Spectra WaveWriter™ SC-1060-BRecalls & reported problems
Boston Scientific Neuromodulation Corporation · Implantable Pulse Generator and Charging Kit · Class 3 · Implant
All models of Spectra WaveWriter (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Boston Scientific Spectra WaveWriter spinal cord stimulator is an implanted device for pain relief. Reports mostly describe charging problems, including needing to charge more often or longer, and loss of pain relief, often leading to replacement of the pulse generator. There are 26 reports naming this identifier, with no deaths and 26 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 26
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 08714729953432), received between May 22, 2019 and May 13, 2026 — classified as 26 injuries.
The most frequently reported specific device problem is “Charging Problem” (13 reports).
17 other recalls cover devices cleared under the same FDA submission (P030017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem13
- Charging Problem13
- Unexpected Therapeutic Results5
- Battery Problem1
- High impedance1
- Impedance Problem1
- Intermittent Energy Output1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions19
- Pain3
- Inadequate Pain Relief2
- Undesired Nerve Stimulation2
- Complaint, Ill-Defined1
- Fall1
- Headache1
- Hematoma1
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
InjuryMay 13, 2026Charging Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) PATIENT UNDERWENT A REVISION PROCEDURE DUE HAVING INCREASED CHARGING TIME AND ALSO WANTED MAGNETIC RESONANCE IMAGING (MRI) CAPABILITIES. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS REPLACED. ELECTROCAUTERY WAS USED. THE PATIENT WAS STABLE POST OPERATION AND DOING GREAT. THE EXPLANTED IPG WILL NOT BE RETURNED AS IT WAS RETAINED BY THE MEDICAL FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25166436 ↗InjuryMar 30, 2026Charging Problem
IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) IMPLANTABLE PULSE GENERATOR (IPG) WAS REPLACED DUE TO CHARGING DIFFICULTIES AS PATIENT HAD TO CHARGING MORE FREQUENTLY THE IPG. THE PATIENT WAS REPORTED TO BE DOING WELL POSTOPERATIVELY. THE EXPLANTED DEVICE WILL NOT BE RETURNED FOR ANALYSIS, AS FACILITY PROTOCOL PROHIBITS DEVICE RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24736649 ↗InjuryMar 20, 2026Unexpected Therapeutic Results
IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) PATIENT UNDERWENT A REPLACEMENT PROCEDURE BECAUSE THE IMPLANTABLE PULSE GENERATOR (IPG) WAS NO LONGER PROVIDING THERAPY. THE DEVICE WAS RETAINED BY THE FACILITY AND WILL NOT BE RETURNED FOR ANALYSIS. POSTOPERATIVELY, THERAPY WAS RESUMED WITHOUT COMPLICATIONS.
Patient problems: Pain
MAUDE report 24652022 ↗InjuryMar 19, 2026Charging Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICE IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Fall; No Clinical Signs, Symptoms or Conditions
MAUDE report 24645967 ↗InjuryJan 28, 2026High impedance; Unexpected Therapeutic Results; Charging Problem; Impedance Problem
IT WAS REPORTED THAT THE PATIENT HAD A CAR ACCIDENT AND WAS NOT GETTING ADEQUATE PAIN RELIEF. IT WAS ALSO NOTED THAT THE PATIENT HAD UNDESIRED STIMULATION AT THE SPINE AND HAD WEAKNESS WHENEVER THE STIMULATION WAS ON. IT WAS ALSO REPORTED THAT THE PATIENT HAD TO CHARGE THE IMPLANTABLE PULSE GENERATOR (IPG) MORE FREQUENTLY AFTER THE ACCIDENT. THERE WERE HIGH IMPEDANCES NOTED ON THE LEADS. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR (SCS) REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT ALL EXPLANTED DEVICES WERE DISCARDED BY THE MEDICAL FACILITY.
