NovaGold 5201Recalls & reported problems
Heraeus Medical Components LLC · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729892045), received between Jun 1, 2018 and Nov 3, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
- Insufficient Device Problem Information1
- Material Separation1
Patient problems
- Foreign Body In Patient2
- Fever1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 3, 2021Material Separation; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Material Separation; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WIRE GOT STUCK AND BROKE INSIDE OF THE PATIENT DURING AND ERCP . THE WIRE WAS RETRIEVED AND NOTHING REMAINED IN THE PATIENT. PER SALESFORCE EVENT (B)(4), IT WAS REPORTED THAT: NOVAGOLD WAS EXPERIENCING A LOT OF RESISTANCE AT THE TIP OF THE CANNULA AND WHEN THE NURSE PULLED THE WIRE BACK IT HAD STRIPPED THE GOLD TIP OFF INTO THE CANNULA TIP AND AWAY FROM THE WIRE CORE. WHAT TROUBLESHOOTING STEPS TOOK PLACE? TRIED TO PUSH WIRE THROUGH THE TIP OF THE CANNULA AND IT DIDN'T PASS THROUGH WHAT IS THE NEXT COURSE OF ACTION? NOTHING RESOLVED IT, WHEN THEY PULLED THE WIRE OUT IT STRIPPED THE GOLD TIP OFF. FAILURE MODES RELATED TO THIS DEVICE INCLUDE: DEVICE-GUIDEWIRE INTERACTION WERE THE ISSUES PRESENT UPON OPENING THE PACKAGE?: NO INDICATE ANY ATTACHED/RELATED DEVICE(S): NOVAGOLD GUIDEWIRE DEVICE 2 OF 2 EVENT DATE: (B)(6) 2021.
Patient problems: Foreign Body In Patient
MAUDE report 12743405 ↗MalfunctionJun 1, 2018Fracture; Fracture
IT WAS REPORTED THAT ERCP WAS PERFORMED FOR REMOVAL OF STONE. SINCE CANNULATION WAS DIFFICULT, CONTOUR 5-4-3 TIP AND NOVA GOLD WERE USED. THE PHYSICIAN ATTEMPTED TO PERFORM CANNULATION AIMING AT THE BILE DUCT. IT COULD ONLY BE INSERTED INTO THE PANCREATIC DUCT, SO IN ORDER TO PERFORM PANCREATIC DUCTAL GW METHOD, THE INSERTED NOVA GOLD WAS PLACED IN THE PANCREATIC DUCT AND WHEN THE PHYSICIAN TRIED TO REMOVE THE CANNULA, THE TIP OF THE NOVA GOLD GOT DETACHED. ACCORDING TO THE ASSISTANT, THERE WAS A FEELING OF RESISTANCE WHEN REMOVING THE CANNULA. THE TIP OF NOVA GOLD REMAINED IN THE PANCREATIC DUCT, SO ACCIDENT REPORT WILL BE REPORTED AT THE FACILITY. DELIBERATE INVESTIGATION HAS BEEN REQUESTED ABOUT THE OCCURRENCE OF THE ISSUE.
Patient problems: Fever; Foreign Body In Patient
MAUDE report 7560847 ↗
FDA clearance and classification
The FDA regulates product code OCY (endoscopic guidewire, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NovaGold 5201
Has the NovaGold 5201 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729892045), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NovaGold 5201?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jun 1, 2018 and Nov 3, 2021. In those reports, the most common specific device problems are “Fracture” (1) and “Material Separation” (1); the most common patient problems are “Foreign Body In Patient” (2) and “Fever” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NovaGold 5201?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NovaGold 5201 FDA approved?
Devices of this type (product code OCY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Heraeus Medical Components LLC.
Who makes the NovaGold 5201, and is it still sold?
The FDA's device database lists Heraeus Medical Components LLC as the company responsible for it (brand NovaGold, model 5201). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic Guidewire, Gastroenterology-Urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology81 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death48 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page