SamuraiRecalls & reported problems across all models
Boston Scientific Corporation · 3 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 43
- Deaths reported
- 1
- Injuries reported
- 16
- Recalls
- 0
We found 3 models of Samurai by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 43 adverse event reports in the FDA's MAUDE database, received between Jan 11, 2017 and Oct 22, 2025 — classified as 1 death, 16 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Samurai problems
Device problems
- Detachment of Device or Device Component14
- Difficult to Remove11
- Entrapment of Device8
- Fracture7
- Material Deformation4
- Material Integrity Problem4
- Adverse Event Without Identified Device or Use Problem3
- Difficult to Advance3
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Consequences Or Impact To Patient11
- Foreign Body In Patient8
- Device Embedded In Tissue or Plaque5
- Chest Pain2
- Death1
- No Known Impact Or Consequence To Patient1
Samurai models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Samurai™ H749393021900WIRE, GUIDE, CATHETER1 death38 reports
- Samurai™ H74939302190J0WIRE, GUIDE, CATHETER3 reports
- Samurai™ H749393023000WIRE, GUIDE, CATHETER2 reports
Recent Samurai adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryOct 22, 2025Fracture; Difficult to Remove; Fracture; Difficult to Remove
IT WAS REPORTED THAT A GUIDEWIRE FRACTURE AND REMOVAL DIFFICULTY OCCURRED. THE DIFFUSELY STENOSED TARGET LESION OF LESS THAN 30% WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT CIRCUMFLEX ARTERY. AFTER WITHDRAWAL OF A BALLOON, THE 190CM STRAIGHT-TIP SAMURAI GUIDEWIRE WAS NOTED TO BE ABNORMAL AND COULD NOT BE WITHDRAWN. A MICROCATHETER WAS INTRODUCED, BUT THE GUIDEWIRE WAS FRACTURED AND COULD NOT BE REMOVED. THE DEVICE FRAGMENT REMAINS INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE AND THE PATIENT WAS STABLE.
Patient problems: Foreign Body In Patient
MAUDE report 23360676 ↗MalfunctionAug 7, 2025Contamination; Device Contamination with Chemical or Other Material; Contamination; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT A FOREIGN OBJECT WAS PRESENT WHICH LEADS TO A DEVICE CONTAMINATION. DURING PREPARATION AND PRIOR TO OPENING THE PACKAGE OF A 190 CM SAMURAI GUIDEWIRE, A FOREIGN OBJECT WAS OBSERVED INSIDE THE PACKAGING AND COMPROMISED THE STERILITY OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22745604 ↗MalfunctionJun 5, 2025Material Deformation; Material Deformation
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 12MAY2025. IT WAS REPORTED THAT A GUIDEWIRE KINKED OCCURRED. A PERCUTANEOUS INTERVENTION PROCEDURE WAS BEING PERFORMED. A 190CM SAMURAI GUIDEWIRE WAS USED. HOWEVER, DURING PREPARATION, IT WAS NOTICED THAT THE GUIDEWIRE WAS ALREADY KINKED OR BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITHOUT ANY PATIENT COMPLICATIONS REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A DISTAL TIP AND CORE WIRE DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22150939 ↗MalfunctionMay 12, 2025Stretched; Stretched
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 17-APR-2025. IT WAS REPORTED THAT GUIDEWIRE TIP STRETCHED WAS ENCOUNTERED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 190CM SAMURAI GUIDEWIRE WAS ADVANCED AND THE GUIDEWIRE WAS FOUND TO BE STRETCHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS STABLE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED CORE WIRE DETACHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22003680 ↗MalfunctionNov 1, 2024Fracture; Fracture