Patient problems: Headache; Inappropriate Device Stimulation of Tissue; Muscle Weakness/Atrophy; Pain; Undesired Nerve Stimulation
MAUDE report 24199638 ↗InjuryJan 13, 2026Charging Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO DIFFICULTIES CHARGING. THE PATIENT UNDERWENT AN IMPLANTABLE PULSE GENERATOR (IPG) REVISION PROCEDURE WHEREIN THEIR IPG WAS EXPLANTED AND REPLACED. THE IPG WAS DISCARDED PER HOSPITAL POLICY AND WILL NOT BE RETURNED FOR ANALYSIS. THE REVISION PROCEDURE WAS COMPLETED SUCCESSFULLY, WITH NO POSTOPERATIVE COMPLICATIONS OBSERVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24059932 ↗InjuryOct 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICE IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23438661 ↗InjuryOct 30, 2025Charging Problem; Adverse Event Without Identified Device or Use Problem; Charging Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICE IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23425757 ↗InjuryOct 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REVISED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICE IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23397779 ↗InjurySep 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) DEVICE WAS REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICE IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23080552 ↗InjuryJul 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) SYSTEM UNDERWENT A REVISION PROCEDURE FOR AN UNKNOWN REASON. THE CURRENT LOCATION OF THE DEVICE REMAINS UNKNOWN. NO ADDITIONAL INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22493961 ↗InjuryMay 27, 2025Charging Problem; Migration; Charging Problem; Migration
IT WAS REPORTED THAT THE PATIENT HAD DIFFICULTY CHARGING THE IMPLANTABLE PULSE GENERATOR (IPG). IT WAS ALSO NOTED THAT IMAGING WAS TAKEN AND SHOWED THAT THE LEFT LEAD WAS ANTERIOR AND MAY HAVE MIGRATED. THE PATIENT UNDERWENT AN IPG AND LEAD REPLACEMENT PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22093315 ↗InjuryApr 29, 2025Charging Problem; Charging Problem
IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) SYSTEM UNDERWENT AN EXPLANT PROCEDURE DUE TO CHARGING DIFFICULTIES, THE PATIENT IS RECOVERING WELL. THE EXPLANTED DEVICE WILL NOT BE RETURN AS IT WAS RETAINED BY FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21925445 ↗InjuryApr 29, 2025Charging Problem; Charging Problem
IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING DIFFICULTY CHARGING THE IMPLANTABLE PULSE GENERATOR (IPG). THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21924031 ↗InjuryJan 30, 2025Battery Problem; Battery Problem
IT WAS REPORTED THAT THE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EXPLANTED DUE TO AN UNKNOWN REASON... THE PATIENT IS DOING GREAT POST OPERATIVELY. THE EXPLANTED DEVICE WILL NOT BE RETURNED AS IT WAS DISPOSED PER FACILITY POLICY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21279801 ↗InjuryOct 15, 2024Unexpected Therapeutic Results; Charging Problem; Wireless Communication Problem; Unexpected Therapeutic Results; Charging Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE PATIENTS IMPLANTABLE PULSE GENERATOR (IPG) HAD DIFFICULTY COMMUNICATING WITH THE REMOTE CONTROL. IT WAS ALSO NOTED THAT THE IPG WOULD NOT HOLD A CHARGE. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY.
Patient problems: No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 20457080 ↗InjuryAug 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) SYSTEM WAS REPLACED DUE TO AN UNKNOWN REASON. THE LOCATION OF THE DEVICES IS UNKNOWN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20083140 ↗InjuryAug 28, 2024Charging Problem; Adverse Event Without Identified Device or Use Problem; Charging Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS SPINAL CORD STIMULATION (SCS) SYSTEM WAS REPLACED DUE TO AN INCREASED CHARGING BURDEN OF THE IMPLANTABLE PULSE GENERATOR (IPG) AS WELL AS A DECREASE IN EFFICACY OF THERAPY WHEN THE BATTERY WAS LOW. THE PATIENT IS GETTING GOOD PAIN RELIEF POST OPERATIVELY. DEVICE WILL NOT RETURN DUE TO FACILITY POLICY AND ELECTROCAUTERY WAS NOT USED NEAR NEW IPG.
Patient problems: Inadequate Pain Relief; No Clinical Signs, Symptoms or Conditions
MAUDE report 20083108 ↗InjuryJun 19, 2024Unexpected Therapeutic Results; Charging Problem; Unexpected Therapeutic Results; Charging Problem
IT WAS REPORTED THAT THE PATIENT HAD TO FREQUENTLY CHARGE THE IPG AND EXPERIENCED AN INADEQUATE STIMULATION. THE LEADS HAD HIGH IMPEDANCES. ALL COMPONENTS WERE EXPLANTED AND DISCARDED PER HOSPITAL POLICY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19571218 ↗InjuryFeb 26, 2024Intermittent Energy Output; Intermittent Energy Output
IT WAS REPORTED THAT THE PATIENTS IMPLANTABLE PULSE GENERATOR (IPG) HAD MOVED TOWARDS THE SPINE CAUSING INTERMITTENT AND UNDESIRED STIMULATION. THE PATIENT UNDERWENT AN EXPLANT PROCEDURE.
Patient problems: Undesired Nerve Stimulation
MAUDE report 18785027 ↗InjuryJun 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN EXPLANT PROCEDURE DUE TO AN UNKNOWN REASON. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17057064 ↗InjuryMay 31, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT WAS EXPERIENCING PAIN AT THE POCKET SITE. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE WHEREIN AN MRI COMPATIBLE IPG WAS IMPLANTED. THE PATIENT WAS DOING WELL POST OPERATIVELY. THE EXPLANTED DEVICE WILL NOT BE RETURNED AS IT WAS KEPT AT THE FACILITY.
