EndoVive Standard PEG Kit M00509070Recalls & reported problems
Boston Scientific Corporation · Percutaneous Endoscopic Gastrostomy Kit ENFit · Class 2
All models of EndoVive Standard PEG Kit (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729880523), received between Jan 26, 2022 and Jan 13, 2026 — classified as 4 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Difficult to Advance” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Advance4
- Deformation Due to Compressive Stress3
- Detachment of Device or Device Component3
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryJan 13, 2026Deformation Due to Compressive Stress; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS USED IN THE STOMACH DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, WHEN ADVANCING THE PEG TUBE OVER WIRE, THE PLASTIC KINKS WHEN TRYING TO PUSH THROUGH INCISION SITE. TO COMPLETE THE PROCEDURE, THE TRACT WAS DILATED THEN THE DEVICE WAS ADVANCE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24064210 ↗InjuryJan 13, 2026Deformation Due to Compressive Stress; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS USED IN THE STOMACH DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, WHEN ADVANCING THE PEG TUBE OVER WIRE, THE PLASTIC KINKS WHEN TRYING TO PUSH THROUGH INCISION SITE. TO COMPLETE THE PROCEDURE, THE TRACT WAS DILATED THEN THE DEVICE WAS ADVANCE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24064207 ↗InjuryJan 9, 2026Deformation Due to Compressive Stress; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS USED IN THE STOMACH DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WHEN ADVANCING THE PEG TUBE OVER WIRE, THE PLASTIC KINKS WHEN TRYING TO PUSH THROUGH INCISION SITE. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THE INSERTION TUBE DEVICE WITH SEVERAL KINKS. TO COMPLETE THE PROCEDURE, THE TRACT WAS DILATED THEN THE DEVICE WAS ADVANCE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24027614 ↗MalfunctionFeb 20, 2025Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation
NOTE: THIS REPORT PERTAINS TO ONE OF TWO ENDOVIVE STANDARD PEG KIT PUSH METHODS USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS USED IN THE STOMACH DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE FIRST PEG TUBE CORKSCREWED AT THE BOTTOM AND THE PHYSICIAN WAS UNABLE TO ADVANCE THE TUBE ALONG THE WIRE AND INTO THE PATIENT. THE 2ND PEG TUBE BROKE IN HALF WHEN THE PHYSICIAN WAS PULLING IT THROUGH THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENDOVIVE STANDARD PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21422547 ↗MalfunctionFeb 19, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NOTE: THIS REPORT PERTAINS TO ONE OF TWO ENDOVIVE STANDARD PEG KIT PUSH METHODS USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS USED IN THE STOMACH DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY (PEG) PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE FIRST PEG TUBE CORKSCREWED AT THE BOTTOM AND THE PHYSICIAN WAS UNABLE TO ADVANCE THE TUBE ALONG THE WIRE AND INTO THE PATIENT. THE 2ND PEG TUBE BROKE IN HALF WHEN THE PHYSICIAN WAS PULLING IT THROUGH THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER ENDOVIVE STANDARD PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21417350 ↗MalfunctionJul 7, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS USED DURING A PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PLACEMENT PROCEDURE ON (B)(6) 2022. DURING PREPARATION, THE PEG TUBE DETACHED. THE PROCEDURE WAS COMPLETED WITH A NEW ENDOVIVE STANDARD PEG KIT PUSH METHOD. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14944176 ↗InjuryJan 26, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ENDOVIVE STANDARD PEG KIT PUSH METHOD WAS USED DURING AN ESOPHAGOGASTRODUODENOSCOPY WITH PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PLACEMENT PROCEDURE ON (B)(6) 2022. DURING THE PROCEDURE, THE PEG TUBE DETACHED AT THE TRANSITION ZONE BETWEEN THE HARD AND SOFT PLASTIC IN THE GASTRIC WALL AND STOMA INCISION. THE PROCEDURE WAS NOT COMPLETED, AND PATIENT WAS SENT FOR EXPLORATORY LAPAROTOMY SURGERY. IT WAS REPORTED THAT PATIENT WAS ADMITTED TO HOSPITAL BEYOND THE STANDARD OF CARE AND WAS EXPECTED TO FULLY RECOVER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13359849 ↗
Other Gastrointestinal Tubes with Enteral Specific Connectors devices
- EndoVive Safety PEG Kit M00509000Boston Scientific Corporation · Gastrointestinal Tubes with Enteral Specific Connectors1 recall32 reports
- EndoVive Safety PEG Kit M00509010Boston Scientific Corporation · Gastrointestinal Tubes with Enteral Specific Connectors1 recall16 reports
- EndoVive Standard PEG Kit M00509050Boston Scientific Corporation · Gastrointestinal Tubes with Enteral Specific Connectors1 recall6 reports
- Salem Sump 8888264986ECardinal Health 200, LLC · Gastrointestinal Tubes with Enteral Specific Connectors1 recall1 death2 reports
- Salem Sump 8888268060ECardinal Health 200, LLC · Gastrointestinal Tubes with Enteral Specific Connectors1 recall0 reports
- Salem Sump 7771810ECardinal Health, Inc. · Gastrointestinal Tubes with Enteral Specific Connectors28 reports
- Salem Sump 7771610ECardinal Health, Inc. · Gastrointestinal Tubes with Enteral Specific Connectors1 death18 reports
- Edi Catheter ENFit 6Fr/49 cm, 5pcs 6883898Maquet Critical Care AB · Gastrointestinal Tubes With Enteral Specific Connectors14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page