DiRex™ Steerable Sheath M004DS300Recalls & reported problems
Oscor Inc. · Steerable sheath, 8.5F · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729858942), received between Nov 11, 2019 and Sep 13, 2021 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak4
- Break1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 13, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT DIREX STEERABLE GUIDING SHEATH WAS LEAKING THROUGH VALVE. THERE WAS NO KNOWN ADVERSE PATIENT SIDE EFFECTS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12463855 ↗MalfunctionDec 2, 2020Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THERE WAS A BUBBLE LEAKAGE FROM INSIDE THE INTRODUCER FROM THE INFERIOR VALVE. THE ISSUE OCCURRED DURING AN ATRIAL FIBRILLATION PROCEDURE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10934440 ↗MalfunctionOct 26, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING PROCEDURE HEMOSTATIC VALVE OF STEERABLE GUIDING SHEATH WAS LEAKING. THERE WAS NO KNOWN PATIENT SIDE EFFECTS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10734645 ↗MalfunctionApr 2, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT DURING A PROCEDURE THERE WAS BLOOD LEAKING FROM THE CATHETER HANDLE. THE SHEATH WAS IRRIGATED WITH AN UNDER PRESSURE SALINE BAG. DUE TO THE LEAK AND BLOOD LOSS THE SHEATH HAD TO BE CHANGED. THE AMOUNT OF BLOOD LOSS IS UNKNOWN. NO SURGERY OR BLOOD PRODUCT NEEDED TO CONTROL THE BLOOD LOSS. THE ISSUE WAS RESOLVED BY REPLACING THE SHEATH. THERE WAS NO PATIENT INJURY REPORTED. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9918097 ↗MalfunctionNov 11, 2019Break; Fluid/Blood Leak; Break; Fluid/Blood Leak
IT WAS REPORTED THAT: AT THE UNIKLINIK REGENSBURG WE HAD TWO DIREX SHEATHS WITH MALFUNCTIONS. THE HEMOSTATIC SEAL OF THE FIRST SHEATH WAS BROKEN AND THEREFORE THE DILATOR COULD NOT BE ADVANCED INTO THE SHEATH AND WAS THEREFORE EXCHANGED. THE SECOND SHEATH HAD A BROKEN STEERING AND ONLY ONE SIDE OF THE CURVE WAS OPERABLE THIS RESULTED IN ANOTHER SHEATH EXCHANGE. THE THIRD ONE WORKED WITHOUT ANY FURTHER ISSUES AND THE CASE WAS COMPLETED. BOTH SHEATHS ARE FROM THE SAME LOT. IT WAS REPORTED NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED, INCLUDING THE EVENT DATE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9306792 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.