Clik™ X MRI SC-4319Recalls & reported problems
Boston Scientific Neuromodulation Corporation · Anchor · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Clik anchor is a small implanted piece that holds the lead of a spinal cord stimulator in place under the skin. The 27 reports mostly describe pain, discomfort or a lump at the anchor site, and sometimes unwanted nerve sensations or stimulation that did not give relief. All 27 report injuries and no deaths. Several needed surgery to reposition, resize the pocket, or remove the anchor because it had loosened, shifted, or become infected. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 27
- Deaths reported
- 0
- Injuries reported
- 27
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 08714729905325), received between Jun 5, 2017 and Apr 14, 2025 — classified as 27 injuries.
The most frequently reported specific device problem is “Defective Device” (1 report).
17 other recalls cover devices cleared under the same FDA submission (P030017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem24
- Defective Device1
- Material Protrusion/Extrusion1
- Unexpected Therapeutic Results1
Patient problems
- Pain13
- Discomfort5
- No Code Available5
- Implant Pain2
- Undesired Nerve Stimulation2
- Bacterial Infection1
- Erosion1
- Fluid Discharge1
Recent adverse event reports
The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.
InjuryApr 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE CLIK ANCHOR SITE OF THE SPINAL CORD STIMULATOR (SCS) SYSTEM WHILE THE STIMULATION IS IN USE. ADJUSTMENTS TO THE STIMULATION WERE PERFORMED, HOWEVER DID NOT ALLEVIATE THE PAIN WHILE THE SYSTEM WAS IN USE. THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS THE PAIN IN WHICH THE PHYSICIAN OPENED THE INCISION SITE AND MADE THE ORIGINAL LOOP AROUND THE ANCHOR STRAIGHT AND THE WHEN THE STIMULATION WAS TURNED ON THE PATIENT EXPERIENCED PAIN AGAIN. THE CLIK ANCHOR WAS REPOSITIONED 15CM AWAY FROM THE ORIGINAL LOCATION, THE STIMULATION WAS TURNED ON AGAIN, AND THE PAIN RETURNED. THE PHYSICIAN THEN REPOSITIONED THE LEAD 10CM AND THEN 5CM BACKWARDS FROM THE EPIDURAL SPACE, STIMULATION WAS TURNED ON AGAIN, AND THE PAIN RETURNED. THE PHYSICIAN ASSESSED THAT THE STIMULATION IRRITATES THE EPIDURAL FAT AND NOTHING COULD BE DONE TO RESOLVE THE PAIN, HE THEREFORE DECIDED TO EXPLANT THE ENTIRE SYSTEM.
Patient problems: Implant Pain
MAUDE report 21836341 ↗InjuryJan 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) PATIENT EXPERIENCED THE CLIK ANCHOR SHAKING WHEN THE OUTPUT OF STIMULATION WAS INCREASED. THE PHYSICIAN STATED THAT HE FELT A PULSATING MOVEMENT WHEN THE CLIK ANCHOR PART WAS SHAKING. NO IMPEDANCE ABNORMALITIES HAVE OCCURRED. THE PATIENT LATER EXPERIENCED INCREASED PAIN AT THE CLIK ANCHOR SITE. THE PHYSICIAN CONSIDERED THE PAIN DUE TO THE FAT LAYER AT THE SITE BEING THIN. AN X-RAY WAS TAKEN AND THE MIGRATION OF THE LEAD AND CLIK ANCHOR WERE NOT CONFIRMED. THE PATIENT UNDERWENT A REVISION PROCEDURE TO REPOSITION THE CLIK ANCHOR A LITTLE DEEPER, SO THAT IT IS NOT PROTRUDING FROM THE SURFACE OF THE BODY. THE PATIENT IS DOING WELL POSTOPERATIVELY AND THERE IS NO PAIN IN THE ANCHOR AREA. STIMULATION ADJUSTMENT WAS PERFORMED AND PARESTHESIA WAS INDUCED IN THE PAIN AREA.
