Skip to content

DiRex™ Steerable Sheath M004DS300Recalls & reported problems

Oscor Inc. · Steerable sheath, 8.5F · Class 2

All models of DiRex Steerable Sheath (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729858942), received between Nov 11, 2019 and Sep 13, 2021 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak4
  • Break1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 13, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT DIREX STEERABLE GUIDING SHEATH WAS LEAKING THROUGH VALVE. THERE WAS NO KNOWN ADVERSE PATIENT SIDE EFFECTS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12463855 ↗
  • MalfunctionDec 2, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THERE WAS A BUBBLE LEAKAGE FROM INSIDE THE INTRODUCER FROM THE INFERIOR VALVE. THE ISSUE OCCURRED DURING AN ATRIAL FIBRILLATION PROCEDURE. NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10934440 ↗
  • MalfunctionOct 26, 2020Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING PROCEDURE HEMOSTATIC VALVE OF STEERABLE GUIDING SHEATH WAS LEAKING. THERE WAS NO KNOWN PATIENT SIDE EFFECTS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10734645 ↗
  • MalfunctionApr 2, 2020Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT DURING A PROCEDURE THERE WAS BLOOD LEAKING FROM THE CATHETER HANDLE. THE SHEATH WAS IRRIGATED WITH AN UNDER PRESSURE SALINE BAG. DUE TO THE LEAK AND BLOOD LOSS THE SHEATH HAD TO BE CHANGED. THE AMOUNT OF BLOOD LOSS IS UNKNOWN. NO SURGERY OR BLOOD PRODUCT NEEDED TO CONTROL THE BLOOD LOSS. THE ISSUE WAS RESOLVED BY REPLACING THE SHEATH. THERE WAS NO PATIENT INJURY REPORTED. NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9918097 ↗
  • MalfunctionNov 11, 2019Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    IT WAS REPORTED THAT: AT THE UNIKLINIK REGENSBURG WE HAD TWO DIREX SHEATHS WITH MALFUNCTIONS. THE HEMOSTATIC SEAL OF THE FIRST SHEATH WAS BROKEN AND THEREFORE THE DILATOR COULD NOT BE ADVANCED INTO THE SHEATH AND WAS THEREFORE EXCHANGED. THE SECOND SHEATH HAD A BROKEN STEERING AND ONLY ONE SIDE OF THE CURVE WAS OPERABLE THIS RESULTED IN ANOTHER SHEATH EXCHANGE. THE THIRD ONE WORKED WITHOUT ANY FURTHER ISSUES AND THE CASE WAS COMPLETED. BOTH SHEATHS ARE FROM THE SAME LOT. IT WAS REPORTED NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED, INCLUDING THE EVENT DATE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9306792 ↗

FDA clearance and classification

Cleared through 510(k) K120459 on Jun 29, 2012 (decision: Substantially Equivalent). Submitted by Oscor, Inc. as “Steerable Guiding Sheath, Model Adelante Destino”. The FDA review took 135 days. FDA record ↗

The FDA regulates product code DYB (introducer, catheter) under 21 CFR 870.1340 as a Class II device. 21 CFR 870.1340 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the DiRex™ Steerable Sheath M004DS300

Has the DiRex™ Steerable Sheath M004DS300 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729858942), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the DiRex™ Steerable Sheath M004DS300?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Nov 11, 2019 and Sep 13, 2021. In those reports, the most common specific device problems are “Fluid/Blood Leak” (4), “Break” (1) and “Leak/Splash” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the DiRex™ Steerable Sheath M004DS300?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the DiRex™ Steerable Sheath M004DS300 FDA approved?

It was cleared through 510(k) K120459 on Jun 29, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the DiRex™ Steerable Sheath M004DS300, and is it still sold?

The FDA's device database lists Oscor Inc. as the company responsible for it (brand DiRex™ Steerable Sheath, model M004DS300). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page