Artisan™ MRI SC-8416-50Recalls & reported problems
Boston Scientific Neuromodulation Corporation · 50cm 2x8 Surgical Lead Kit · Class 3 · Implant
All models of Artisan MRI (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Artisan MRI spinal cord stimulator lead kit has 107 reports, no deaths and 103 injuries. Many reports describe the lead migrating out of place, which led to poor pain relief or unwanted stimulation, and some describe high electrical resistance, a broken lead, or infection, with some patients needing revision or removal surgery. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 108
- Deaths reported
- 0
- Injuries reported
- 104
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 08714729951124), received between Sep 8, 2017 and Jul 7, 2026 — classified as 104 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Unexpected Therapeutic Results” (39 reports).
17 other recalls cover devices cleared under the same FDA submission (P030017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Unexpected Therapeutic Results39
- Migration38
- Adverse Event Without Identified Device or Use Problem36
- High impedance11
- Fracture4
- Migration or Expulsion of Device3
- Device Dislodged or Dislocated2
- Positioning Problem2
Patient problems
- No Code Available33
- Pain21
- No Clinical Signs, Symptoms or Conditions17
- Inadequate Pain Relief12
- Undesired Nerve Stimulation12
- Bacterial Infection11
- Paralysis8
- Hematoma5
Recent adverse event reports
The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.
InjuryJul 7, 2026Unexpected Therapeutic Results; Migration
IT WAS REPORTED THAT THE PATIENT EXPERIENCED LEAD MIGRATION FOLLOWING A FALL, WHICH WAS CONFIRMED THROUGH IMAGING STUDIES. UPON EVALUATION, THE PHYSICIAN OBSERVED THAT THE MIDLINE SPINAL INCISION APPEARED SUSPICIOUS, THOUGH NO EXTERNAL DRAINAGE WAS PRESENT. THE PATIENT WAS TAKEN TO THE OPERATING ROOM FOR FURTHER ASSESSMENT. DURING SURGICAL EXPLORATION OF THE SPINAL INCISION, SUSPICIOUS FLUID WAS ENCOUNTERED WITHIN THE MIDLINE INCISION CAVITY. BOTH SUPERFICIAL AND DEEP CULTURES WERE OBTAINED FOR ANALYSIS. SUBSEQUENTLY, THE PATIENT UNDERWENT A SPINAL CORD STIMULATION (SCS) EXPLANT PROCEDURE. POSTOPERATIVELY, THE PATIENT WAS ADMITTED FOR INPATIENT CARE AND WILL BE INITIATED ON ANTIBIOTIC THERAPY. EXPLANTS DISCARDED PER FACILITY BIOHAZARD PROTOCOL.
Patient problems: Bacterial Infection; Fluid Discharge; Pain; Wound Dehiscence
MAUDE report 25740370 ↗InjuryMar 30, 2026Unexpected Therapeutic Results; Migration
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INADEQUATE STIMULATION DUE TO LEAD MIGRATION THAT WAS CONFIRMED VIA X-RAY. THE PATIENT UNDERWENT A LEAD REVISION PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.
Patient problems: Pain
MAUDE report 24714016 ↗InjuryNov 27, 2025Unexpected Therapeutic Results; Migration; Unexpected Therapeutic Results; Migration
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE PAIN RELIEF AND FEELING STRONG UNWANTED STIMULATION. THE SPINAL CORD STIMULATION (SCS) PADDLE MIGRATED. THE PATIENT UNDERWENT A SPINAL CORD STIMULATION (SCS) EXPLANT PROCEDURE. THE EXPLANTED DEVICE WILL NOT BE RETURNED.
Patient problems: Pain; Undesired Nerve Stimulation
MAUDE report 23665497 ↗InjuryNov 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD PAIN AT THE LEAD INCISION SITE AND WENT THE EMERGENCY ROOM. THE PATIENTS PAIN WAS DEEMED TO BE POSTOPERATIVE PAIN. THE PATIENT WAS PRESCRIBED WITH PAIN MEDICATIONS AND WAS DISCHARGED FROM THE HOSPITAL. THE PATIENT WAS DOING WELL.
Patient problems: Implant Pain
MAUDE report 23638234 ↗InjuryJun 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT FOLLOWING A PADDLE LEAD IMPLANT PROCEDURE, THE PATIENT WAS EXPERIENCING LOWER EXTREMITY WEAKNESS AND WAS UNABLE TO MOVE THE LOWER EXTREMITIES. THE PHYSICIAN INSTRUCTED THE PATIENT TO BE TAKEN TO THE HOSPITAL.
