMAINED STABLE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865551 ↗MalfunctionApr 6, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED SHUNT. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE FIFTH INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24788737 ↗MalfunctionApr 5, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE DYNAMIC AND STATIC FISTULA. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 80% STENOSED, MODERATELY TORTUOUS AND NON-CALCIFIED. THE BALLOON RUPTURED UPON FIRST INFLATION AT 15 ATMOSPHERES FOR 60 SECONDS. THE DEVICE WAS SIMPLY PULLED OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24787043 ↗MalfunctionApr 2, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE ARTERIOVENOUS FISTULA. VASCULAR ACCESS WAS VIA CEPHALIC VEIN PUNCTURE. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING INFLATION, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24763917 ↗MalfunctionApr 1, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LEFT EXTERNAL ILIAC ARTERY (ITEIA). A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 8 ATMOSPHERES FOR 15 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24758133 ↗MalfunctionApr 1, 2026Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE ARTERIOVENOUS FISTULA. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24757283 ↗MalfunctionMar 31, 2026Fluid/Blood Leak; Inflation Problem; Nonstandard Device; Material Rupture
IT WAS REPORTED THAT THE BALLOON COULD NOT INFLATE AND LEAKING LIQUID DUE TO DAMAGE. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED LOWER UPPER LIMB ARTERIOVENOUS FISTULA. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION, IT WAS NOTED THAT THE BALLOON WAS LEAKING LIQUID DUE TO DAMAGE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS STABLE POST-PROCEDURE. IT WAS FURTHER REPORTED THAT THE LIQUID LEAKAGE WAS DUE TO A BALLOON RUPTURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24746708 ↗MalfunctionMar 30, 2026Inflation Problem
REPORTABLE BASED ON THE DEVICE ANALYSIS COMPLETED ON 4MAR2026. IT WAS REPORTED THAT BALLOON FAILURE TO INFLATE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LESION. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON DID NOT INFLATE PROPERLY. THE PROCEDURE WAS THEN COMPLETE USING WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATION AS A RESULT OF THIS EVENT. HOWEVER, COMPLETED DEVICE ANALYSIS REVEALED A LONGITUDINAL TEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24737246 ↗MalfunctionFeb 25, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED ARM ARTERIOVENOUS SHUNT. A 5.0 X 40, 75CM MUSTANG BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION, THE BALLOON BURST BEFORE IT WAS INFLATED TO THE NOMINAL BURST PRESSURE. THE DEVICE WAS REMOVED SLOWLY AND CAREFULLY, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS STABLE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24445063 ↗MalfunctionJan 26, 2026Fluid/Blood Leak; Inflation Problem
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 05JAN2026. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ARTERIOVENOUS FISTULA. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER ADVANCED FOR DILATION. HOWEVER, DURING DILATION, THE BALLOON DID NOT EXPAND, AND CONTRAST AGENT LEAKED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED A LONGITUDINAL TEAR WAS IDENTIFIED IN THE BALLOON MATERIAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24176183 ↗MalfunctionJan 15, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 5.0 X 40, 75CM MUSTANG BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES FOR 3 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24082273 ↗MalfunctionJan 14, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED VESSEL IN THE LEFT FOREARM. A 5.0 X 40, 75CM MUSTANG BALLOON WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 20 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24070298 ↗MalfunctionJan 14, 2026Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE CEPHALIC VEIN. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURE DURING THE SECOND INFLATION AT 12 ATM. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT 99% IS THE STENOSED TARGET LESION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24065944 ↗MalfunctionJan 8, 2026Fluid/Blood Leak; Material Rupture
IT WAS REPORTED THAT BALLOON HOLE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS BALLOON DILATATION OF ARTIFICIAL ARTERIOVENOUS FISTULA IN THE LEFT FOREARM UNDER LOCAL ANESTHESIA AND B-MODE ULTRASOUND LOCALIZATION. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, IT WAS NOTED THAT THE THERE WAS A HOLE AND GAS LEAKAGE IN THE BALLOON BODY. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24003509 ↗MalfunctionDec 25, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ILIAC ARTERY. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING SECOND INFLATION AT 20 ATMOSPHERES FOR 30 SECONDS, THE BALLOON VERTICALLY RUPTURED AT THE MIDDLE OF THE BALLOON. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION WERE GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23893049 ↗MalfunctionNov 21, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY TORTUOUS AND MODERATELY CALCIFIED VESSEL IN THE LEFT FOREARM. A 5.0 X 40 MM, 75 CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE THIRD INFLATION AT 24 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT DIFFICULTY, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23618074 ↗MalfunctionNov 10, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED INTERNAL SHUNT. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS SELECTED FOR USE. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 10 ATMOSPHERES FOR FEW SECOND. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23518726 ↗MalfunctionOct 21, 2025Inflation Problem; Material Rupture; Inflation Problem; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED ARTERIOVENOUS FISTULA. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON SECOND INFLATION AT 22 ATMOSPHERES FOR 40 SECONDS. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23353940 ↗MalfunctionOct 6, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT FOREARM VESSEL. A 5.0 X 40, 75CM MUSTANG WAS SELECTED FOR USE IN A SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY. DURING THE FIRST INFLATION AT 14 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23223583 ↗MalfunctionSep 16, 2025Leak/Splash; Material Rupture; Leak/Splash; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND MODERATELY CALCIFIED VESSEL. A 5.0 X 40, 75 CM MUSTANG BALLOON CATHETER WAS SELECTED FOR USE. DURING INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. A DAMAGE LOCATED AT MARKED POINT 1CM WAS NOTED. THE DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED A. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23063623 ↗MalfunctionAug 25, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 8 ATMOSPHERES FOR 2 SECONDS, THE BALLOON RUPTURED NEAR THE BALLOON SHOULDER. THE DEVICE WAS REMOVED, AND A PINHOLE WAS NOTED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22882538 ↗MalfunctionAug 5, 2025Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 11JUL2025. IT WAS REPORTED THAT BALLOON FAILED TO INFLATE AND LEAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE ARTERIOVENOUS FISTULA (AVF). A 5.0 X 40, 75CM MUSTANG WAS SELECTED FOR PROCEDURE. DURING THE PROCEDURE, IT WAS NOTED THAT CONTRAST MEDIA WAS LEAKING FROM THE CATHETER TIP AND THE BALLOON WAS DEFECTIVE AS IT COULD NOT EXPAND. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A BALLOON TEAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22722654 ↗MalfunctionJun 24, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 75% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ABDOMINAL AORTA. A 5.0 X 40, 75CM MUSTANG BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING SECOND INFLATION AT 20 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED AT THE MIDDLE PART OF THE BALLOON. THE DEVICE WAS REMOVED, AND A PINHOLE WAS NOTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22319435 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
- MUSTANG™ H74939171060440Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC202 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page