Samurai™ H749393023000Recalls & reported problems

Boston Scientific Corporation · Guidewire · Class 2

All models of Samurai (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729887843), received between Feb 10, 2020 and May 15, 2020 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Difficult to Remove1

Patient problems

  • Chest Pain1
  • Device Embedded In Tissue or Plaque1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT TIP DETACHMENT OCCURRED AND REMAINED IN THE PATIENT. THE TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND HIGHLY CALCIFIED DISTAL LEFT ANTERIOR DESCENDING ARTERY. A 300CM, STRAIGHT-TIP SAMURAI GUIDEWIRE WAS USED AND 1 INCH DETACHED IN THE DISTAL LAD. MULTIPLE BALLOONS WERE USED IN THIS CASE AND STENTING WAS NOT SUCCESSFUL. THE PATIENT WENT TO SURGERY WITH CHEST PAINS. SURGERY WAS PERFORMED BUT THE DETACHED TIP COULD NOT BE RECOVERED. NO FURTHER COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS FINE.

    Patient problems: Chest Pain; Device Embedded In Tissue or Plaque

    MAUDE report 10061563 ↗
  • MalfunctionFeb 10, 2020Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT DIFFICULTY REMOVAL OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. A 3.00 X 32 SYNERGY DRUG-ELUTING STENT AND 300CM STRAIGHT TIP SAMURAI GUIDEWIRE WAS SELECTED FOR USE. HOWEVER, THE STENT GOT DAMAGED BY ATTEMPTING TO PULL OFF FROM THE WIRE, IT GOT STUCK. THE PORTION OF THE WIRE THAT GOT STUCK WAS CUT OFF. THE PROCEDURE WAS COMPLETED WITH A NEW SYNERGY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS GREAT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9687319 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page