Boston Scientific Neuromodulation Corporation · Implantable Pulse Generator and Charger Kit · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Precision SC-1010C is an implanted pulse generator and charger kit used for spinal cord stimulation to relieve chronic pain. The 46 reports mostly describe charging difficulties, units that would no longer charge or charge only partly, and generators that were removed or replaced during surgery; some patients reported inadequate pain relief. Two reports describe deaths and 37 describe injuries. No recalls name this device. Reports in the last 12 months (4) are down from 11 in the prior12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
2
Injuries reported
36
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 08714729821519), received between Dec 2, 2016 and Aug 11, 2026 — classified as 2 deaths, 36 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Charging Problem” (14 reports).

17 other recalls cover devices cleared under the same FDA submission (P030017); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem21
  • Charging Problem14
  • Unexpected Therapeutic Results5
  • Failure to Charge3
  • Device Inoperable2
  • Defective Device1
  • Delayed Charge Time1
  • High impedance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions18
  • No Information7
  • No Known Impact Or Consequence To Patient7
  • Inadequate Pain Relief6
  • No Code Available5
  • Pain4
  • Death1
  • Discomfort1

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • DeathAug 11, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the spinal cord stimulation (SCS) patient passed away. It was stated that the cause was not device related, however, the specific cause is not known. No further information was obtained.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50201467 ↗
  • InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE IMPLANTABLE PULSE GENERATOR (IPG) IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742497 ↗
  • InjuryNov 27, 2025Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) PATIENT UNDERWENT AN EXPLANT PROCEDURE WHEREIN THE IMPLANTABLE PULSE GENERATOR (IPG) WAS REMOVED FOR AN UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23667624 ↗
  • InjuryOct 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICE IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23438915 ↗
  • InjuryOct 2, 2025Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WOULD NO LONGER CHARGE. THE PATIENT UNDERWENT AN IMPLANTABLE PULSE GENERATOR (IPG) REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE WAS DISCARDED BY THE MEDICAL FACILITY. NO DEVICE MALFUNCTION WAS SUSPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23198185 ↗
  • InjurySep 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE IMPLANTABLE PULSE GENERATOR (IPG) WAS REVISED FOR AN UNKNOWN REASON.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23139957 ↗
  • InjuryJul 24, 2025Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WOULD NOT FULLY CHARGE DESPITE TROUBLESHOOTING ATTEMPTS. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE AND THERE WERE NO REPORTED COMPLICATIONS POSTOPERATIVELY. THE EXPLANTED DEVICE WAS KEPT BY THE MEDICAL FACILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22602334 ↗
  • InjuryJul 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THESE SPINAL CORD STIMULATION (SCS) DEVICES WERE EXPLANTED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICES IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22470778 ↗
  • InjuryJul 8, 2025Charging Problem; Impedance Problem; Charging Problem; Impedance Problem

    IT WAS REPORTED THAT THE PATIENT HAD DIFFICULTY CHARGING THE IMPLANTABLE PULSE GENERATOR (IPG) AND THE LEADS HAD IMPEDANCES. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR (SCS) REPLACEMENT PROCEDURE, WAS DOING WELL POSTOPERATIVELY AND THE EXPLANTED DEVICES WERE DISPOSED BY THE FACILITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22440824 ↗
  • InjuryJun 9, 2025Charging Problem; Charging Problem

    IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO CHARGING DIFFICULTIES AND MAGNETIC RESONANCE IMAGING (MRI) PREFERENCES. THE PATIENT UNDERWENT AN IMPLANTABLE PULSE GENERATOR (IPG) REVISION PROCEDURE WHEREIN THEIR IPG WAS EXPLANTED AND REPLACED. THE IPG WAS DISPOSED BY THE FACILITY AND WILL NOT BE RETURNED FOR ANALYSIS. POSTOPERATIVELY, THE PATIENT WAS DOING GREAT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22172428 ↗
  • InjuryJun 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THESE SPINAL CORD STIMULATION (SCS) IMPLANTABLE PULSE GENERATOR (IPG) AND LEAD WERE REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICES IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22162604 ↗
  • InjuryMay 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED BECAUSE THE PATIENT HAD NOT USED THEIR STIMULATOR FOR YEARS, THOUGH THE UNDERLYING REASON REMAINS UNKNOWN. ADDITIONALLY, THE PATIENT REQUIRED A BATTERY UPGRADE FOR MAGNETIC RESONANCE IMAGING (MRI) COMPATIBILITY AND UNDERWENT AN IMPLANTABLE PULSE GENERATOR (IPG) REVISION PROCEDURE, DURING WHICH THEIR IPG WAS EXPLANTED AND REPLACED. THE IPG WILL NOT BE RETURNED FOR ANALYSIS. THE PATIENT IS DOING WELL POST-OPERATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22111132 ↗
  • InjuryMar 24, 2025Charging Problem; Charging Problem

    IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO CHARGING DIFFICULTIES. THE PATIENT UNDERWENT AN IMPLANTABLE PULSE GENERATOR (IPG) REVISION PROCEDURE WHEREIN THEIR IPG WAS EXPLANTED AND REPLACED. THE IPG WAS DISPOSED BY THE HOSPITAL AND WILL NOT BE RETURNED FOR ANALYSIS. THE PATIENT IS DOING WELL POST OPERATIVELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21683480 ↗
  • InjuryFeb 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EXPLANTED DUE TO UNKNOWN REASON. THE PATIENT IS HAPPY AND DOING GREAT POST OPERATIVELY. THE EXPLANTED DEVICE WILL NOT BE RETURNED AS IT WAS DISPOSED PER FACILITY POLICY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21292812 ↗
  • InjuryOct 15, 2024Premature Discharge of Battery; Delayed Charge Time; Premature Discharge of Battery; Delayed Charge Time

    IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) SYSTEM WAS REPLACED DUE TO THE PATIENT COMPLAINED OF IMPLANTABLE PULSE GENERATOR (IPG) DEPLETING QUICKLY AND TAKING LONGER TO CHARGE. THE PATIENT UNDERWENT AN IPG REPLACEMENT, AND HE HAS GREAT COVERAGE OVER HIS PAIN AREAS. EXPLANTED DEVICE WILL NOT RETURN DUE TO FACILITY POLICY AND ELECTROCAUTERY WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20452731 ↗
  • InjuryAug 20, 2024Charging Problem; Adverse Event Without Identified Device or Use Problem; Charging Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN EXPLANT PROCEDURE DUE TO AN UNKNOWN REASON. THE EXPLANTED DEVICE WILL NOT BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20008210 ↗
  • InjuryAug 22, 2023Unexpected Therapeutic Results; Charging Problem; Unexpected Therapeutic Results; Charging Problem

    IT WAS REPORTED THAT THE PATIENT HAD DIFFICULTY CHARGING THE IPG. IT WAS NOTED ALSO THAT PATIENT EXPERIENCED INADEQUATE STIMULATION. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE AND WAS DOING WELL POST OPERATIVELY. THE EXPLANTED DEVICE WILL NOT BE RETURNED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 17594061 ↗
  • InjuryApr 14, 2023High impedance; Charging Problem; Wireless Communication Problem; High impedance; Charging Problem; Wireless Communication Problem

    IT WAS REPORTED THAT THE PATIENT HAD DIFFICULTY CHARGING THE IPG. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE COMPONENTS WILL NOT BE RETURNED AS IT WAS DISCARDED BY THE FACILITY. ADDITIONAL INFORMATION RECEIVED THAT ALL CONTACTS ON PATIENTS LEAD WERE OUT. IT WAS NOTED ALSO THAT ONLY ONE LEAD WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16744230 ↗
  • InjuryMar 16, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN EXPLANT PROCEDURE FOR AN UNKNOWN REASON. NO FURTHER INFORMATION COULD BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11497074 ↗
  • InjuryAug 25, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING PAIN AND WAS HOSPITALIZED. IT WAS UNKNOWN IF MEDICAL OR SURGICAL INTERVENTION WAS GIVEN.

    Patient problems: Pain

    MAUDE report 10449791 ↗
  • InjuryOct 11, 2019Defective Device; Defective Device

    A REPORT WAS RECEIVED THAT THE PATIENTS IPG WAS NO LONGER WORKING. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE. NO FURTHER INFORMATION COULD BE OBTAINED.

