NA 170538-00Recalls & reported problems
Nucletron B.V. · Oncentra External Beam 4.5.3/Oncentra Brachy v4.6 · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (K121448, K132816); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Dec 5, 2022Open, Classified
A reconstruction error may occur when using the Catheter Bending functionality in Applicator Modeling or Implant Modeling
Z-0719-2023 · Root cause: Software design · FDA record ↗
- Oct 13, 2021Open, Classified
Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.
Z-0182-2022 · Root cause: Software design · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121448, K132816. One submission often covers many models, so check the product and lot details in each record.
- Z-0050-2018 · Aug 10, 2017 · TerminatedOncentra Brachy 4.5 radiation therapy software The firm name on the label is Nucletron B.V. Oncentra is a radiation therapy planning software designed to analyze and plan radiati
- Z-0987-2017 · Dec 23, 2016 · TerminatedOncentra External Beam Oncentra Brachy Product Usage: Oncentra is radiation therapy planning software designed to analyze and plan radiation treatment is three dimensions for the
- Z-0604-2016 · Dec 18, 2015 · TerminatedOncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radia
- Z-1712-2015 · May 21, 2015 · TerminatedOncentra Brachy radiation therapy planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.
Other System, planning, radiation therapy treatment devices
- RayStation 9Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment12 recalls0 reports
- RayStation 8.0.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment11 recalls0 reports
- RayStation 9B SP1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment11 recalls0 reports
- RayStation 10.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment10 recalls1 report
- RayStation 8.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment10 recalls0 reports
- RayStation 8.1.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 10.0Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
- RayStation 10.0.1Raysearch Laboratories AB (Publ) · System, Planning, Radiation Therapy Treatment9 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.