Patient problems: Muscle Weakness/Atrophy
MAUDE report 22305262 ↗MalfunctionMay 27, 2025Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT'S LEAD WAS BROKEN IN HALF WHICH WAS CONFIRMED THROUGH X-RAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22088036 ↗InjuryMar 26, 2025High impedance; High impedance
IT WAS REPORTED THAT THE LEADS HAD HIGH IMPEDANCES. THE PATIENT NEEDED AN MRI; THEREFORE, THE SPINAL CORD STIMULATOR SYSTEM WAS EXPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21704047 ↗InjuryFeb 20, 2025Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT SPINAL CORD STIMULATION (SCS) LEAD REPLACEMENT PROCEDURE DUE TO MIGRATION AS CONFIRMED VIA X-RAY. IT WAS ALSO NOTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS REPLACED DUE TO AN UNKNOWN REASON. NO FURTHER INFORMATION COULD BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21427676 ↗InjuryNov 26, 2024Unexpected Therapeutic Results; Migration; Unexpected Therapeutic Results; Migration
IT WAS REPORTED THAT PATIENTS SPINAL CORD STIMULATOR LEAD HAD MIGRATED AND WAS NOTED THAT PATIENT HAD LOSS ITS COVERAGE. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR LEAD REPOSITIONING PROCEDURE AND WAS DOING WELL POST OPERATIVELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20785074 ↗InjuryNov 1, 2024High impedance; Failure to Disconnect; High impedance; Failure to Disconnect
IT WAS REPORTED THAT THE LEAD HAD HIGH IMPEDANCES. DURING THE PROCEDURE TO REPLACE THE LEAD, THE LEAD WOULD NOT DISCONNECT FROM THE IMPLANTABLE PULSE GENERATOR (IPG) AS THE PORT SCREWS WOULD NOT DISENGAGE. THE PHYSICIAN OPTED TO REPLACE BOTH LEAD AND IPG. THE PATIENT WAS DOING WELL POSTOPERATIVELY AND THE EXPLANTED DEVICES WERE NOT RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20588499 ↗InjuryOct 18, 2024High impedance; High impedance
IT WAS REPORTED THAT THE LEAD HAD HIGH IMPEDANCES. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE WAS NOT RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20479328 ↗InjurySep 20, 2024Migration; Migration
IT WAS REPORTED THAT THE PATIENTS PADDLE LEAD HAD MIGRATED WHICH WAS CONFIRMED THROUGH AN X-RAY. THE PHYSICIAN WAS UNSUCCESSFUL IN REPOSITIONING THE LEAD AND DECIDED TO REMOVE THE PADDLE LEAD. THE PATIENT WAS DOING WELL POSTOPERATIVELY, AND THE EXPLANTED DEVICE WAS KEPT BY THE MEDICAL FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20274265 ↗InjuryAug 20, 2024Unexpected Therapeutic Results; Migration; Unexpected Therapeutic Results; Migration
IT WAS REPORTED THAT THE PADDLE LEAD HAD BACKED OUT OF PLACE AND PATIENTS THERAPY WAS LOST. THE PATIENT UNDERWENT A LEAD REPOSITIONING PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE DEVICE REMAINS IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20001351 ↗MalfunctionAug 3, 2024Fracture; Unexpected Therapeutic Results; Fracture; Unexpected Therapeutic Results
IT WAS REPORTED THAT PATIENT HAD STOPPED CHARGING AS STIMULATION WAS NOT PROVIDING ADEQUATE RELIEF. IT WAS FOUND DURING X-RAY THAT PATIENT HAD A SPINAL CORD STIMULATION (SCS) LEAD FRACTURED. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE. THE EXPLANTED DEVICE WAS KEPT BY THE FACILITY.
Patient problems: Pain
MAUDE report 19902342 ↗InjuryJul 12, 2024Migration; Migration
IT WAS REPORTED THAT THE PATIENTS LEAD HAD MIGRATED AS REVEALED FROM XRAY. THE PATIENT UNDERWENT A LEAD REPOSITIONED PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. NOTHING WAS EXPLANTED OR ADDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19733575 ↗MalfunctionJul 5, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENTS PADDLE LEAD CONTACTS CAME OFF AND THIS WAS CONFIRMED THROUGH IMAGING. NO FURTHER INFORMATION COULD BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19684824 ↗InjuryJun 18, 2024Unexpected Therapeutic Results; Unexpected Therapeutic Results
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INADEQUATE PAIN RELIEF AND RIB STIMULATION FOLLOWING AN UPGRADE PROCEDURE. REPROGRAMMING WAS ATTEMPTED, HOWEVER, UNSUCCESSFUL. THE PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE PADDLE LEAD WAS REPLACED AND RELOCATED TO ACHIEVE BETTER COVERAGE. THE PATIENT WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEAD WAS NOT RETURNED AS IT WAS KEPT BY THE MEDICAL FACILITY.
Patient problems: Pain; Undesired Nerve Stimulation
MAUDE report 19562795 ↗InjuryMay 16, 2024Unexpected Therapeutic Results; Unexpected Therapeutic Results
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE PAIN RELIEF DUE TO PADDLE LEAD POSITIONED TOO LATERAL. THE PATIENT UNDERWENT A SPINAL CORD STIMULATION (SCS) REPLACEMENT AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE COMPONENTS WERE DISCARDED BY THE FACILITY.
