Navigator HD M0062502280Recalls & reported problems
Boston Scientific Corporation · Ureteral Access Sheath Set · Class 2
All models of Navigator HD (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729839088), received between Nov 30, 2017 and Oct 2, 2023 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated2
- Material Deformation1
- Material Integrity Problem1
- Material Split, Cut or Torn1
- Torn Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Perforation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 2, 2023Material Deformation; Material Split, Cut or Torn; Material Deformation; Material Split, Cut or Torn
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR ACCESS SHEATH DEVICE WAS USED IN THE KIDNEY DURING A TRANSURETHRAL LITHOTRIPSY (TUL) PROCEDURE PERFORMED ON (B)(6), 2023. DURING PREPARATION, THE TIP OF THE DILATOR WAS DAMAGED AND BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR ACCESS SHEATH DEVICE. THERE WERE NO PATIENT COMPLICATION REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF SHEATH FRACTURED. PLEASE SEE DEVICE ANALYSIS RESULTS FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17854645 ↗MalfunctionAug 9, 2021Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR ACCESS SHEATH WAS USED DURING A TRANSURETHRAL LITHOTRIPSY (TUL) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN PACKAGING WAS UNPACKED TO PREPARE THE DEVICE, COATING WAS FOUND TO BE DEFECTIVE. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR ACCESS SHEATH. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON AUGUST 10, 2021: IT WAS NOTED THAT THE OUTER PART OF THE SHEATH PEELED OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12294520 ↗InjuryNov 13, 2018Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR HD ACCESS SHEATH WAS USED IN THE URINARY BLADDER DURING A CYSTOSCOPY PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SURGEON NOTICED THAT THE END OF THE ACCESS SHEATH FEELS ROUGH. REPORTEDLY, THE ACCESS SHEATH POSSIBLY CAUSED A LACERATION/TEAR IN THE PATIENTS URETER. THE LACERATION WAS STITCHED AND MONITORED AFTER.
Patient problems: Perforation
MAUDE report 8066186 ↗MalfunctionJun 28, 2018Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR HD ACCESS SHEATH WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, IT WAS NOTED THAT THE COATING OF THE SHEATH PEELED OFF FROM THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR HD ACCESS SHEATH. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7645650 ↗MalfunctionNov 30, 2017Torn Material; Torn Material
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIGATOR HD ACCESS SHEATH WAS USED IN THE UPPER URETER DURING A URINARY TRACT STONE REMOVAL PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SHEATH INNER COATING FROM THE TIP WAS PEELED OFF. IT WAS REPORTED THAT NO PARTS OF THE DEVICE FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER NAVIGATOR HD ACCESS SHEATH. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS SHOWED THAT THE INNER LINER OF THE SHEATH PEELED OFF; THEREFORE, THIS IS NOW AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7070388 ↗
Other endoscopic access overtube, gastroenterology-urology devices
- EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE OLYMPUS SIF TYPE Q180Olympus Medical Systems Corp. · endoscopic access overtube, gastroenterology-urology1 recall42 reports
- Trokamed WA2PS20STROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall2 deaths2 reports
- Trokamed WA2PS20LTROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall2 deaths2 reports
- Trokamed WA2PS18LTROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall0 reports
- Trokamed WA2PS18STROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall0 reports
- Navigator HD M0062502220Boston Scientific Corporation · endoscopic access overtube, gastroenterology-urology1 death31 reports
- Navigator HD M0062502230Boston Scientific Corporation · endoscopic access overtube, gastroenterology-urology29 reports
- Proxis™ Ureteral Access Sheath 231045C. R. Bard, Inc. · endoscopic access overtube, gastroenterology-urology22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.