Evoke® Active Anchor Kit 3043 (UDI 09352307001718)Recalls & reported problems
Saluda Medical Pty Limited · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
The Evoke Active Anchor Kit is part of an implanted spinal cord stimulator system used to deliver mild pulses for chronic pain relief. The 30 reports mostly describe leads shifting position inside the body, which patients noticed as a change in stimulation and returning pain; several ended with the system removed or a revision surgery, and one case involved a lead found disconnected before an MRI. All 30 were injuries, with no deaths and no malfunctions reported. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 30
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 09352307001718), received between Mar 12, 2024 and Dec 18, 2025 — classified as 30 injuries.
The most frequently reported specific device problem is “Migration” (26 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Migration26
- Inappropriate/Inadequate Shock/Stimulation15
- Unintended Movement4
- High impedance2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Inadequate Pain Relief19
- No Clinical Signs, Symptoms or Conditions6
- Undesired Nerve Stimulation4
- Impaired Healing1
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
InjuryDec 18, 2025Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED INADEQUATE PAIN RELIEF AFTER A MOTOR VEHICLE ACCIDENT. THE IMPLANTED LEADS WERE REPORTED TO HAVE MIGRATED. A REVISION PROCEDURE WAS COMPLETED AND THERAPY WAS RESTORED.
Patient problems: Inadequate Pain Relief
MAUDE report 23846018 ↗InjuryDec 18, 2025Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED INADEQUATE PAIN RELIEF AFTER A MOTOR VEHICLE ACCIDENT. THE IMPLANTED LEADS WERE REPORTED TO HAVE MIGRATED. A REVISION PROCEDURE WAS COMPLETED AND THERAPY WAS RESTORED.
Patient problems: Inadequate Pain Relief
MAUDE report 23846000 ↗InjuryDec 10, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED A CHANGE IN STIMULATION. AN X-RAY CONFIRMED LEAD MIGRATION. REPROGRAMMING WAS PERFORMED BUT UNABLE TO RESTORE ADEQUATE THERAPY. THE EVOKE SCS SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 23773915 ↗InjuryDec 10, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED A CHANGE IN STIMULATION. AN X-RAY CONFIRMED LEAD MIGRATION. REPROGRAMMING WAS PERFORMED BUT UNABLE TO RESTORE ADEQUATE THERAPY. THE EVOKE SCS SYSTEM WAS EXPLANTED.
Patient problems: Inadequate Pain Relief
MAUDE report 23773914 ↗InjuryOct 31, 2025High impedance; Migration; High impedance; Migration
WHEN PERFORMING THE EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM DEVICE CHECK BEFORE MAGNETIC RESONANCE IMAGING (MRI), DISCONNECTED ELECTRODES WERE IDENTIFIED ON THE RIGHT LEAD. X-RAY CONFIRMED MIGRATION OF RIGHT LEAD. A REVISION PROCEDURE WAS PERFORMED, AND THE PHYSICIAN ELECTED TO REPLACE THE EVOKE SCS SYSTEM AND RESTORE MRI COMPATIBILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23435290 ↗InjuryOct 31, 2025High impedance; Migration; High impedance; Migration
WHEN PERFORMING THE EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM DEVICE CHECK BEFORE MAGNETIC RESONANCE IMAGING (MRI), DISCONNECTED ELECTRODES WERE IDENTIFIED ON THE RIGHT LEAD. X-RAY CONFIRMED MIGRATION OF RIGHT LEAD. A REVISION PROCEDURE WAS PERFORMED, AND THE PHYSICIAN ELECTED TO REPLACE THE EVOKE SCS SYSTEM AND RESTORE MRI COMPATIBILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23435281 ↗InjurySep 2, 2025Migration; Migration
SALUDA REPRESENTATIVES WERE NOTIFIED OF INADEQUATE PAIN RELIEF AND LEAD MIGRATION FOR A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM. AN EXPLANT WAS REPORTED TO BE COMPLETED TO RESOLVE THE REPORTED EVENT. SALUDA REPRESENTATIVES WERE INITIALLY NOTIFIED OF THIS EVENT THROUGH A 3RD PARTY AND NO FURTHER INFORMATION HAS BEEN RECEIVED. THE EXPLANT DATE WAS NOT CONFIRMED AND THE EXACT DATE OF EXPLANT UNKNOWN. (B)(6) 2025 IS SELECTED AS AN APPROXIMATE EVENT DATE.
Patient problems: Inadequate Pain Relief
MAUDE report 22936914 ↗InjurySep 2, 2025Migration; Migration
SALUDA REPRESENTATIVES WERE NOTIFIED OF INADEQUATE PAIN RELIEF AND LEAD MIGRATION FOR A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM. AN EXPLANT WAS REPORTED TO BE COMPLETED TO RESOLVE THE REPORTED EVENT. SALUDA REPRESENTATIVES WERE INITIALLY NOTIFIED OF THIS EVENT THROUGH A 3RD PARTY AND NO FURTHER INFORMATION HAS BEEN RECEIVED. THE EXPLANT DATE WAS NOT CONFIRMED AND THE EXACT DATE OF EXPLANT UNKNOWN. (B)(6) 2025 IS SELECTED AS AN APPROXIMATE EVENT DATE.
