Skip to content

Navigator HD M0062502280Recalls & reported problems

Boston Scientific Corporation · Ureteral Access Sheath Set · Class 2

All models of Navigator HD (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729839088), received between Nov 30, 2017 and Oct 2, 2023 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated2
  • Material Deformation1
  • Material Integrity Problem1
  • Material Split, Cut or Torn1
  • Torn Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • Perforation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 2, 2023Material Deformation; Material Split, Cut or Torn; Material Deformation; Material Split, Cut or Torn

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR ACCESS SHEATH DEVICE WAS USED IN THE KIDNEY DURING A TRANSURETHRAL LITHOTRIPSY (TUL) PROCEDURE PERFORMED ON (B)(6), 2023. DURING PREPARATION, THE TIP OF THE DILATOR WAS DAMAGED AND BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR ACCESS SHEATH DEVICE. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF SHEATH FRACTURED. PLEASE SEE DEVICE ANALYSIS RESULTS FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17854645 ↗
  • MalfunctionAug 9, 2021Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR ACCESS SHEATH WAS USED DURING A TRANSURETHRAL LITHOTRIPSY (TUL) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN PACKAGING WAS UNPACKED TO PREPARE THE DEVICE, COATING WAS FOUND TO BE DEFECTIVE. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR ACCESS SHEATH. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON AUGUST 10, 2021: IT WAS NOTED THAT THE OUTER PART OF THE SHEATH PEELED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12294520 ↗
  • InjuryNov 13, 2018Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR HD ACCESS SHEATH WAS USED IN THE URINARY BLADDER DURING A CYSTOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SURGEON NOTICED THAT THE END OF THE ACCESS SHEATH FEELS ROUGH. REPORTEDLY, THE ACCESS SHEATH POSSIBLY CAUSED A LACERATION/TEAR IN THE PATIENTS URETER. THE LACERATION WAS STITCHED AND MONITORED AFTER.

    Patient problems: Perforation

    MAUDE report 8066186 ↗
  • MalfunctionJun 28, 2018Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR HD ACCESS SHEATH WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, IT WAS NOTED THAT THE COATING OF THE SHEATH PEELED OFF FROM THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR HD ACCESS SHEATH. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7645650 ↗
  • MalfunctionNov 30, 2017Torn Material; Torn Material

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR HD ACCESS SHEATH WAS USED IN THE UPPER URETER DURING A URINARY TRACT STONE REMOVAL PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SHEATH INNER COATING FROM THE TIP WAS PEELED OFF. IT WAS REPORTED THAT NO PARTS OF THE DEVICE FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR HD ACCESS SHEATH. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS SHOWED THAT THE INNER LINER OF THE SHEATH PEELED OFF; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7070388 ↗

Other endoscopic access overtube, gastroenterology-urology devices

All endoscopic access overtube, gastroenterology-urology devices →

Sources and updates