Skip to content

DiRex™ Steerable Sheath M004DS200Recalls & reported problems

Oscor Inc. · Steerable sheath, 8.5F · Class 2

All models of DiRex Steerable Sheath (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729858928), received between Sep 3, 2021 and Mar 26, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak4
  • Difficult to Flush1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THE SHEATH WAS BACK BLEEDING UPON INSERTION. PATIENT PRESENT AT TIME OF EVENT. PROCEDURE COMPLETED USING DIFFERENT DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21700133 ↗
  • MalfunctionMar 6, 2023Fluid/Blood Leak; Fluid/Blood Leak

    DURING PROCEDURE DIREX SHEATH WAS LEAKING OUT OF THE FLUSH TUBING AT THE SITE OF HANDLE. THERE WAS NO PATIENT COMPLICATION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16491772 ↗
  • MalfunctionJan 27, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED, LAB STAFF NOTIFIED EP REP THAT DIREX STEERABLE SHEATH HEMOSTASIS VALUE WAS LEAKING DURING PROCEDURE. PHYSICIAN REPLACED SHEATH WITH ANOTHER DIREX WHICH RESOLVED THE ISSUE. THE LOT NUMBER IS UNKNOWN AND THE DEVICE WAS DISPOSED OF ON SITE. NO PATIENT HEALTH CONSEQUENCES OR IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13366030 ↗
  • MalfunctionDec 14, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT THE END OF THE SHEATH WAS LEAKING DURING AN ABLATION PROCEDURE, THERE WERE NO PATIENT COMPLICATIONS REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13001280 ↗
  • MalfunctionSep 3, 2021Difficult to Flush; Difficult to Flush

    IT WAS REPORTED THAT A DIREX STEERABLE SHEATH WAS USED IN A PROCEDURE. HOWEVER IT WAS NOTED THAT DURING THE PROCEDURE A UNSPECIFIED SHEATH COMPLICATION OCCURRED DURING THE PROCEDURE. IT WAS FURTHER REPORTED ON 21JUN2021 THAT DURING THE CASE A SHEATH COMPLICATION OCCURRED WHILE THE PATIENT WAS ON THE TABLE; TWO OF THE SHEATHS PORTS WERE NOT ABLE TO FLUSH. THE THIRD SHEATH WAS USED. NO KNOWN PATIENT COMPLICATIONS. NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12421798 ↗

FDA clearance and classification

Cleared through 510(k) K120459 on Jun 29, 2012 (decision: Substantially Equivalent). Submitted by Oscor, Inc. as “Steerable Guiding Sheath, Model Adelante Destino”. The FDA review took 135 days. FDA record ↗

The FDA regulates product code DYB (introducer, catheter) under 21 CFR 870.1340 as a Class II device. 21 CFR 870.1340 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the DiRex™ Steerable Sheath M004DS200

Has the DiRex™ Steerable Sheath M004DS200 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729858928), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the DiRex™ Steerable Sheath M004DS200?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Sep 3, 2021 and Mar 26, 2025. In those reports, the most common specific device problems are “Fluid/Blood Leak” (4) and “Difficult to Flush” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the DiRex™ Steerable Sheath M004DS200?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the DiRex™ Steerable Sheath M004DS200 FDA approved?

It was cleared through 510(k) K120459 on Jun 29, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the DiRex™ Steerable Sheath M004DS200, and is it still sold?

The FDA's device database lists Oscor Inc. as the company responsible for it (brand DiRex™ Steerable Sheath, model M004DS200). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page