Patient problems: Pain; Undesired Nerve Stimulation
MAUDE report 21107944 ↗InjuryOct 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING DISCOMFORT AT THE ANCHOR SITE. THE PATIENT UNDERWENT A PROCEDURE WHEREIN THE CASING OF THE ANCHOR WAS REMOVED. THE EXPLANTED DEVICE WAS DISCARDED BY THE MEDICAL FACILITY AND WILL NOT BE RETURNED.
Patient problems: Discomfort
MAUDE report 20513263 ↗InjurySep 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A LUMP ON THE MIDLINE INCISION. THE PATIENT UNDERWENT A REVISION WHEREIN THE PHYSICIAN CUT A SCAR TISSUE AND SUTURED THE SPINAL CORD STIMULATION (SCS) LEAD WITH THE SAME ANCHOR. THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: Scar Tissue
MAUDE report 20186078 ↗InjurySep 22, 2023Defective Device; Defective Device
IT WAS REPORTED THAT THE PATIENTS CLIK ANCHOR WAS LOOSE. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE CLIK ANCHOR WAS REPOSITIONED. THE PATIENT WAS DOING WELL POST-OPERATIVELY. THE DEVICE REMAINED IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17797195 ↗InjuryMay 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE ANCHOR SITE. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE POCKET WAS OPENED AND THEN DISSECTED LATERALLY ALONG THE SPINE IN ORDER TO CREATE A BIGGER POCKET AND BE ABLE TO HIDE THE ANCHOR. NOTHING WAS EXPLANTED AND THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: Implant Pain
MAUDE report 16961266 ↗InjurySep 6, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION AT THE CLIK ANCHOR SITE. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE WHERE THE CLIK ANCHOR, LEAD, AND LEAD EXTENSION WERE EXPLANTED. POST OPERATIVELY, THE PATIENT WAS DOING WELL. CULTURES TAKEN CONFIRMED INFECTION. THE DEVICES WERE RETAINED BY THE FACILITY AND WERE NOT RETURNED FOR ANALYSIS.
Patient problems: Unspecified Infection
MAUDE report 15363446 ↗InjuryJan 7, 2022Unexpected Therapeutic Results; Unexpected Therapeutic Results
IT WAS REPORTED THAT THE PATIENT HAD INADEQUATE STIMULATION. THE LEAD AND IPG WERE ADJUSTED AND THE CLIK ANCHOR WAS REPLACED. THE PATIENT WAS DOING WELL POSTOPERATIVELY AND THE EXPLANTED DEVICE WILL NOT BE RETURNED PER HOSPITAL POLICY.
Patient problems: Inadequate Pain Relief
MAUDE report 13189337 ↗InjuryJan 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS HAVING PAIN NEAR THE ANCHOR SITE. THERE WERE NO DEVICE MALFUNCTION SUSPECTED. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR REVISION PROCEDURE WHEREIN THE LEADS WERE REPLACED AS THEY WERE DAMAGED DURING THE PROCEDURE. THE PATIENT WAS DOING WELL POSTOPERATIVELY AND EXPLANTED DEVICES WILL NOT BE RETURNED.
Patient problems: Pain
MAUDE report 13163864 ↗InjuryOct 8, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING PAIN AND DISCOMFORT AT THE ANCHOR SITE AND THE PAIN WORSEN WHEN LAYING ON THE SIDE. THE PAIN UNDERWENT A REVISION WHERE IN THE ANCHOR WAS BURIED DEEPER. THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: Discomfort; Pain
MAUDE report 12598492 ↗InjuryApr 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A CLIK ANCHOR THAT HAD PROTRUDED AND A BED SORE WAS CONFIRMED AROUND THAT AREA. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE TO REMOVE THE ENTIRE SPINAL CORD STIMULATOR SYSTEM IN CONSIDERATION OF THE RISK OF INFECTION. THE PHYSICIAN STATED THAT THE PATIENT HAD EXPERIENCED WEIGHT LOSS AND THERE IS NO RELATION BETWEEN THE IMPLANTED DEVICES AND THIS EVENT. NO INFECTION WAS OBSERVED DURING THE EXPLANT PROCEDURE. ALL EXPLANTED DEVICES WERE DISCARDED BY THE FACILITY.
