Expect Slimline SL M00555500Recalls & reported problems

Boston Scientific Corporation · Endoscopic Ultrasound Aspiration Needle · Class 2

All models of Expect Slimline SL (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729848141), received between Mar 2, 2017 and Feb 3, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Difficult to Advance2
  • Appropriate Device Problem Term/Code Not Available1
  • Device-Device Incompatibility1
  • Material Deformation1
  • Material Perforation1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Hemorrhage/Blood Loss/Bleeding1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 3, 2026Use of Device Problem; Device-Device Incompatibility; Difficult to Advance

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EXPECT SLIMLINE NEEDLE WAS USED DURING A PROCEDURE TO INSERT A COIL TO CONTROL A BLEEDING GASTRIC VARIX ON (B)(6) 2025. DURING PROCEDURE, THE EXPECT SLIMLINE NEEDLE WAS INSERTED IN THE VARIX, THE AREA WAS VERY MOBILE DUE TO RESPIRATIONS AND HICCUPPING. THE STYLET WAS REMOVED FROM THE NEEDLE; A COIL WAS PLACED IN THE NEEDLE AND THE STYLET WAS ATTEMPTED TO BE ADVANCED TO PUSH THE COIL INTO THE VARIX AND STOP THE BLEEDING. THE COIL WOULD ONLY ADVANCE TO ABOUT THE HALFWAY POINT AND BECAME STUCK THE SCOPE WAS ATTEMPTED TO BE REPOSITIONED BUT WAS UNSUCCESSFUL. A GASTROSCOPE WAS PUT UP TO GLUE THE VARIX UNDER ENDOSCOPIC VIEW, BUT THE VARIX STARTED BLEEDING AND THE PATIENT BECAME UNSTABLE. A CODE 66 WAS CALLED AND RESUSCITATION WAS INITIATED; THE PATIENT WAS INTUBATED AFTER THE SCOPE WAS REMOVED. THE PATIENT WAS ADMITTED TO ICU WHERE THE VARIX WAS INJECTED WITH GLUE TO STOP THE BLEEDING. ANOTHER EXPECT SLIMLINE NEEDLE WAS OPENED OUTSIDE THE PATIENT AND A COIL WAS ATTEMPTED BE ADVANCED THROUGH THE NEEDLE AND IT ALSO BECAME STUCK. ACCORDING TO THE INDICATION FOR USE SET BY THE MANUFACTURER, THE DEVICE IS INTENDED TO BE USED FOR THE DEL

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246542 ↗
  • MalfunctionOct 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EXPECT SLIMLINE NEEDLE WAS USED IN THE LIVER DURING AN ENDOSCOPIC ULTRASOUND (EUS) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE NEEDLE SEPARATED FROM THE SHAFT. THE DISTAL TIP DID NOT DETACH INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20421082 ↗
  • InjuryOct 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EXPECT SLIMLINE (SL) WAS USED DURING AN ENDOSCOPIC ULTRASOUND (EUS) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE NEEDLE DETACHED ITSELF FROM THE HANDLE. THE NEEDLE WAS SUCCESSFULLY REMOVED WITHOUT DAMAGING THE SCOPE OR INJURY TO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20346608 ↗
  • MalfunctionApr 20, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EXPECT SLIMLINE SL NEEDLE WAS USED IN A PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE NEEDLE DETACHED FROM THE DEVICE. A PHOTO WAS PROVIDED BY THE CUSTOMER SHOWING THAT THE DISTAL END OF THE NEEDLE DETACHED. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14164157 ↗
  • MalfunctionSep 17, 2021Difficult to Advance; Material Deformation; Appropriate Device Problem Term/Code Not Available; Difficult to Advance; Material Deformation; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EXPECT SLIMLINE NEEDLE WAS USED DURING AN ENDOSCOPIC GUIDED BILIARY DRAINAGE PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE NEEDLE COULD NOT ADVANCE THROUGH THE ENDOSCOPE. THE NEEDLE WAS WITHDRAWN AND IT WAS NOTED THAT THE WORKING LENGTH WAS BENT. THE ENDOSCOPE CHANNEL WAS DAMAGED AND THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12487983 ↗
  • MalfunctionMar 2, 2017Material Perforation; Material Perforation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EXPECT¿ NEEDLE DEVICE WAS USED IN THE PANCREATIC DURING AN ENDOSCOPIC ULTRASOUND FINE NEEDLE ASPIRATION (EUS FNA) PROCEDURE PERFORMED ON UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE THE COMPLAINANT REPORTED THE NEEDLE WAS UNABLE TO ADVANCE AND ADJUST THE HANDLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THE INVESTIGATION FOUND THAT THE NEEDLE PUNCTURED THROUGH THE SHEATH AND THE DISTAL TIP WAS BENT, THIS IS NOW DEEMED AN MDR REPORTABLE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6372818 ↗