EndoVive Safety PEG Kit M00509030Recalls & reported problems
Boston Scientific Corporation · Percutaneous Endoscopic Gastrostomy Kit ENFit · Class 2
All models of EndoVive Safety PEG Kit (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 08714729880448), received between Aug 3, 2022 and Mar 17, 2025 — classified as 3 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component12
- Adverse Event Without Identified Device or Use Problem1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, A PEG TUBE WAS INSERTED OVER THE GUIDEWIRE THROUGH THE MOUTH AND ADVANCED UNTIL THE END OF THE TUBE WAS VISIBLE AT THE PUNCTURE SITE ON THE ANTERIOR ABDOMINAL WALL. TRACTION WAS THEN APPLIED TO FACILITATE THE TUBE PASSAGE THROUGH THE ABDOMINAL WALL. WHEN PASSING THE CONNECTOR THROUGH THE ABDOMINAL WALL, THE TWO PORTIONS OF THE TUBE BECAME SEPARATED, ONE PART INSIDE THE PATIENT AND THE OTHER OUTSIDE THE PATIENT. THE GUIDEWIRE WAS IN PLACE ALLOWING BOTH THE EXTERNAL AND INTERNAL PORTIONS TO BE RETRIEVED. THE PROCEDURE WAS COMPLETED USING A NON-BOSTON SCIENTIFIC DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21627926 ↗MalfunctionOct 29, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, WHILE PLACING THE PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) DEVICE, IT SEPARATED AT THE TRANSITION JOINT. A PHOTO OF THE DEVICE OUTSIDE THE PATIENT WAS PROVIDED BY THE CUSTOMER AND SHOWED THE TRANSITION JOINT BETWEEN THE PEG TUBE AND THE INTRODUCER DILATOR WAS DETACHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20559676 ↗MalfunctionOct 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED IN THE STOMACH DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE SILICONE PLASTIC JUNCTION WAS DETACHED. A PHOTO OF THE DEVICE WAS PROVIDED BY THE CUSTOMER AND SHOWED THAT THE TRANSITION JOINT BETWEEN PEG TUBE AND INTRODUCER DILATOR WAS DETACHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20545565 ↗MalfunctionOct 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NOTE: THIS REPORT PERTAINS TO THE FIRST OF TWO ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE PEG TUBE DISASSOCIATED AT THE TRANSITION ZONE. A SECOND PEG KIT WAS OPENED AND ATTEMPTED TO BE USED BUT, THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20519058 ↗MalfunctionOct 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NOTE: THIS REPORT PERTAINS TO THE SECOND OF TWO ENDOVIVE SAFETY PEG KITS PUSH METHOD THAT WERE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE PEG TUBE DISASSOCIATED AT THE TRANSITION ZONE. A SECOND PEG KIT WAS OPENED AND ATTEMPTED TO BE USED BUT, THE SAME PROBLEM OCCURRED. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20518173 ↗MalfunctionOct 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE PHYSICIAN REPORTED THAT THE PEG TUBE DETACHED AT THE TRANSITION ZONE WHERE THE HARD PLASTIC MEETS THE SILICONE TUBING. THE PROCEDURE WAS COMPLETED WITH A SECOND ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20385302 ↗MalfunctionOct 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE PHYSICIAN REPORTED THAT THE PEG TUBE DETACHED AT THE TRANSITION ZONE WHERE THE HARD PLASTIC MEETS THE SILICONE TUBING. THE PROCEDURE WAS COMPLETED WITH A SECOND ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20385183 ↗MalfunctionSep 9, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED IN THE STOMACH DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, WHEN IT WAS ATTEMPTED TO PULL THE INTRODUCER PART OF THE PUSH TYPE PEG IT SEPARATES OFF OF THE SILICONE TUBE. A PHOTO WAS PROVIDED BY THE CUSTOMER AND SHOWED THAT THE TRANSITION JOINT BETWEEN THE PEG TUBE AND THE INTRODUCER DILATOR WAS SEPARATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20173727 ↗MalfunctionAug 8, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS ATTEMPTED TO BE USED IN THE GASTRIC DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE PEG TUBE BROKE UPON PULLING. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19940775 ↗MalfunctionFeb 1, 2024Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT PASS THROUGH THE TRANSITION ZONE BETWEEN THE CONNECTOR, BETWEEN THE HARD AND SOFT PLASTIC. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18623510 ↗InjuryJun 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. IT WAS REPORTED THAT, DURING DEVICE PLACEMENT, THE PEG TUBE TORE APART. REPORTEDLY, AIR LEAKED INTO THE PERITONEUM REQUIRING AN ANGIOCATH TO BE INSERTED TO RELEASE THAT AIR. THE PROCEDURE WAS COMPLETED WITH A NON-BSC DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17040235 ↗InjuryApr 28, 2023Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY WITH PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PLACEMENT PROCEDURE ON (B)(6) 2023. IT WAS REPORTED THAT, DURING DEVICE PLACEMENT, THE PEG TUBE BROKE OFF. THE DETACHED PORTION WAS SURGICALLY RETRIEVED. THE PROCEDURE WAS COMPLETED WITH A NEW ENDOVIVE STANDARD PEG KIT PUSH METHOD. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16831990 ↗InjuryAug 3, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE SAFETY PEG KIT PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PLACEMENT PROCEDURE. THE PROCEDURE DATE IS UNKNOWN. DURING THE PROCEDURE, THE PEG TUBE DETACHED INSIDE THE STOMA SITE. THEY WERE ABLE TO RETRIEVE THE TUBE. THE PROCEDURE WAS COMPLETE A NEW ENDOVIVE SAFETY PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15160528 ↗
Other Gastrointestinal Tubes with Enteral Specific Connectors devices
- EndoVive Safety PEG Kit M00509000Boston Scientific Corporation · Gastrointestinal Tubes with Enteral Specific Connectors1 recall32 reports
- EndoVive Safety PEG Kit M00509010Boston Scientific Corporation · Gastrointestinal Tubes with Enteral Specific Connectors1 recall16 reports
- EndoVive Standard PEG Kit M00509050Boston Scientific Corporation · Gastrointestinal Tubes with Enteral Specific Connectors1 recall6 reports
- Salem Sump 8888264986ECardinal Health 200, LLC · Gastrointestinal Tubes with Enteral Specific Connectors1 recall1 death2 reports
- Salem Sump 8888268060ECardinal Health 200, LLC · Gastrointestinal Tubes with Enteral Specific Connectors1 recall0 reports
- Salem Sump 7771810ECardinal Health, Inc. · Gastrointestinal Tubes with Enteral Specific Connectors28 reports
- Salem Sump 7771610ECardinal Health, Inc. · Gastrointestinal Tubes with Enteral Specific Connectors1 death18 reports
- Edi Catheter ENFit 6Fr/49 cm, 5pcs 6883898Maquet Critical Care AB · Gastrointestinal Tubes With Enteral Specific Connectors14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page