Patient problems: Erosion; Pressure Sore/Ulcer; Weight Changes
MAUDE report 11675870 ↗InjuryMar 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS HAVING PAIN AT MIDLINE DUE TO A LOOSE ANCHOR. IT WAS ALSO REPORTED THAT THE PATIENT WAS EXPERIENCING NAUSEA AND VOMITING. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE ANCHOR WAS REATTACHED AND IPG AND LEADS TOGETHER WITH THE CLIK ANCHORS WERE REPLACED. THE PATIENT WAS DOING WELL POST-OPERATIVELY. ADDITIONAL INFORMATION WAS RECEIVED THAT PATIENT HAS POST-OPERATIVE PAIN. THE PHYSICIAN SUSPECTED A DURA PUNCTURE AND ADMINISTERED A BLOOD PATCH AT THE END OF THE REVISION SURGERY.
Patient problems: Nausea; Pain; Vomiting
MAUDE report 11478199 ↗InjuryDec 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD AN INFECTION. SYMPTOMS OF SMALL LUMP AT THE LEAD ENTRY INCISION SITE AND SMALL OOZING CYST AT THE ANCHOR SITE WERE NOTED. IT WAS NOT CONFIRMED IF THE INFECTION WAS DEVICE OR PROCEDURE RELATED. THE PATIENT WAS PRESCRIBED WITH ANTIBIOTICS AND UNDERWENT AN EXPLANT PROCEDURE. ALL DEVICES WERE REMOVED AND WERE NOT RETURNED.
Patient problems: Bacterial Infection; Fluid Discharge; No Code Available
MAUDE report 10964332 ↗InjuryDec 9, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT EXPERIENCED SOME PAIN AND IRRITATION AT THE ANCHOR SITE. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE ANCHOR WAS REPOSITIONED. THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: No Code Available; Pain; Skin Irritation
MAUDE report 9441188 ↗InjuryJul 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENTS CLIK ANCHOR WAS SUPERFICIAL AND CAUSING PAIN. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE CIK ANCHOR WAS REPLACED. THE PATIENT WAS REPORTEDLY DOING WELL POSTOPERATIVELY.
Patient problems: No Code Available; Pain
MAUDE report 8769997 ↗InjuryMar 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING UNCOMFORTABLE STIMULATION EVEN IF THE DEVICE HAD BEEN TURNED OFF. IT WAS ALSO REPORTED THAT THE PATIENT HAD SORENESS AROUND THE ANCHOR SITE. THE PATIENT WAS SENT TO THE EMERGENCY ROOM AND WAS GIVEN PAIN MEDICATION.
Patient problems: Discomfort; Undesired Nerve Stimulation
MAUDE report 8452363 ↗InjuryMar 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING TENDERNESS OVER THE MIDLINE INCISION. THE PATIENT UNDERWENT AN ANCHOR REVISION PROCEDURE WHEREIN THE PHYSICIAN BURIED IT DEEPER. THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: Discomfort; No Code Available
MAUDE report 8428656 ↗InjuryMar 13, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT WAS HAVING PAIN AT THE ANCHOR SITE. IT WAS MENTIONED THAT THE ANCHOR HAD COME LOOSE AND WAS PUSHING THROUGH THE SKIN. NO SKIN BREAKAGE WAS NOTED. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE AND WAS REPORTEDLY DOING WELL POSTOPERATIVELY.
Patient problems: No Code Available; Pain
MAUDE report 8419046 ↗InjuryJul 20, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING PAIN AT THE ANCHOR SITE. THE PATIENT WILL UNDERGO A REVISION PROCEDURE WHEREIN THE ANCHOR WILL BE MOVED DEEPER.
Patient problems: Pain
MAUDE report 7708343 ↗InjuryJul 12, 2018Material Protrusion/Extrusion; Material Protrusion/Extrusion
A REPORT WAS RECEIVED THAT THE PATIENTS ANCHOR WAS SOMEWHAT PROTRUDING BUT NO SKIN BREAKAGE WAS NOTED. THE PATIENT UNDERWENT SURGERY TO SUTURE THE ANCHOR DOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7684404 ↗InjuryApr 2, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING DISCOMFORT ON WHERE THE ANCHORS WAS BURIED. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE ANCHORS WHERE REPOSITIONED. THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: Discomfort
MAUDE report 7390989 ↗InjuryFeb 15, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING PAIN AT THE LEAD SITE. THE PATIENT UNDERWENT A REVISION WHEREIN THE CLIK ANCHOR WAS REPOSITIONED. THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: Pain
MAUDE report 7273497 ↗InjuryOct 3, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE ANCHORS WERE CAUSING THORACIC PAIN TO THE PATIENT. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE ANCHORS WERE BURIED DEEPER.