IT WAS REPORTED THAT A GUIDEWIRE FRACTURE OCCURRED. THE 83% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING (LAD) ARTERY. A 190 CM STRAIGHT-TIP (5PK) WAS SELECTED FOR USE. HOWEVER, DURING THE PROCEDURE, THE GUIDEWIRE FRACTURED. THE DEVICE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THE PATIENT'S CONDITION REMAINED STABLE FOLLOWING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20593580 ↗MalfunctionSep 6, 2024Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT A CATHETER ISSUE OCCURRED. THE PATIENT PRESENTED FOR A BIFURCATION PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE. THE TARGET LESION WAS LOCATED IN A CALCIFIED LEFT ANTERIOR DESCENDING CORONARY ARTERY (LAD). A NON-BOSTON SCIENTIFIC (NON-BSC) GUIDEWIRE WAS PLACED AND AN OPTICROSS HD VASCULAR IMAGING CATHETER WAS ADVANCED OVER A SAMURAI GUIDEWIRE FOR ULTRASOUND EXAMINATION OF THE TARGET LESION. DURING PULLBACK, THE IMAGE DISAPPEARED. THE PULLBACK STOPPED AFTER APPROXIMATELY 4CM. THE PHYSICIAN ATTEMPTED TO BRING THE CATHETER BACK INTO STARTING POSITION, BUT IT DID NOT WORK, AND THE OPTICROSS CATHETER WAS STUCK WITH THE GUIDEWIRE. IT WAS NOTED THAT THE CATHETER LOOKED LIKE A CORKSCREW. EVERYTHING WAS REMOVED TOGETHER, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH OTHER DEVICES. THERE WAS NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20158880 ↗MalfunctionJun 28, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT COATING PEELING OCCURRED. THE 85% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY. A 190CM SAMURAI GUIDEWIRE WAS SELECTED FOR USE. HOWEVER, IT WAS NOTED THAT THE GUIDEWIRE WAS PEELED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19637169 ↗InjuryMay 6, 2024Fracture; Difficult to Remove; Difficult to Advance; Fracture; Difficult to Remove; Difficult to Advance
IT WAS REPORTED THAT GUIDEWIRE FRACTURE OCCURRED REQUIRING ADDITIONAL INTERVENTION. THE PATIENT UNDERWENT CORONARY ANGIOGRAPHY AND SUBSEQUENT INTRAVASCULAR ULTRASOUND (IVUS)-GUIDED STENT IMPLANTATION IN THE 80% STENOSED MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. ACCESS WAS GAINED IN THE RIGHT BRACHIAL ARTERY AND A SAMURAI GUIDEWIRE WAS SELECTED AND PLACED IN THE DISTAL SECTION OF THE LAD BRANCH. ANOTHER NON-BOSTON SCIENTIFIC (BSC) GUIDEWIRE WAS ADVANCED INTO THE DISTAL END OF THE SPIRAL BRANCH. IVUS WAS INSERTED FOR EXAMINATION AND THEN SAMURAI GUIDEWIRE COULD NOT BE MOVED FORWARD OR BACKWARD. UPON WITHDRAWAL OF THE SAMURAI GUIDEWIRE, PART OF IT REMAINED IN THE LEFT CORONARY ARTERY AND LEFT CORONARY SINUS. AN ATTEMPT TO USE THE NON-BSC GUIDEWIRE TO WRAP THE FRACTURED SAMURAI GUIDEWIRE FAILED. AN ADDITIONAL ACCESS SITE WAS THEN GAINED ON THE LEFT BRACHIAL ARTERY ALLOWING FOR AN ATTEMPT TO RETRIEVE THE FRACTURED GUIDE WIRE FROM THE LEFT CORONARY ARTERY USING A SNARE DEVICE, BUT IT WAS UNSUCCESSFUL. STENTS WERE SUCCESSFULLY IMPLANTED IN THE LAD ARTERY USING BALLOONS AND STENTS OF VARIOUS SIZES. THE IMPROVEMENT IN THE STENOTIC LESIONS WAS CONFIRMED THROUGH ANGIOGRAPHY AND IVUS. THE PRO
Patient problems: Foreign Body In Patient
MAUDE report 19255737 ↗InjuryJan 5, 2024Fracture; Fracture
IT WAS REPORTED THAT GUIDEWIRE FRACTURE OCCURRED REQUIRING INTERVENTION. A 190CM, STRAIGHT-TIP SAMURAI GUIDEWIRE WAS SELECTED FOR USE. DURING THE PROCEDURE, IT BROKE OFF IN THE PATIENT'S BODY. THE REMAINING PIECE WAS RETRIEVED VIA SNARE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18460683 ↗InjuryJan 5, 2024Entrapment of Device; Failure to Advance; Material Deformation; Entrapment of Device; Failure to Advance; Material Deformation
IT WAS REPORTED THAT DEVICE ENTRAPMENT OCCURRED REQUIRING ADDITIONAL INTERVENTION. THE TARGET LESION WAS LOCATED IN THE MID LEFT ANTERIOR DESCENDING ARTERY (LAD). A 190CM, STRAIGHT-TIP SAMURAI GUIDEWIRE WAS ADVANCED FOR TREATMENT, HOWEVER, WHEN THE PHYSICIAN TRIED TO INSERT THE WIRE, IT FAILED TO CROSS AND STUCK IN THE LESION. A BOSTON SCIENTIFIC FIGHTER WIRE WAS INSERTED AND TWISTED WITH THE SAMURAI. THE WIRES WERE REMOVED TOGETHER, AND IT WAS NOTED THAT THE SAMURAI WIRE WAS KINKED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18451641 ↗MalfunctionNov 29, 2023Failure to Advance; Failure to Advance