Patient problems: Pain
MAUDE report 19336694 ↗InjuryMay 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD IMAGING DONE WHICH CONFIRMED A POOR LEAD PLACEMENT. THE PATIENT UNDERWENT A LEAD REPOSITIONING PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE DEVICE REMAINS IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19244446 ↗InjuryMar 25, 2024Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE DUE TO LEAD MIGRATION. THE EXPLANTED DEVICE WILL NOT BE RETURNED AS IT WAS NOT RELEASED BY THE FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18970671 ↗InjuryJan 27, 2024Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION DUE TO THE LEAD MIGRATED IN THE SPINE. THE PATIENT WAS DOING WELL POSTOPERATIVELY, AND NO DEVICE WILL BE RETURN AS THE PRODUCT WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18592791 ↗InjuryJan 5, 2024Unexpected Therapeutic Results; Migration; Unexpected Therapeutic Results; Migration
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE PAIN RELIEF DESPITE REPROGRAMMING. X-RAY CONFIRMED LEAD MIGRATION. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE WAS NOT RETURNED.
Patient problems: Pain
MAUDE report 18451300 ↗InjuryJan 4, 2024Migration; Migration
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING PAIN AT THE MIDLINE INCISION SITE WHEN PATIENT LEANS ON IT. X-RAY SHOWED THAT THE PADDLE LEAD HAD MIGRATED. THE PATIENT UNDERWENT A REVISION PROCEDURE, WHEREIN THE PHYSICIAN FIXED THE LEAD TAIL WHICH HERNIATED THROUGH THE PATIENTS FASCIA. NO PRODUCT WAS ADDED OR REMOVED. THE PATENT WAS DOING WELL POST-OPERATIVELY.
Patient problems: Implant Pain
MAUDE report 18443808 ↗InjuryDec 6, 2023High impedance; High impedance
IT WAS REPORTED THAT PATIENT WAS EXPERIENCING SHARP PAIN. HIGH IMPEDANCES WERE ALSO NOTED. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POST OPERATIVELY. THE EXPLANTED DEVICE WAS DISCARDED BY THE FACILITY.
Patient problems: Pain; Undesired Nerve Stimulation
MAUDE report 18274786 ↗InjuryDec 2, 2023Unexpected Therapeutic Results; Unexpected Therapeutic Results
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING INADEQUATE STIMULATION. THE PATIENT UNDERWENT A PADDLE REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. NO DEVICE MALFUNCTION SUSPECTED. THE EXPLANTED PRODUCT WILL NOT BE RETURNED BY THE FACILITY.
Patient problems: Pain
MAUDE report 18253381 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030017. One submission often covers many models, so check the product and lot details in each record.
- Z-2879-2026 · Jun 17, 2026 · Open, ClassifiedINFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spin
- Z-2404-2026 · Apr 29, 2026 · Open, ClassifiedBoston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
- Z-2617-2024 · Jul 17, 2024 · Open, ClassifiedWAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
- Z-2121-2020 · Apr 24, 2019 · TerminatedLinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's Spinal Cord Stimulation (SCS) systems by delivering electric
- Z-2120-2020 · Apr 24, 2019 · TerminatedThirty-five (35) cm 8 Contact Extension Kit, part number M365SC3138350. Package Contents: Lead Extension Kit (1) Lead Extension (1) Hex Wrench (1) Tunneling Tool Assembly (1) Devic
- Z-2123-2020 · Apr 24, 2019 · TerminatedPrecision TM M8 Adapter, 15 cm, part number M365SC9218150 - Product Usage: The adapter is part of a spinal cord stimulation system for prescription use only indicated as an aid in
- Z-2122-2020 · Apr 24, 2019 · TerminatedThe LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for prescription use only as an aid in the management of chronic i
- Z-2124-2020 · Apr 24, 2019 · TerminatedFifty-five (55) cm 8 Contact Extension Kit, part number M365SC3138550 - Product Usage: Lead Extensions are designed to connect the percutaneous leads to the Implantable Pulse Gene
- Z-2119-2020 · Apr 24, 2019 · TerminatedArtisanTM 70 cm 2 x 8 Surgical Lead Kit, part number M365SC8216700. 2 x 8 Surgical Paddle Lead Kit: (1) One Paddle Lead (4); Four Suture Sleeves; (2) Lead Position Labels, left
- Z-1409-2010 · Mar 19, 2010 · TerminatedBionic NAVIGATOR Clinician Programmer Model Number SC-7150-4
- Z-0271-2009 · Sep 22, 2008 · TerminatedPrecision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System labeled as BIONICS SCS-CHARGING KIT CONTENTS Charger (1) Base Station (1) Belt Po
- Z-1441-2012 · Apr 22, 2008 · TerminatedOMG (Observational Mechanical Gateway) Connector - A Model Number SC-9305 The Precision Spinal Cord Stimulator System is indicated as an aid in the management of chronic intractabl
- Z-0960-2008 · Oct 5, 2007 · TerminatedAdvanced Bionics Precision Implantable Pulse Generator Kit, Model Number: SC-1110 with firmware version prior to Revision 3.02, Advanced Bionics Corportation, Sylmar, CA 91342
- Z-0055-2008 · Aug 3, 2007 · TerminatedPrecision Implantable Pulse Generator (IPG), Model Number SC-110
- Z-0700-2007 · Jan 30, 2007 · TerminatedThe Precision Implantable Pulse Generator, Model Number SC-1110
…and 2 more.
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page