Evoke® 12C Percutaneous Lead Kit - 60 cm 3008Recalls & reported problems
Saluda Medical Pty Limited · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 09352307000087), received between Jan 3, 2024 and Apr 23, 2026 — classified as 6 injuries.
The most frequently reported specific device problem is “High impedance” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High impedance4
- Defective Device2
- Insufficient Device Problem Information2
Patient problems
- Inadequate Pain Relief4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2026High impedance
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM UNDERWENT A DEVICE CHECK PRIOR TO A MAGNETIC RESONANCE IMAGERY (MRI). DISCONNECTED ELECTRODE WAS IDENTIFIED AND THE PATIENT REQUESTED A SYSTEM EXPLANT TO RESTORE MRI COMPATIBILITY. AT THE TIME OF THIS REPORT, NO ADDITIONAL INFORMATION ON THE EXPLANT PROCEDURE IS AVAILABLE TO SALUDA. SALUDA REPRESENTATIVES ARE NOT EXPECTED TO BE CONSULTED FURTHER ON THIS EVENT.
Patient problems: Inadequate Pain Relief
MAUDE report 24965466 ↗InjuryApr 23, 2026High impedance
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM UNDERWENT A DEVICE CHECK PRIOR TO A MAGNETIC RESONANCE IMAGERY (MRI). DISCONNECTED ELECTRODE WAS IDENTIFIED AND THE PATIENT REQUESTED A SYSTEM EXPLANT TO RESTORE MRI COMPATIBILITY. AT THE TIME OF THIS REPORT, NO ADDITIONAL INFORMATION ON THE EXPLANT PROCEDURE IS AVAILABLE TO SALUDA. SALUDA REPRESENTATIVES ARE NOT EXPECTED TO BE CONSULTED FURTHER ON THIS EVENT.
Patient problems: Inadequate Pain Relief
MAUDE report 24965462 ↗InjuryNov 24, 2025Defective Device; Insufficient Device Problem Information; Defective Device; Insufficient Device Problem Information
IT WAS REPORTED IN MEDWATCH: MW5177282, THAT A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM UNDERWENT A REVISION PROCEDURE DURING WHICH THE SALUDA LEAD WAS REMOVED BECAUSE IT WAS ASSESSED AS DEFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23630357 ↗InjuryNov 24, 2025Defective Device; Insufficient Device Problem Information; Defective Device; Insufficient Device Problem Information
IT WAS REPORTED IN MEDWATCH: MW5177282, THAT A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM UNDERWENT A REVISION PROCEDURE DURING WHICH THE SALUDA LEAD WAS REMOVED BECAUSE IT WAS ASSESSED AS DEFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23630356 ↗InjuryJan 5, 2024High impedance; High impedance
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED THEIR SCS SYSTEM PROGRAM WAS NO LONGER WORKING. IT WAS IDENTIFIED THAT THE SYSTEM HAD DISCONNECTED ELECTRODES. THE PATIENT'S SYSTEM WAS REPROGRAMMED WITH TWO CLOSED-LOOP PROGRAMS. THE PROGRAMS DID NOT PROVIDE THE STABILITY DESIRED, SO THE PATIENT UNDERWENT A REVISION PROCEDURE. ALL IMPEDANCES WERE IN RANGE AND THE PATIENT'S THERAPY WAS RESTORED FOLLOWING THE PROCEDURE.
Patient problems: Inadequate Pain Relief
MAUDE report 18451654 ↗InjuryJan 3, 2024High impedance; High impedance
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED THEIR SCS SYSTEM PROGRAM WAS NO LONGER WORKING. IT WAS IDENTIFIED THAT THE SYSTEM HAD DISCONNECTED ELECTRODES. THE PATIENT'S SYSTEM WAS REPROGRAMMED WITH TWO CLOSED-LOOP PROGRAMS. THE PROGRAMS DID NOT PROVIDE THE STABILITY DESIRED, SO THE PATIENT UNDERWENT A REVISION PROCEDURE. ALL IMPEDANCES WERE IN RANGE AND THE PATIENT'S THERAPY WAS RESTORED FOLLOWING THE PROCEDURE.
Patient problems: Inadequate Pain Relief
MAUDE report 18440048 ↗
Other Stimulator, Spinal-Cord, Totally Implanted For Pain Relief devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page