Patient problems: Pain
MAUDE report 6909673 ↗InjurySep 12, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING MIDLINE PAIN. IT WAS ALSO REPORTED THAT THE MAIN REASON FOR ANCHOR REVISION WAS DISCOMFORT. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE CLIK ANCHOR WAS RELOCATED BY CUTTING THE SUTURE AND RE-SUTURED WITH A NEW FIXATE. THE PATIENT WAS DOING POSTOPERATIVELY.
Patient problems: Pain
MAUDE report 6857833 ↗InjuryJul 17, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING IRRITATION AT THE CLIK ANCHOR SITE. THE PATIENT UNDERWENT A REVISION PROCEDURE WHEREIN THE CLIK ANCHOR WAS REPLACED AND PLACED DEEPER. NO DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT WAS DOING WELL POSTOPERATIVELY.
Patient problems: Irritation
MAUDE report 6720084 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030017. One submission often covers many models, so check the product and lot details in each record.
- Z-2879-2026 · Jun 17, 2026 · Open, ClassifiedINFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spin
- Z-2404-2026 · Apr 29, 2026 · Open, ClassifiedBoston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
- Z-2617-2024 · Jul 17, 2024 · Open, ClassifiedWAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
- Z-2123-2020 · Apr 24, 2019 · TerminatedPrecision TM M8 Adapter, 15 cm, part number M365SC9218150 - Product Usage: The adapter is part of a spinal cord stimulation system for prescription use only indicated as an aid in
- Z-2120-2020 · Apr 24, 2019 · TerminatedThirty-five (35) cm 8 Contact Extension Kit, part number M365SC3138350. Package Contents: Lead Extension Kit (1) Lead Extension (1) Hex Wrench (1) Tunneling Tool Assembly (1) Devic
- Z-2121-2020 · Apr 24, 2019 · TerminatedLinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's Spinal Cord Stimulation (SCS) systems by delivering electric
- Z-2124-2020 · Apr 24, 2019 · TerminatedFifty-five (55) cm 8 Contact Extension Kit, part number M365SC3138550 - Product Usage: Lead Extensions are designed to connect the percutaneous leads to the Implantable Pulse Gene
- Z-2122-2020 · Apr 24, 2019 · TerminatedThe LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for prescription use only as an aid in the management of chronic i
- Z-2119-2020 · Apr 24, 2019 · TerminatedArtisanTM 70 cm 2 x 8 Surgical Lead Kit, part number M365SC8216700. 2 x 8 Surgical Paddle Lead Kit: (1) One Paddle Lead (4); Four Suture Sleeves; (2) Lead Position Labels, left
- Z-1409-2010 · Mar 19, 2010 · TerminatedBionic NAVIGATOR Clinician Programmer Model Number SC-7150-4
- Z-0271-2009 · Sep 22, 2008 · TerminatedPrecision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System labeled as BIONICS SCS-CHARGING KIT CONTENTS Charger (1) Base Station (1) Belt Po
- Z-1441-2012 · Apr 22, 2008 · TerminatedOMG (Observational Mechanical Gateway) Connector - A Model Number SC-9305 The Precision Spinal Cord Stimulator System is indicated as an aid in the management of chronic intractabl
- Z-0960-2008 · Oct 5, 2007 · TerminatedAdvanced Bionics Precision Implantable Pulse Generator Kit, Model Number: SC-1110 with firmware version prior to Revision 3.02, Advanced Bionics Corportation, Sylmar, CA 91342
- Z-0055-2008 · Aug 3, 2007 · TerminatedPrecision Implantable Pulse Generator (IPG), Model Number SC-110
- Z-0700-2007 · Jan 30, 2007 · TerminatedThe Precision Implantable Pulse Generator, Model Number SC-1110
…and 2 more.
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page