    Patient problems: No Code Available

    MAUDE report 9180151 ↗
  • InjuryJul 22, 2019Unexpected Therapeutic Results; Unexpected Therapeutic Results

    A REPORT WAS RECEIVED THAT THE PATIENT WAS HAVING PAIN OVER THE IPG SITE. IT WAS ALSO MENTIONED THAT THE SCS WAS NOT PROVIDING ADEQUATE RELIEF AND WAS CAUSING INCREASED PAIN AND JOLTING. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE.

    Patient problems: No Code Available; Pain; Undesired Nerve Stimulation

    MAUDE report 8814325 ↗
  • InjuryJun 19, 2019Unexpected Therapeutic Results; Unexpected Therapeutic Results

    A REPORT AS RECEIVED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE.

    Patient problems: No Code Available

    MAUDE report 8713866 ↗
  • InjuryJun 17, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A REPORT WAS RECEIVED THAT THE PATIENT UNDERWENT AN EXPLANT PROCEDURE FOR AN UNKNOWN REASON.

    Patient problems: No Code Available

    MAUDE report 8705571 ↗
  • InjuryJun 13, 2019Unexpected Therapeutic Results; Charging Problem; Unexpected Therapeutic Results; Charging Problem

    A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION AND IPG FREQUENT CHARGING. NO DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.

    Patient problems: No Code Available

    MAUDE report 8697465 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030017. One submission often covers many models, so check the product and lot details in each record.

  • Z-2879-2026 · Jun 17, 2026 · Open, Classified
    INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spin
  • Z-2404-2026 · Apr 29, 2026 · Open, Classified
    Boston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
  • Z-2617-2024 · Jul 17, 2024 · Open, Classified
    WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
  • Z-2121-2020 · Apr 24, 2019 · Terminated
    LinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's Spinal Cord Stimulation (SCS) systems by delivering electric
  • Z-2124-2020 · Apr 24, 2019 · Terminated
    Fifty-five (55) cm 8 Contact Extension Kit, part number M365SC3138550 - Product Usage: Lead Extensions are designed to connect the percutaneous leads to the Implantable Pulse Gene
  • Z-2120-2020 · Apr 24, 2019 · Terminated
    Thirty-five (35) cm 8 Contact Extension Kit, part number M365SC3138350. Package Contents: Lead Extension Kit (1) Lead Extension (1) Hex Wrench (1) Tunneling Tool Assembly (1) Devic
  • Z-2122-2020 · Apr 24, 2019 · Terminated
    The LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for prescription use only as an aid in the management of chronic i
  • Z-2119-2020 · Apr 24, 2019 · Terminated
    ArtisanTM 70 cm 2 x 8 Surgical Lead Kit, part number M365SC8216700. 2 x 8 Surgical Paddle Lead Kit: (1) One Paddle Lead (4); Four Suture Sleeves; (2) Lead Position Labels, left
  • Z-2123-2020 · Apr 24, 2019 · Terminated
    Precision TM M8 Adapter, 15 cm, part number M365SC9218150 - Product Usage: The adapter is part of a spinal cord stimulation system for prescription use only indicated as an aid in
  • Z-1409-2010 · Mar 19, 2010 · Terminated
    Bionic NAVIGATOR Clinician Programmer Model Number SC-7150-4
  • Z-0271-2009 · Sep 22, 2008 · Terminated
    Precision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System labeled as BIONICS SCS-CHARGING KIT CONTENTS Charger (1) Base Station (1) Belt Po
  • Z-1441-2012 · Apr 22, 2008 · Terminated
    OMG (Observational Mechanical Gateway) Connector - A Model Number SC-9305 The Precision Spinal Cord Stimulator System is indicated as an aid in the management of chronic intractabl
  • Z-0960-2008 · Oct 5, 2007 · Terminated
    Advanced Bionics Precision Implantable Pulse Generator Kit, Model Number: SC-1110 with firmware version prior to Revision 3.02, Advanced Bionics Corportation, Sylmar, CA 91342
  • Z-0055-2008 · Aug 3, 2007 · Terminated
    Precision Implantable Pulse Generator (IPG), Model Number SC-110
  • Z-0700-2007 · Jan 30, 2007 · Terminated
    The Precision Implantable Pulse Generator, Model Number SC-1110

…and 2 more.