Patient problems: Implant Pain
MAUDE report 17031413 ↗InjuryMar 13, 2023Unexpected Therapeutic Results; Unexpected Therapeutic Results
IT WAS REPORTED THAT PATIENT EXPERIENCED INADEQUATE STIMULATION. NO DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED IPGS WERE NOT RETURNED PER HOSPITAL POLICY.
Patient problems: Inadequate Pain Relief
MAUDE report 16533710 ↗InjuryJan 26, 2023Charging Problem; Charging Problem
IT WAS REPORTED THAT THE PATIENTS IPG WOULD NOT HOLD A CHARGE. THE PATIENT UNDERWENT AN IPG REPLACEMENT PROCEDURE AND DOING WELL POST OPERATIVELY. THE EXPLANTED DEVICE WAS DISCARDED AT THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16245253 ↗InjurySep 10, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A REPORT WAS RECEIVED THAT THE PATIENT HAD AN AUTOIMMUNE DISORDER FROM THE PAST. IT WAS ALSO REPORTED THAT FOLLOWING POST PLACEMENT OF SCS THE PATIENT WAS EXPERIENCING SERIOUS DRAINAGE AND WAS TREATED BY INFECTIOUS DISEASE PHYSICIAN. THE PHYSICIAN STATED THAT PATIENT HAD GROWN A NON-STAPHYLOCOCCAL ORGANISM. THE PATIENT WILL UNDERGO EXPLANT.
Patient problems: Complaint, Ill-Defined; Unspecified Infection
MAUDE report 8985886 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030017. One submission often covers many models, so check the product and lot details in each record.
- Z-2879-2026 · Jun 17, 2026 · Open, ClassifiedINFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spin
- Z-2404-2026 · Apr 29, 2026 · Open, ClassifiedBoston Scientific Model SC-4318, Clik X Anchor REF: M365SC43180
- Z-2617-2024 · Jul 17, 2024 · Open, ClassifiedWAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
- Z-2123-2020 · Apr 24, 2019 · TerminatedPrecision TM M8 Adapter, 15 cm, part number M365SC9218150 - Product Usage: The adapter is part of a spinal cord stimulation system for prescription use only indicated as an aid in
- Z-2120-2020 · Apr 24, 2019 · TerminatedThirty-five (35) cm 8 Contact Extension Kit, part number M365SC3138350. Package Contents: Lead Extension Kit (1) Lead Extension (1) Hex Wrench (1) Tunneling Tool Assembly (1) Devic
- Z-2121-2020 · Apr 24, 2019 · TerminatedLinearTM 3-4 50 cm 8 Contact Lead Kit, part number M365SC2352500. Leads function as a component of Boston Scientific's Spinal Cord Stimulation (SCS) systems by delivering electric
- Z-2124-2020 · Apr 24, 2019 · TerminatedFifty-five (55) cm 8 Contact Extension Kit, part number M365SC3138550 - Product Usage: Lead Extensions are designed to connect the percutaneous leads to the Implantable Pulse Gene
- Z-2119-2020 · Apr 24, 2019 · TerminatedArtisanTM 70 cm 2 x 8 Surgical Lead Kit, part number M365SC8216700. 2 x 8 Surgical Paddle Lead Kit: (1) One Paddle Lead (4); Four Suture Sleeves; (2) Lead Position Labels, left
- Z-2122-2020 · Apr 24, 2019 · TerminatedThe LinearTM 3-4 70 cm 8 Contact Lead Kit, part number M365SC2352700 is part of a spinal cord stimulation system for prescription use only as an aid in the management of chronic i
- Z-1409-2010 · Mar 19, 2010 · TerminatedBionic NAVIGATOR Clinician Programmer Model Number SC-7150-4
- Z-0271-2009 · Sep 22, 2008 · TerminatedPrecision Charger 1.0 (Model Number SC-5300) for the Precision Spinal Cord Stimulator System labeled as BIONICS SCS-CHARGING KIT CONTENTS Charger (1) Base Station (1) Belt Po
- Z-1441-2012 · Apr 22, 2008 · TerminatedOMG (Observational Mechanical Gateway) Connector - A Model Number SC-9305 The Precision Spinal Cord Stimulator System is indicated as an aid in the management of chronic intractabl
- Z-0960-2008 · Oct 5, 2007 · TerminatedAdvanced Bionics Precision Implantable Pulse Generator Kit, Model Number: SC-1110 with firmware version prior to Revision 3.02, Advanced Bionics Corportation, Sylmar, CA 91342
- Z-0055-2008 · Aug 3, 2007 · TerminatedPrecision Implantable Pulse Generator (IPG), Model Number SC-110
- Z-0700-2007 · Jan 30, 2007 · TerminatedThe Precision Implantable Pulse Generator, Model Number SC-1110
…and 2 more.
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page