Samurai™ H74939302190J0Recalls & reported problems
Boston Scientific Corporation · Guidewire · Class 2
All models of Samurai (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729887836), received between Sep 4, 2019 and Jul 22, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult to Remove2
- Material Deformation2
- Detachment of Device or Device Component1
Patient problems
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 22, 2022Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED THAT THE DISTAL TIP OF THE GUIDEWIRE DETACHED AND REMAINED IN THE PATIENT. A 190CM SAMURAI GUIDEWIRE WAS SELECTED FOR USE IN A PERCUTANEOUS CORONARY INTERVENTION (PCI). THE TARGET LESION WAS 40% STENOSED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED OSTIUM OF A DIAGONAL BRANCH OF THE LEFT ANTERIOR DESCENDING ARTERY (LAD).THE GUIDEWIRE WAS USED TO DELIVER A STENT AND IT BECAME STUCK IN THE CORONARY AFTER THE STENT DEPLOYED AND RETRIEVAL WAS ATTEMPTED. THE PHYSICIAN TRIED TO PUSH THE DEVICE DISTALLY IN THE VESSEL BUT MOVEMENT WAS NOT SUCCESSFUL. THE DEVICE WAS ATTEMPTED TO BE PULLED, HOWEVER RESISTANCE WAS NOTED, AND A BALLOON WAS ADVANCED IN ORDER TO ATTEMPT TO MOVE THE DEVICE, WITHOUT SUCCESS. A FURTHER ATTEMPT WAS MADE TO PULL HARDER, HOWEVER, THE GUIDEWIRE BROKE ALONG THE RADIOPAQUE PORTION, AT AROUND 3 CM FROM THE TIP. THE PROXIMAL PART OF THE GUIDEWIRE WAS PULLED OUT, AND REMOVED SUCCESSFULLY. THE DISTAL PORTION OF THE GUIDEWIRE WAS LEFT INSIDE THE CORONARY OF THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN ADDITIONAL STENT AND NON-BOSTON SCIENTIFIC GUIDEWIRE. AN ADDITIONAL STENT WAS PLACED IN THE LEFT MAIN IN ORDER TO KEEP THE DETACHED PORTION O
Patient problems: Foreign Body In Patient
MAUDE report 15085588 ↗MalfunctionNov 18, 2019Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation
IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCUNTERED. A 190CM, J-TIP SAMURAI GUIDEWIRE WAS ADVANCED TO CROSS THE LESION. HOWEVER, IT WAS NOTED THAT THE GUIDEWIRE WAS ENTANGLED AND WAS TRAPPED WITHIN THE LESION. IT WAS NOT SPECIFIED HOW THE GUIDEWIRE WAS REMOVED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9335475 ↗MalfunctionSep 4, 2019Material Deformation; Material Deformation
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 21-AUG-2019. IT WAS REPORTED THAT GUIDE WIRE DAMAGE OCCURRED. A 190CM SAMURAI GUIDE WIRE WAS SELECTED FOR USE; HOWEVER, IT WAS NOTED THAT THE GUIDE WIRE WAS DAMAGED DURING INTRODUCTION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED WIRE CORE DETACHMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8962179 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page