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 03-NOV-2023. IT WAS REPORTED THAT CROSSING DIFFICULTIES OCCURRED. THE TARGET LESION IS LOCATED IN THE LEFT CIRCUMFLEX ARTERY. A 190CM, STRAIGHT-TIP SAMURAI GUIDEWIRE FAILED TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED THAT DISTAL TIP DETACHMENT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18231380 ↗MalfunctionOct 20, 2023Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT ENTRAPMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE NON-CALCIFIED AND NON-TORTUOUS LEFT CIRCUMFLEX (LCX) ARTERY EXTENDING INTO THE FIRST OBTUSE MARGINAL BRANCH. THERE WAS ANGULATION INTO THE LCX. AFTER A SAMURAI GUIDE WIRE, ALONG WITH ANOTHER WIRE, WAS PLACED, A 2.50 MM X 12 MM EMERGE BALLOON CATHETER WAS ADVANCED ON THE SAMURAI. WHEN THE BALLOON WAS IN THE GUIDE CATHETER, IT GOT STUCK ON THE SAMURAI WIRE. IT WAS NOTED THAT THE WIRES WERE NOT CROSSED BUT THE BALLOON GOT STUCK MIDWAY IN THE GUIDE. THE EMERGE BALLOON AND SAMURAI WIRE REMAINED INTACT AND WERE REMOVED TOGETHER. THERE WAS NO VISIBLE DAMAGE TO THE EMERGE BALLOON CATHETER. THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICES AND THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17980653 ↗InjurySep 11, 2023Difficult to Remove; Unraveled Material; Difficult to Remove; Unraveled Material
IT WAS REPORTED THAT REMOVAL DIFFICULTIES OCCURRED, AND THE PATIENT WAS SENT FOR BYPASS SURGERY. THE PATIENT UNDERWENT A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE. THE STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED MID RIGHT CORONARY ARTERY. A 190CM, STRAIGHT-TIP (5PK) SAMURAI WAS SELECTED FOR USE. HOWEVER, WHEN THE PHYSICIAN WIRED THE RIGHT CORONARY ARTERY (RCA) WITH THE SAMURAI PRIOR TO DILATING AND STENTING THE BLOCKAGE, THE WIRE BECAME STUCK IN A SMALL VESSEL, AND IT WOULD NOT RELEASE. THE GUIDEWIRE WAS ENTRAPPED IN THE ARTERY, AND IT APPEARED TO BE KNOTTED AND STRETCHED OUT. THE PATIENT WAS GIVEN NITROGLYCERIN TO SEE IF THE WIRE WOULD RELEASE BUT IT DID NOT. SINCE THERE WAS ALSO SIGNIFICANT DISEASE IN THE LEFT ANTERIOR DESCENDING ARTERY (LAD), THE PATIENT WAS SENT FOR CORONARY ARTERY BYPASS SURGERY (CABG) FOR THE BLOCKAGES AND GUIDEWIRE REMOVAL.
Patient problems: Foreign Body In Patient
MAUDE report 17725917 ↗MalfunctionJul 19, 2023Fracture; Fracture
IT WAS REPORTED THAT GUIDEWIRE FRACTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE TORTUOUS MID RIGHT CORONARY ARTERY. A 190CM, STRAIGHT-TIP SAMURAI GUIDEWIRE WAS SELECTED FOR USE. HOWEVER, THE GUIDEWIRE WAS TWISTED AND FRACTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17352281 ↗InjuryMar 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. THE PATIENT PRESENTED FOR PROVISIONAL STENTING ACROSS THE LEFT MAIN (LM) AND LEFT ANTERIOR DESCENDING ARTERY (LAD). THE 60-70% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED VESSEL. THE FIRST SAMURAI GUIDEWIRE CROSSED THE LAD AND THE SECOND SAMURAI THE LEFT CIRCUMFLEX ARTERY (LCX). A 3.5 SYNERGY MEGATRON STENT WAS DEPLOYED FROM THE LAD ACROSS THE LEFT MAIN (LM) AND AN AGENT DCB WAS DEPLOYED AT 6 ATMOSPHERES AT THE MID LCX. THE MEGATRON WAS THEN POST DILATED WITH A 4.00 X 8 NON-COMPLAINT (NC) BALLOON AT THE LAD AND A 4.50 X 8 NC BALLOON AT THE LM. THE VESSEL WAS EXAMINED USING AN OPTICROSS HD IMAGING CATHETER AND SINCE THE LM APPEARED TO BE AROUND 5.00MM, IT WAS POST-DILATED WITH A 5.00 X 8 NC BALLOON. THE OPTICROSS HD WAS USED AGAIN, AND THE PHYSICIAN WAS SATISFIED WITH THE RESULTS. THE PHYSICIAN DECIDED TO REMOVE THE SAMURAI WIRE FROM THE LCX BUT IT WAS SNAPPED AND 4CM REMAINED IN THE VESSEL. A NON-BOSTON SCIENTIFIC WIRE WAS INSERTED IN THE LCX AND A 2.00 EMERGE BALLOON ADVANCED IN AN ATTEMPT TO PULL OUT THE WIRE. THE SAMURAI REMAINED STUCK SO TWO STENTS, A 3.00 X 38 AND A 3.50 X 24, WERE IMPLANTED FROM
Patient problems: Foreign Body In Patient
MAUDE report 16535090 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page