Patient problems: Inadequate Pain Relief
MAUDE report 22936357 ↗InjuryJul 23, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR INADEQUATE PAIN RELIEF. THE PATIENT REPORTED INCREASING STIMULATION IN THE ABDOMEN AREA AND ONE (1) EVENT OF UNEXPECTED INTERRUPTION OF THERAPY IN CLOSED LOOP. X-RAY IMAGING WAS OBTAINED WHICH CONFIRMED A LEAD MIGRATION. THE PATIENT WAS REPROGRAMMED TO TEMPORARILY RESTORE THERAPY, PENDING A LEAD REVISION TO RESOLVE THE LEAD MIGRATION. A LEAD REVISION PROCEDURE WAS COMPLETED THERAPY RESTORED TO THE PATIENT.
Patient problems: Inadequate Pain Relief
MAUDE report 22579632 ↗InjuryApr 11, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR INSUFFICIENT PAIN RELIEF.. TROUBLESHOOTING WAS PERFORMED, INCLUDED REPROGRAMMING. X-RAYS SHOW MIGRATION OCCURRED. DURING THE LEAD REVISION, REPOSITIONING WAS NOT AVAILABLE DUE TO SCAR TISSUE, THEREFORE; BOTH LEADS WERE REPLACED RESTORING ADEQUATE PAIN RELIEF.
Patient problems: Inadequate Pain Relief
MAUDE report 21827891 ↗InjuryApr 11, 2025Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR INSUFFICIENT PAIN RELIEF. TROUBLESHOOTING WAS PERFORMED, INCLUDED REPROGRAMMING. X-RAYS SHOW MIGRATION OCCURRED. DURING THE LEAD REVISION, REPOSITIONING WAS NOT AVAILABLE DUE TO SCAR TISSUE, THEREFORE; BOTH LEADS WERE REPLACED RESTORING ADEQUATE PAIN RELIEF.
Patient problems: Inadequate Pain Relief
MAUDE report 21827890 ↗InjuryMar 25, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR STIMULATION AT THE CLOSED LOOP STIMULATOR (CLS) IMPLANT SITE RECEIVING THERAPY. AN X-RAY WAS OBTAINED WHICH CONFIRMED LEAD MIGRATION TO THE CLS IMPLANT SITE. A REVISION PROCEDURE WAS COMPLETED TO RESOLVE THE REPORTED EVENT.
Patient problems: Inadequate Pain Relief
MAUDE report 21689994 ↗InjuryMar 25, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR STIMULATION AT THE CLOSED LOOP STIMULATOR (CLS) IMPLANT SITE RECEIVING THERAPY. AN X-RAY WAS OBTAINED WHICH CONFIRMED LEAD MIGRATION TO THE CLS IMPLANT SITE. A REVISION PROCEDURE WAS COMPLETED TO RESOLVE THE REPORTED EVENT.
Patient problems: Inadequate Pain Relief
MAUDE report 21689990 ↗InjuryFeb 26, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR A CHANGE IN STIMULATION. A LEAD MIGRATION WAS CONFIRMED. THE PATIENT REPORTED A FALL PRIOR TO THE REPORTED EVENT. THE EVOKE SCS SYSTEM DID NOT CAUSE OR CONTRIBUTE TO THE PATIENT¿S FALLS. A LEAD REVISION WAS COMPLETED AND THERAPY RESTORED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21469447 ↗InjuryFeb 26, 2025Migration; Migration
A PATIENT WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR A CHANGE IN STIMULATION. A LEAD MIGRATION WAS CONFIRMED. THE PATIENT REPORTED A FALL PRIOR TO THE REPORTED EVENT. THE EVOKE SCS SYSTEM DID NOT CAUSE OR CONTRIBUTE TO THE PATIENT¿S FALLS. A LEAD REVISION WAS COMPLETED AND THERAPY RESTORED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21469445 ↗InjuryFeb 7, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED INADEQUATE PAIN RELIEF. A LEAD MIGRATION WAS CONFIRMED. A LEAD REVISION WAS PERFORMED TO REPLACE THE LEAD AND ANCHOR. THERAPY HAS BEEN RESTORED TO THE PATIENT.
Patient problems: Inadequate Pain Relief
MAUDE report 21332460 ↗InjuryFeb 7, 2025Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED INADEQUATE PAIN RELIEF. A LEAD MIGRATION WAS CONFIRMED. A LEAD REVISION WAS PERFORMED TO REPLACE THE LEAD AND ANCHOR. THERAPY HAS BEEN RESTORED TO THE PATIENT.
Patient problems: Inadequate Pain Relief
MAUDE report 21332459 ↗InjuryJan 31, 2025Migration; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR INADEQUATE PAIN STIMULATION. X-RAY IMAGING WAS OBTAINED WHICH CONFIRMED LEAD MIGRATION. A LEAD REVISION WAS COMPLETED TO RESTORE THERAPY. LEAD INFORMATION: BRAND NAME - EVOKE CAP12 PERCUTANEOUS LEAD KIT - 60CM (US), MODEL - 102878, CATALOG - 3008, LOT/BATCH NUMBER - 9016004179.
