Evoke ® CAP12 Percutaneous Lead Kit - 60 cm 3008Recalls & reported problems

Saluda Medical Pty Limited · Class 3 · Implant

All models of Evoke CAP12 Percutaneous Lead Kit - 60 cm (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 09352307001589), received between Apr 16, 2024 and Apr 16, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “High impedance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High impedance1
  • Migration1

Patient problems

  • Inadequate Pain Relief1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 16, 2024High impedance; Migration; High impedance; Migration

    A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM RUNNING OPEN LOOP PROGRAMS REPORTED TO NO LONGER BE RECEIVING ADEQUATE STIMULATION. THE PATIENT HAD BEEN PROGRAMMED IN OPEN LOOP PREVIOUSLY DUE TO DISCONNECTED ELECTRODES. TROUBLESHOOTING WAS PERFORMED BUT UNSUCCESSFUL. AN X-RAY WAS PERFORMED WHICH REVEALED THE LEADS HAD MIGRATED. IT WAS REPORTED THAT THE PATIENT HAD EXPERIENCED FREQUENT FALLS, NOT RELATED TO SALUDA SCS SYSTEM. A REVISION PROCEDURE WAS PERFORMED, AND THE SYSTEM WAS REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 19122612 ↗

Other Stimulator, Spinal-Cord, Totally Implanted For Pain Relief devices

All Stimulator, Spinal-Cord, Totally Implanted For Pain Relief devices →