Skip to content

Esprit™ 1203300-18Recalls & reported problems

Abbott Vascular Inc. · Esprit™ BTK Everolimus Eluting Resorbable Scaffold System 3.00mm x 18mm Rapid-Exchange · Class 3 · Implant

All models of Esprit (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648356971), received between Sep 2, 2025 and Sep 2, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2025Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    IT WAS REPORTED THAT THE PROCEDURE WAS TREAT THE OSTIUM OF THE ANTERIOR TIBIAL ARTERY. A 3.0X18MM ESPRIT BELOW THE KNEE (BTK) WAS IMPLANTED WITHOUT ISSUE; HOWEVER, THE 3.0X18MM WAS A SMALLER STENT THAN WHAT THE PHYSICIAN WANTED TO USE BUT WAS ONLY SIZE AVAILABLE AND PLANNED FOR RE-INTERVENTION ON A LATER DATE. ON (B)(6) 2025, WHEN ADVANCING A 4FR X 90CM ANGLED GUIDE CATHETER ATTEMPTING TO CROSS THE PREVIOUSLY IMPLANTED ESPRIT RESISTANCE WAS MET AND THE SCAFFOLD ACCORDIONED AND FOLDED UP ON ITSELF. IT WAS OBSERVED ANGIOGRAPHICALLY AS THE PLATINUM MARKERS WENT FROM 18MM APART TO ABOUT 2MM APART. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22932729 ↗

Other Scaffold, Infrapopliteal, Absorbable devices

All Scaffold, Infrapopliteal, Absorbable devices →

Sources and updates