NA 170537-04Recalls & reported problems
Nucletron B.V. · Oncentra External Beam/Oncentra Brachy v4.5.2 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (K121448, K132816); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Dec 5, 2022Class IIOpen, Classified
A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling
Z-0719-2023 · Root cause: Software design · Recall details · FDA record ↗
- Oct 13, 2021Class IIOpen, Classified
Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.
Z-0182-2022 · Root cause: Software design · Recall details · FDA record ↗
FDA clearance and classification
Cleared through 510(k) K132816 on Jun 13, 2014 (decision: Substantially Equivalent). Submitted by Nucletron B.V. as “Oncentra Brachy”. The FDA review took 277 days. FDA record ↗
Cleared through 510(k) K121448 on Jul 2, 2012 (decision: Substantially Equivalent). Submitted by Elekta Inc. D/B/A Nucletron Corporation as “Oncentra, Oncentra External Beam, Oncentra Brachy”. The FDA review took 48 days. FDA record ↗
The FDA regulates product code MUJ (system,planning,radiation therapy treatment) under 21 CFR 892.5050 as a Class II device. 21 CFR 892.5050 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K121448, K132816. One submission often covers many models, so check the product and lot details in each record.
- Z-0050-2018 · Aug 10, 2017 · TerminatedOncentra Brachy 4.5 radiation therapy software The firm name on the label is Nucletron B.V. Oncentra is a radiation therapy planning software designed to analyze and plan radiati
- Z-0987-2017 · Dec 23, 2016 · TerminatedOncentra External Beam Oncentra Brachy Product Usage: Oncentra is radiation therapy planning software designed to analyze and plan radiation treatment is three dimensions for the
- Z-0604-2016 · Dec 18, 2015 · TerminatedOncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radia
- Z-1712-2015 · May 21, 2015 · TerminatedOncentra Brachy radiation therapy planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.
Common questions about the NA 170537-04
Has the NA 170537-04 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0719-2023 (Class II), began on Dec 5, 2022: A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Is the NA 170537-04 FDA approved?
It was cleared through 510(k) K121448 on Jul 2, 2012; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA 170537-04, and is it still sold?
The FDA's device database lists Nucletron B.V. as the company responsible for it (brand NA, model 170537-04). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System,planning,radiation therapy treatment devices
- RayStation 9Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment12 recalls0 reports
- RayStation 9B SP1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment11 recalls0 reports
- RayStation 8.0.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment11 recalls0 reports
- RayStation 10.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment10 recalls1 report
- RayStation 8.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment10 recalls0 reports
- RayStation 10.0.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 8.1.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 10.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page