Skip to content

AirLife™ 2K8004Recalls & reported problems

Airlife · Adult Manual Resuscitator • Oxygen Reservoir Bag, Adult Mask · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10190752114104), received between Sep 24, 2021 and Oct 23, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 10, 2024Class IOpen, Classified

    Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

    Z-1065-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inflation Problem2
  • Defective Device1
  • Gas/Air Leak1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1
  • Pulmonary Infarction1
  • Respiratory Insufficiency1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED, WHEN THE RESERVOIR BAG, ON THE ADULT RESUSCITATOR MASK WAS SQUEEZED, AIR ESCAPED. THERE WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23375232 ↗
  • InjuryJun 21, 2024Material Integrity Problem; Material Integrity Problem

    RESERVOIR BAG FAILED TO INFLATE WHILE BEING ATTACHED TO HIGH FLOW OXYGEN. PATIENT WAS ADMITTED TO THE ICU.

    Patient problems: Respiratory Insufficiency

    MAUDE report 19586649 ↗
  • InjuryJan 19, 2024Inflation Problem; Inflation Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 2K8004 - RESUS, ADLT W/MASK, RESV BAG, 6/CS DID NOT INFLATE WHILE THE PATIENT WAS CODING. THE NURSE PULLED ANOTHER BAG, AND IT ALSO WOULDN'T INFLATE. A SECOND AND THIRD BAG WERE PULLED, AND THEY ALSO DID NOT INFLATE. FINALLY, THE FOURTH BAG INFLATED. THE PATIENT'S IMPACT OR HARM IS CURRENTLY UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 18544358 ↗
  • MalfunctionNov 7, 2021Inflation Problem; Inflation Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT THE 2K8004- AIRLIFE¿ ADULT MANUAL RESUSCITATOR, 40'' (1.0M) OXYGEN RESERVOIR TUBING, ADULT M DID NOT INFLATE DUE TO MISSING PARTS. THIS ISSUE OCCURRED WHILE USED ON A PATIENT. AS AN INTERVENTION, THE SAMPLE WAS REPLACED. AT THIS TIME, THERE IS NO INFORMATION ON PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12763360 ↗
  • InjurySep 24, 2021Defective Device; Defective Device

    THE CUSTOMER REPORTED THAT THE 2K8004 RESUS, ADLT W/MASK, RESV BAG RESERVOIR ON THE AMBU BAGS WAS WRAPPED AT THE DISTAL END OPENING WHERE THE TUBING GOES THROUGH THE RESERVOIR, IT WAS FOLDED OVER BLOCKING AIR FROM ESCAPING CAUSING EXCESSIVE PRESSURES WHICH RESULTED IN A PNEUMOTHORAX WITH ONE PATIENT.

    Patient problems: Pulmonary Infarction

    MAUDE report 12525437 ↗

Other Cardiopulmonary resuscitation aid kit devices

All Cardiopulmonary resuscitation aid kit devices →

Sources and updates