Patient problems: Inadequate Pain Relief
MAUDE report 21285051 ↗InjuryJan 31, 2025Migration; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR INADEQUATE PAIN STIMULATION. X-RAY IMAGING WAS OBTAINED WHICH CONFIRMED LEAD MIGRATION. A LEAD REVISION WAS COMPLETED TO RESTORE THERAPY.
Patient problems: Inadequate Pain Relief
MAUDE report 21284978 ↗InjuryJan 16, 2025Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR LEAD MIGRATION. AN X-RAY WAS OBTAINED AND CONFIRMED MIGRATION OF BOTH LEADS. A REVISION PROCEDURE WAS COMPLETED TO RESOLVE THE ISSUE AND RESTORE THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21175667 ↗InjuryJan 16, 2025Migration; Migration
A PATIENT IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM WAS EVALUATED FOR LEAD MIGRATION. AN X-RAY WAS OBTAINED AND CONFIRMED MIGRATION OF BOTH LEADS. A REVISION PROCEDURE WAS COMPLETED TO RESOLVE THE ISSUE AND RESTORE THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21175620 ↗InjuryDec 4, 2024Inappropriate/Inadequate Shock/Stimulation; Unintended Movement; Inappropriate/Inadequate Shock/Stimulation; Unintended Movement
DURING ROUTINE PROGRAMMING, A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED STIMULATION IN THEIR RIGHT UPPER EXTREMITY. AN X-RAY WAS OBTAINED AND CONFIRMED A LEAD MIGRATION. A LEAD REVISION WAS COMPLETED TO RESOLVE THE ISSUE AND RESTORE THERAPY. IT WAS ADDITIONALLY REPORTED THAT THE PATIENT¿S CLOSED LOOP STIMULATOR (CLS) WAS REPLACED AND A NEWER DEVICE MODEL IMPLANTED. THE REASON FOR THE DEVICE REPLACEMENT WAS NOT DISCLOSED TO SALUDA.
Patient problems: Undesired Nerve Stimulation
MAUDE report 20846677 ↗InjuryDec 4, 2024Inappropriate/Inadequate Shock/Stimulation; Unintended Movement; Inappropriate/Inadequate Shock/Stimulation; Unintended Movement
DURING ROUTINE PROGRAMMING, A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED STIMULATION IN THEIR RIGHT UPPER EXTREMITY. AN X-RAY WAS OBTAINED AND CONFIRMED A LEAD MIGRATION. A LEAD REVISION WAS COMPLETED TO RESOLVE THE ISSUE AND RESTORE THERAPY. IT WAS ADDITIONALLY REPORTED THAT THE PATIENT¿S CLOSED LOOP STIMULATOR (CLS) WAS REPLACED AND A NEWER MODEL IMPLANTED. THE REASON FOR THE DEVICE REPLACEMENT WAS NOT DISCLOSED TO SALUDA.
Patient problems: Undesired Nerve Stimulation
MAUDE report 20845816 ↗InjuryOct 31, 2024Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED INADEQUATE DECREASE IN PAIN RELIEF AND UNINTENDED STIMULATION. THE PATIENT REPORTED THAT THEY HAD BEEN INVOLVED IN PHYSICAL ACTIVITIES WHICH INVOLVED, INCLUDING LIFTING HEAVY OBJECTS. AN X-RAY WAS OBTAINED WHICH REVEALED MIGRATION OF THE RIGHT LEAD. A REVISION WAS PERFORMED AND THE EVOKE SCS SYSTEM WAS REPLACED. THERAPY HAS BEEN RESTORED FOLLOWING THE REVISION PROCEDURE.
Patient problems: Inadequate Pain Relief
MAUDE report 20584907 ↗InjuryOct 31, 2024Inappropriate/Inadequate Shock/Stimulation; Migration; Inappropriate/Inadequate Shock/Stimulation; Migration
A PATIENT PREVIOUSLY IMPLANTED WITH AN EVOKE SPINAL CORD STIMULATION (SCS) SYSTEM REPORTED DECREASE IN PAIN RELIEF AND UNINTENDED STIMULATION. THE PATIENT REPORTED THAT THEY HAD BEEN INVOLVED IN PHYSICAL ACTIVITIES, INCLUDING LIFTING HEAVY OBJECTS. AN X-RAY WAS OBTAINED WHICH REVEALED MIGRATION OF THE RIGHT LEAD. A REVISION WAS PERFORMED AND THE EVOKE SCS SYSTEM WAS REPLACED. THERAPY HAS BEEN RESTORED FOLLOWING THE REVISION PROCEDURE.
Patient problems: Inadequate Pain Relief
MAUDE report 20584906 ↗
Other Stimulator, Spinal-Cord, Totally Implanted For Pain Relief devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page