Evoke® Active Anchor Kit 3043 (UDI 09352307000810)Recalls & reported problems
Saluda Medical Pty Limited · Class 3 · Implant
All models of Evoke Active Anchor Kit (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 09352307000810), received between Nov 27, 2023 and Sep 25, 2024 — classified as 10 injuries.
The most frequently reported specific device problem is “Migration” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Migration7
- Adverse Event Without Identified Device or Use Problem3
- Inappropriate/Inadequate Shock/Stimulation2
- Unintended Movement1
Patient problems
- Inadequate Pain Relief6
- Implant Pain2
- Impaired Healing1
- Undesired Nerve Stimulation1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjurySep 25, 2024Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED A CHANGE IN THERAPY COVERAGE. AN X-RAY WAS OBTAINED WHICH REVEALED MIGRATION OF BOTH LEADS. THE EVOKE SCS SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 20301428 ↗InjurySep 23, 2024Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED A CHANGE IN THERAPY COVERAGE. AN X-RAY WAS OBTAINED WHICH REVEALED MIGRATION OF BOTH LEADS. THE EVOKE SCS SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 20286034 ↗InjurySep 3, 2024Adverse Event Without Identified Device or Use Problem; Unintended Movement; Adverse Event Without Identified Device or Use Problem; Unintended Movement
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EXPLANTED DUE TO A LEAD MIGRATION AND WOUND BREAKDOWN.
Patient problems: Impaired Healing
MAUDE report 20129067 ↗InjuryApr 1, 2024Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED THE LOSS OF STIMULATION. DURING TROUBLESHOOTING, THE REPORTED COVERAGE DID NOT MATCH THE INITIAL LEAD PLACEMENT. AN EPIDURAL INJECTION WAS ADMINISTERED USING FLUOROSCOPY, WHICH VERIFIED THAT THE LEADS HAD MIGRATED. HOWEVER THE PATIENT IS RECEIVING ADEQUATE COVERAGE WITH THEIR CURRENT PROGRAM.
Patient problems: Inadequate Pain Relief
MAUDE report 19016671 ↗InjuryMar 28, 2024Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED THE LOSS OF STIMULATION. DURING TROUBLESHOOTING, THE REPORTED COVERAGE DID NOT MATCH THE INITIAL LEAD PLACEMENT. AN EPIDURAL INJECTION WAS ADMINISTERED USING FLUOROSCOPY, WHICH VERIFIED THAT THE LEADS HAD MIGRATED. HOWEVER THE PATIENT IS RECEIVING ADEQUATE COVERAGE WITH THEIR CURRENT PROGRAM.
Patient problems: Inadequate Pain Relief
MAUDE report 18999711 ↗InjuryMar 13, 2024Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED PAIN AND A MOTOR RESPONSE. THE EVOKE SCS SYSTEM AND A PHYSICIAN EXAM CONFIRMED THE PAIN WAS DUE TO LEAD MIGRATION CAUSING UNINTENDED STIMULATION. A LEAD REVISION WAS PERFORMED THREE DAYS LATER. POST PROCEDURE, THE PATIENT WAS REPORTED TO BE RECEIVING ADEQUATE THERAPY.
Patient problems: Undesired Nerve Stimulation
MAUDE report 18899757 ↗InjuryDec 27, 2023Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED TURNING OVER IN BED AND FEELING A POPPING SENSATION AT THE SPINE INCISION SITE. AN X-RAY REVEALED THE IMPLANTED LEADS HAD MIGRATED. A REVISION PROCEDURE WAS COMPLETED REPLACING THE LEADS AND ANCHORS. FOLLOWING THE REVISION PROCEDURE, THE PATIENT WAS PROGRAMMED INTO CLOSED-LOOP THERAPY WITH ADEQUATE PAIN RELIEF.
Patient problems: Inadequate Pain Relief
MAUDE report 18406552 ↗InjuryDec 23, 2023Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED TURNING OVER IN BED AND FEELING A POPPING SENSATION AT THE SPINE INCISION SITE. AN X-RAY REVEALED THE IMPLANTED LEADS HAD MIGRATED. A REVISION PROCEDURE WAS COMPLETED REPLACING THE LEADS AND ANCHORS. FOLLOWING THE REVISION PROCEDURE, THE PATIENT WAS PROGRAMMED INTO CLOSED-LOOP THERAPY WITH ADEQUATE PAIN RELIEF.
Patient problems: Inadequate Pain Relief
MAUDE report 18389821 ↗InjuryDec 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED EXPERIENCING PAIN AT THE REGION OF THE ANCHOR USED WITH THE SCS LEAD. DURING THE REVISION PROCEDURE, THE LEADS COULD NOT BE FULLY RE-INSERTED INTO THE CLOSED LOOP STIMULATOR (CLS) SO THE SCS SYSTEM WAS REPLACED.
Patient problems: Implant Pain
MAUDE report 18288646 ↗InjuryNov 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED EXPERIENCING PAIN AT THE REGION OF THE ANCHOR USED WITH THE SCS LEAD. DURING THE REVISION PROCEDURE, THE LEADS COULD NOT BE FULLY RE-INSERTED INTO THE CLOSED LOOP STIMULATOR (CLS) SO THE SCS SYSTEM WAS REPLACED.
Patient problems: Implant Pain
MAUDE report 18210978 ↗
Other Stimulator, Spinal-Cord, Totally Implanted For Pain Relief devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